Pfeiffer Adixen ASM380 Leak Detectors Family Brochure - Ideal Vac

Transcription

Leak detectorsThe widest offer of leak testing solutions,using helium and hydrogen1

Leak detectors2

ContentsLeak detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Applications and benefits . . . . . . . . . . . . . . . . . . . . . . . . .Overview of series and applications . . . . . . . . . . . . . . . . .Overview of test methods . . . . . . . . . . . . . . . . . . . . . . . . .4568Portable – MiniTest 300 . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Technical data / Accessories / Order numbers . . . . . . . . . 13Portable – ASM 310 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Technical data / Accessories / Order numbers . . . . . . . . . 17Multipurpose – ASM 340, ASM 340 D . . . . . . . . . . . . . . . 18Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Technical data / Accessories / Order numbers . . . . . . . . . 21High performance mobile – ASM 380 . . . . . . . . . . . . . . . 22Dimensions / Technical data . . . . . . . . . . . . . . . . . . . . . . . 24Accessories / Order numbers. . . . . . . . . . . . . . . . . . . . . . . 25High performance compact –ASM 182 T, ASM 182 TD . . . . . . . . . . . . . . . . . . . . . . . . 26Dimensions / Technical data . . . . . . . . . . . . . . . . . . . . . . . 28Order numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29High performance console –ASM 192 T, ASM 192 T2D . . . . . . . . . . . . . . . . . . . . . . . 30Dimensions / Technical data . . . . . . . . . . . . . . . . . . . . . . . 32Order numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Workstation – ASM 1002 . . . . . . . . . . . . . . . . . . . . . . . . . . 34Dimensions / Technical data . . . . . . . . . . . . . . . . . . . . . . . 36Order numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Modular – ASI 30 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40Technical data / Order numbers . . . . . . . . . . . . . . . . . . . . . 41Sniffer leak detector – ASM 102 S . . . . . . . . . . . . . . . . . . 42Dimensions / Technical data . . . . . . . . . . . . . . . . . . . . . . . 44Order numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45General accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Order numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Helium leak detection systems . . . . . . . . . . . . . . . . . . . . . 48Helium recovery units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Vacuum solutions from a single source . . . . . . . . . . . . . . 52Market segments and applications . . . . . . . . . . . . . . . . . . 52Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 563

Leak detectorsAdvantages and applications4

Leak detectors forvarious needs:Our portable leak detectors are used both for mobile leakdetection „on site“, as well as for worldwide service jobs in avariety of different locations. They stand out above all due totheir high performance combined with low weight.These multipurpose leak detectors are used mainly for stationary applications to localize leaks and to check leak tightness.They are also movable thanks to a dedicated cart. Compacthelium leak detectors are used in a wide range of industries,from heavy industry right up to clean room applications. Inthis latter case, the dry versions of the leak detector are used.High-performance leak detectors are used where extremelyshort cycle times and fast reduction of helium background arerequired. From testing of industrial components in highthroughput production up to highly sensitive applications invacuum and medical technology or in research and development.Customer benefits:Portable:Ultralight and easy to operateQ Ideal for global servicing workQ Remote controlled for comfortable operationQMultipurpose:Q Suitable for a large variety of leak detection applicationsQ Powerful and dependableQ Simple design for easy operationHigh performance:Q Extremely short pump down times even on large volumesQ Designed for ultra sensitive leak detection limits whateversize and volume of the test partsQ High reliability in clean processes as well as roughenvironmentsWorkstation:Q Ideal for tests in mass production with short cycle timesQ Fully automatic operation on sealed componentsQ Ergonomic design and operation for seated operatorsModular:Q Easy and flexible integration into leak detection systemsQ Low maintenance for continuous operationQ Broad selection of interfaces to answer all commonindustry standardsSniffing:Q Ideal solution for accurately locating minor leaksQ Status messages by voice synthesizer for the user‘sconvenienceQ Remote controlled for comfortable operation5

