Reference Manual00809-0100-3156Rev. ADMarch 2020Rosemount 1056Dual-Input Intelligent Analyzer

Essential InstructionsREAD THIS PAGE BEFORE PROCEEDING!Your instrument purchase from Emerson is one ofthe finest available for your particular application.These instruments have been designed, and testedto meet many national and international standards.Experience indicates that its performance is directlyrelated to the quality of the installation and knowledge of the user in operating and maintaining theinstrument. To ensure their continued operation tothe design specifications, personnel should read thismanual thoroughly before proceeding with installation, commissioning, operation, and maintenance ofthis instrument. If this equipment is used in a manner not specified by the manufacturer, the protectionprovided by it against hazards may be impaired. Failure to follow the proper instructions maycause any one of the following situations tooccur: Loss of life; personal injury; property damage; damage to this instrument; and warrantyinvalidation. Ensure that you have received the correct modeland options from your purchase order. Verify thatthis manual covers your model and options. Ifnot, call 1-800-854-8257 or 949-757-8500 torequest correct manual. For clarification of instructions, contact yourRosemount representative. Follow all warnings, cautions, and instructionsmarked on and supplied with the product. Use only qualified personnel to install, operate,update, program and maintain the product. Educate your personnel in the proper installation,operation, and maintenance of the product. Install equipment as specified in the Installationsection of this manual. Follow appropriate localand national codes. Only connect the product toelectrical and pressure sources specified in thismanual. Use only factory documented components forrepair. Tampering or unauthorized substitution ofparts and procedures can affect the performanceand cause unsafe operation of your process. All equipment doors must be closed and protective covers must be in place unless qualified personnel are performing maintenance.WARNINGRISK OF ELECTRICAL SHOCKEquipment protected throughout by double insulation. Installation and servicing of this product may expose personelto dangerous voltages. Main power wired to separate power source must bedisconnected before servicing. Do not operate or energize instrument with case open! Signal wiring connected in this box must be rated at least240 V. Non-metallic cable strain reliefs do not provide groundingbetween conduit connections! Use grounding type bushingsand jumper wires. Unused cable conduit entries must be securely sealed bynon-flammable closures to provide enclosure integrity incompliance with personal safety and environmental protectionrequirements. Unused conduit openings must be sealed withNEMA 4X or IP65 conduit plugs to maintain the ingressprotection rating (NEMA 4X). Electrical installation must be in accordance with the NationalElectrical Code (ANSI/NFPA-70) and/or any other applicablenational or local codes. Operate only with front panel fastened and in place. Safety and performance require that this instrument beconnected and properly grounded through a three-wirepower source. Proper use and configuration is the responsibility of theuser.CAUTIONThis product generates, uses, and can radiate radio frequencyenergy and thus can cause radio communication interference.Improper installation, or operation, may increase such interference. As temporarily permitted by regulation, this unit has notbeen tested for compliance within the limits of Class A computing devices, pursuant to Subpart J of Part 15, of FCC Rules,which are designed to provide reasonable protection againstsuch interference. Operation of this equipment in a residentialarea may cause interference, in which case the user at his ownexpense, will be required to take whatever measures may berequired to correct the interference.WARNINGPhysical accessUnauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment.This could be intentional or unintentional and needs to be protected against.Physical security is an important part of any security program and fundamental to protecting your system. Restrictphysical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.