Leak detectorsSeries and applicationsOverview of series and applicationsQ recommendedPortableMiniTest 300ASM 310MultipurposeASM 340ASM 340 DHigh-performance –mobileASM 380High-performance –compactASM 182 TASM 182 TD High-performance –consoleASM 192 TASM 192 T2D WorkstationASM 1002ModularASI 30SniffingASM 102 S6SemiconductorCoatingIndustryR&DElectron microscopyLeak detectionMass spectrometrySurface analysisPlasma monitoringResidual gas analysisLithographyPVD (Physical Vapor Deposition)CVD (Chemical Vapor Deposition)Plasma etchingImplantation – SourceImplantation – BeamlineInspectionBondingMBE (Molecular Beam Epitaxy)Load-locks, transfer chambers, handling systemsFlat Panel Display (FPD)LED / OLEDHard disk coatingPhotovoltaicsGlass coating (PVD)CD- / DVD- / Blu-Ray manufacturing (PVD)Optical coating (PVD)Wear protection (PVD, CVD)Web coatingMedical technologyIndustrial leak detectionElectron beam weldingIsolation vacuumsLamp and tube manufacturingHeat treatmentVacuum dryingVacuum furnacesNuclear researchFusion technologyPlasma researchParticle acceleratorsSpace simulationCryogenic researchElementary particle physicsNanotechnologyBiotechnologyAnalyticsQ Q Q QQQQ Q Q Q Q Q Q Q Q Q QQ Q Q Q QQ Q Q QQ Q Q Q Q Q Q Q QQ Q Q QQQ Q Q Q Q Q Q Q Q Q Q QQ Q Q QQ Q Q QQ Q Q Q Q Q Q Q Q Q Q Q Q Q Q QQ Q Q Q Q Q Q Q Q Q Q Q Q Q Q QQ Q Q Q Q Q Q Q Q Q Q Q Q Q Q Q Q Q Q QQ Q Q Q Q Q Q Q Q Q Q Q Q Q Q Q Q Q Q QQ Q Q QQ Q Q Q Q Q Q Q Q Q Q Q QQ QQ Q Q Q QQ Q Q QQ Q Q QQQQ Q Q Q Q Q Q Q Q Q Q Q QQ QQ QQ Q QQ Q Q Q Q Q Q Q Q Q Q Q Q Q QQ Q Q Q QQQ Q Q Q Q Q Q QQQ Q Q Q Q Q Q Q Q Q Q QQQQQQ Q Q Q Q Q Q Q QQ Q Q Q Q Q Q Q Q Q Q Q QQ Q Q Q Q Q Q QQ QQQ Q QQQ Q Q Q Q Q Q Q

Series at a glancePortableMiniTest 300High-performance –mobileASM 380WorkstationASM 1002MultipurposeASM 310High-performance –compactASM 182 TASM 182 TD ModularASI 30ASM 340ASM 340 DHigh-performance –consoleASM 192 TASM 192 T2D SniffingASM 102 S7

Leak detectorsSix test methods for leak detectionLeak localizationApplications in production and maintenance as well as quality controlMethodVacuum test:Spraying testSniffing testFor leak localization on partsunder vacuumFor leak localization onpressurized partsDescriptionThe leak detector evacuatesthe air inside the test part.After that, the tracer gas issprayed on the external surface of the part. The detectormeasures the gas flowthrough the leak channel inthe wall of the test part. Theleak can be localized.The test part is pressurizedwith tracer gas. After that,the sniffer probe is movedaround the part. If a leak ispresent, the leak detector willdetect the escaping tracergas, allowing to locate theleak.Customer benefitsQQQQ8Localization of the leakVery high sensitivityEasy to performLocal or integral testQQQLocalization of the leakNot necessary to put thetest part under vacuumEasy to perform

Vacuum test:Spraying testSniffing testIntegralvacuum testVacuum test:Bombing testIntegral test ofenclosed partsunder vacuumSniffing test:Integral test atathomsphericpressureMiniTestQASM 310QQASM 340 / 340 DQQQQQASM 380QQQQQASM 182 T / TD QQQQQQASM 192 T / T2D QQQQQQASM 1002QQQQQQASI 30QQQQASM 102 SQIntegral testIndustrial applications and quality controlStep 11112Step 211011284712110396 52398476 5Integral vacuum testVacuum test:Bombing testIntegral test of enclosedparts under vacuumSniffing test:Integral test atatmospheric pressureUsed in productionenvironmentsThe test part is sealed andcannot be evacuated orpressurized.It needs to be tested whethertracer gas can enter into thepart.The test part can withstandoverpressure and needs to betested as a complete product.The test part is placed in avacuum test chamber andfilled with tracer gas. Shoulda leak be present, the tracergas will escape from the partinto the test chamber andwill be measured by the leakdetector.Place the test part in a specific chamber and pressurizeit with tracer gas (bombingchamber). Should a leak bepresent, the tracer gas isforced into the part due tothe gas pressure. After that,the part is placed in a vacuum chamber which is evacuated.Any tracer gas that wasforced inside the part willnow escape and be measured by the leak detector.Inside a test chamber, thepart is connected to a leakdetector and evacuated.The chamber is filled withtracer gas. Should a leakbe present, the tracer gaswill penetrate the part andbe measured by the leakdetector.The test part is pressurizedwith tracer gas in a simpleaccumulation chamber whichis under atmospheric pressure. After an accumulationtime, the detector analyzesthe air inside the chamberand determines if an increasein the tracer gas concentration can be measured.QQQQQVery high sensitivityHigh throughputEasy to integrate into aproduction lineEasy calibrationHigh repeatabilityQQThe only solution to testsealed components withhigh sensitivityHigh repeatabilityQQQQHigh automationlevel possibleVery high sensitivityHigh throughputHigh repeatabilityQEasy to integrate into aproduction line9

Portable – MiniTest 300Portable helium leak detector for industrial useInnovative technology andlow weightThe MiniTest 300 helium leak detector weighs only a sensational 5 kg. It is the first vacuum leak detector on the marketwhich works with a quartz window sensor. Thanks to thisinnovative technology, it stands out due to its lightweightdesign and small footprint.The uncomplicated operation and an optional wireless remotecontrol enable helium leak detection to be carried out even inlarge systems by only one person.The MiniTest 300 can work at a high inlet pressure of up to200 hPa and is not sensitive to water vapor. This makes it theideal solution for leak detection in power plants, vacuum furnaces and vacuum systems as well as in heat exchangers.The data recording function on the remote control via USBinterface makes this unit a user-friendly data logger. Since itis practically maintenance-free, it is also cost-efficient both inoperation and in long-term use.Remote control RC 500 WLAll functions of the helium leak detector can be controlledfrom a distance of up to 100 m. The measured data can beeasily read on the 3.5 inch touchscreen display and saved inthe remote controlRemote control RC 500 WLCustomer benefitsQQQQQQQQ10Vacuum leak detection with a quartz window sensor forlocalizing and quantifying even the small leaksCompact size and lightweight (5 kg) -Ideal for servicingworkWireless remote control with extensive range and memoryEasy touchscreen operation with intuitive menu navigationGraphic displayDirect connection possible to systems with a maximumworking pressure of up to 200 hPaLeaks of up to 5 · 10-8 Pa m3/s detectable at a pumpingspeed of 1 l/sLow maintenance costs

ApplicationsQQQQQQQVacuum furnacesLamp manufacturingVacuum coating systemsAir-conditioning and refrigerationPharmaceuticsPower engineeringChemistry11

Portable – MiniTest 300The portable helium leak detector for industrial useDimensions265,215224,8306,826,8204,8Dimensions in mm12

Technical dataMiniTest 300Flange (in)DN 25 KFSound pressure level 70 dB (A)Power consumptionmax. 30 WPower inputmax. 2 AInstallation altitude, max.2,000 mCold start to ready30 minInlet pressure max.200 hPa absoluteWeight5 kgSmallest detectable leak rate for 4He (to AVS 2.1)5 · 10-8 Pa m3/s1)InterfacesRS-485, gauge, remote control, radio transmitter, I/O D-Sub 9Operating temperature: 10 to 35 CStorage temperature:-10 to 55 CPower supply100-240 V, 50/60 HzBacking pumpnone1)at a pumping speed of the system pump of S 1 l/sOrder numbersOrder numberMiniTest 300PT L03 000Order numbers accessoriesAccessoriesOrder numberRemote control RC 500 WL, wirelessPT 445 420-TRemote control RC 500PT 445 421-TTest leak in pin-type casing with adapter DN 16 ISO-KFLeakrate 3 · 10-3 Pa m3/sPT 445 425Leakrate 3 · 10-4 Pa m3/sPT 445 426Leakrate 3 · 10-5 Pa m3/sPT 445 427Transport casePT 445 428Extension cablePT 445 402Helium spray gunBG 512 125 -TCentering ring with poral filterPF 117 225 -T13