Quick Start GuideRosemount 1056 Dual-Input Intelligent Analyzer1. Refer to Section 2.0 for mechanical installation instructions.2.Wire sensor(s) to the signal boards. See Section 3.0 for wiring instructions. Refer to the sensor instructionsheet for additional details. Make current output, alarm relay and power connections.3. Once connections are secured and verified, apply power to the analyzer.WARNINGRISK OF ELECTRICAL SHOCKElectrical installation must be in accordance withthe National Electrical Code (ANSI/NFPA-70)and/or any other applicable national or local codes.4. When the analyzer is powered up for the first time, Quick Start screens appear. Quick Start operating tipsare as follows:a. A backlit field shows the position of the cursor.b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or downor to increase or decrease the value of a digit use the keys above and below the ENTER key . Use theleft or right keys to move the decimal point.c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT during QuickStart returns the display to the initial start-up screen (select language).5. Complete the steps as shown in the Quick Start Guide flow diagram, Fig. A on the following page.6. After the last step, the main display appears. The outputs are assigned to default values.7.To change output, and temperature-related settings, go to the main menu and choose Program. Follow theprompts. For a general guide to the Program menu, see the Quick Reference Guide, Fig.B.8. To return the analyzer to the default settings, choose Reset Analyzer under the Program menu.

Figure A. Quick Start GuideQuick Start Guide

Figure B. Model 1056 Menu TreeQuick Reference Guide

Reference Manual00809-0100-3156Table of ContentsMODEL 1056DUAL INPUT INTELLIGENT ANALyzERMarch 2020TABLE OF CONTENTSQUICK START GUIDEQUICK REFERENCE GUIDETABLE OF CONTENTSSection Title1.0DESCRIPTION AND SPECIFICATIONS .2.0INSTALLATION .2.1Unpacking and 3.4WIRING.General .Preparing Conduit Openings.Preparing Sensor Cable .Power, Output, Alarms and Sensor Connections.5.15.25.35.45.55.65.75.85.96.0General .Changing the StartUp Settings .Choosing Temperature units and Automatic/Manual Temperature Compensation .Configuring and Ranging the Current Outputs.Setting a Security Code .Security Access.Using Hold .Resetting Factory Defaults – Reset Analyzer .Alarm Relays.PROGRAMMING - MEASUREMENTS .333334343637373839436.6.1 Free Chlorine .534.04.14.24.34.45.06.16.26.36.46.56.6DISPLAy AND OPERATION .User Interface .Instrument Keypad.Main Display .Menu System .PROGRAMMING – BASICS .Programming Measurements – Introduction .pH .ORP .Contacting Conductivity .Toroidal Conductivity.Chlorine.6.6.2 Total Chlorine .6.6.3 Monochloramine .6.76.86.6.4 pH-independent Free Chlorine .Oxygen.Ozone .i21212122222929293031334344454750535556575960

Reference ManualTable of Contents00809-0100-31566.9March 2020TABLE OF CONTENTS CONT’DTurbidity .626.11Current Input .667.2pH Calibration .6.107.07.17.37.47.57.6Flow .CALIBRATION .Calibration – Introduction .7777ORP Calibration .80Contacting Conductivity Calibration .7.6.1 Free Chlorine .86Chlorine Calibration .889092Ozone Calibration .97Oxygen Calibration .7.9Temperature Calibration.8.0867.6.4 pH-Independent Free Chlorine .7.77.1181847.6.3 Monochloramine .7.1078Toroidal Conductivity Calibration .7.6.2 Total Chlorine .7.8659499Turbidity .100RETURN OF MATERIAL .Warranty.115115Pulse Flow .ii102

Reference Manual00809-0100-3156Section 1.0: Description and SpecificationSection 1.0Description and SpecificationsMarch 2020 MULTI-PARAMETER INSTRUMENT – single or dual input. Choose from pH/ORP/ISE,Resistivity/Conductivity, % Concentration, Chlorine, Oxygen, Ozone, Temperature, Turbidity, Flow,and 4-20mA Current Input. LARGE DISPLAY – large easy-to-read process measurements. EASY TO INSTALL – modular boards, removable connectors, easy to wire power, sensors, and outputs. INTUITIVE MENU SCREENS with advanced diagnostics and help screens. SEVEN LANGUAGES incl

Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facili-ty. Quick Start Guide Rosemount 1056 Dual-Input Intelligent Analyzer 1. Refer to Section 2.0 for mechanical installation instructions. 2. Wire sensor(s) to .