Portable – ASM 310Combining light weight and superior performancePowerfulHighly compact and light in weight, the leak detectorASM 310 requires low maintenance. Its low weight anduniversal voltage enable the ASM 310 to be easily operatedanywhere in the world. A transport case to safeguard againstshipping damage and a trolley are available as accessories.Superior performanceQQQQUser-friendlyBacking capacityPumping speed for HeInlet pressureMinimum detectable leak rate1.7 m³/h1.1 l/s15 hPa5 · 10-13 Pa m3/sThe control panel integrates magnets, enabling the operatorto place it on a metallic base.The unique color touchscreen display can be individuallyadjusted. As a result, only information that the user actuallyneeds for his task are displayed. The menu is password-protected to prevent unauthorized access to the settings of thedevice.CompatibleDetachable control panel(with approx. 2 m long cable)The ASM 310 can likewise be operated with the wirelessremote control RC 500 WL. This enables the leak detector tobe operated even from a distance of up to 100 meters.Remote control RC 500 WLCustomer benefitsQQQQQQQQQQQQQ14Lightweight and portable, only 21 kgSmart design with retractable handleEasy to moveDetachable control panelOn-demand operating interfaceIntuitive and customizable menuSmall footprint, small sizeCan be operated in any positionLarge, bright color touchscreenColor graphics functionalityPassword-protected displayIntegrated SD memory card for recording, downloadingdata and parameter settingsVoice synthesizer

ApplicationsQQQQSemiconductor industryAnalytics and researchIndustrial applicationsPower plants15

Portable – ASM 310Combining light weight and superior 50Dimensions in mm16

Technical dataPortable leak detectorASM 310Test methodsVacuum and sniffing leak detectionMinimum detectable leak rate for Helium (vacuum leak detection)5 · 10-13 Pa m3/sMinimum detectable leak rate for Helium (sniffer leak detection)1 · 10-8 Pa m3/sDetectable gases4Maximum inlet test pressure15 hPaBacking pump capacity1.7 m3/hPumping speed for He1.1 l/sFlange (in)DN 25 ISO-KFStart-up time (20 C) 2 min ( 3.5 min with auto-calibration)He, 3He, H2Response time (sniffing leak detection) 1sInterfaceRS-232I/O interfacesDimensionsDigital inputAnalog output (mantissa)Analog output (exponent)350 x 245 x 414 mmWeight21 kgUniversal voltage100-240 V AC – 50/60 HzMaximum power consumption300 WHelium background suppressionAutomatic / manualOperator interfaceColor touchscreenSelectable languagesEnglish, French, German, Italian, Spanish, Russian, Japanese,Chinese, KoreanOrder numbersOrder numberASM 310BSAA0000MM9AOrder numbers accessoriesAccessoriesOrder numberCountry-specific power cablesUKItaly104411104758Switzerland103718Remote control RC 500 WL, wirelessPT 445 432-TTrolley114820Transportation case119594General accessories for our leak detectors will be found on page 46.17

Multipurpose – ASM 340, ASM 340 DBest in class leak detector for high reliability testingin various applications, using helium and hydrogenTop performance*5 · 10-10 Pa m3/s min.detectable leak rate for HeThe new ASM 340 guarantees top performance in vacuumor sniffing leak detection for various applications – frommaintenance to applications in small production environments. This dependable leak detector can be used both forqualitative localization of leaks as well as quantitative globalor local testing.The ASM 340 is characterized by its powerful pumpingsystem and available in conventional or dry versions. It is theonly leak detector in its class that offers qualitative leakdetection starting at 100 hPa before reaching the inlet testpressure. With an unrivaled performance in sniffing* theASM 340 is the ideal partner for leak localization on pressurized parts. Easy operation, ultra fast response time and shortrecovery time are among the outstanding features of thiscompact multi-purpose unit. The ASM 340 is the perfectsolution for everyday testing even in severe test conditions.Easy operationThe detachable control panel with magnets enhances ergonomics for leak detection on medium or large size parts. Itslarge, bright and color touch screen makes for a maximumreadability of the leak test results. The menu is customizableand can be protected to avoid any unintended operation.Thanks to the wireless remote control, the ASM 340 caneasily be operated from a distance of up to 100 m.More convenienceQQQCustomer benefitsQQQQQQQQQQ18Sniffer probe with LEDs that indicate the measurementstatus and with a zero button facilitates sniffing leakdetectionBypass option for the installation of an auxiliaryprimary pumpComplete range of I/O and Profibus communicationavailable for an easy integration into production linesFastest time to test in its classHigh backing pump capacity for versatile useRapid response time due to high helium pumping speedUnique capability to detect leaks starting at 100 hPaImpressive results in sniffing test mode, with 5·10-10 Pam3/s minimum detectable leak rate for heliumLow maintenance due to rugged designIntegrated SD memory card for recording, downloadingand parameter settingDetachable color control panel for enhanced ergonomicsUser friendly and customizable interfaceComplete range of I/O and Profibus communicationavailableRemote control RC 500 WLSniffer probe LP 505

ApplicationsQQQQQQQR&DAeronautic industryMechanical engineeringMeasuring technologyRefrigerationAir conditioningSemiconductor technologie19

Multipurpose – ASM 340, ASM 340 DBest in class leak detector for high reliability testingin various applications, using helium and hydrogen253139393Dimensions547350Dimensions in mm375Technical dataGeneral technical dataFlange (in)Test methodDetectable gasesMinimum detectable leak rate for He (Sniffing leak detection)Minimum detectable leak rate for He (Vacuum leak detection)Pumping speed for HeMaximum inlet test pressureStart-up time (20 C) without calibrationI/O interfacesInterfaces (see ordering matrix)Selectable languagesDimensionsASM 340DN 25 ISO-KFVacuum and sniffing leak detection4He, 3He, H25 · 10-10 Pa m3/s 5 · 10-13 Pa m3/s2.5 l/s25 hPa 3 minDigital and Analog I/O, RelaysRS-232, Ethernet, Bluetooth English, French, German, Italian, Spanish, Russian, Japanese, Chinese, Korean393 x 547 x 375 mmSpecific technical dataBacking pumpBacking pump capacityPower consumption max.SupplyASM 340Rotary vane pump15 m3/h850 W90 – 130 V, 50/60 Hz200 – 240 V, 50/60 Hz56 kg0 – 45 C (vacuum)0 – 40 C (sniffing)WeightOperating temperature20ASM 340 D (dry version)Diaphragm pump3.4 m3/h600 WUniversal 90 – 240 V, 50/60 Hz45 kg0 – 35 C

Order numberOrder number matrixASM 340JSVA00AaMb9cInterface boardaBasic 15 pins I/O0Basic 15 pins I/O Bluetooth137 pins I/O237 pins I/O Wifi337 pins I/O Ethernet437 pins I/O Bluetooth5Supplyb90/130 V, 50/60 Hz, US Power cableL200/240 V, 50/60 Hz, EU Power cableHPersonalizationcNoAJapanBOrder numberOrder number matrixASM 340 D (dry version)KSBA00AaMM9AInterface boardaBasic 15 pins I/O0Basic 15 pins I/O Bluetooth137 pins I/O237 pins I/O Wifi337 pins I/O Ethernet437 pins I/O Bluetooth5Order number accessoriesAccessoriesCountry-specific power cableUKItalySwitzerlandRemote control, 5 m cableLeak rate in mbar l/s, front panel in EnglishLeak rate in Pa m3/s, front panel in EnglishWireless remote controlRC 500 WLSpray GunStandard“Elite kit” spray gun with accessories in a compact caseStandard sniffer probe5 m hose length, rigid 9 cm nozzleOther nozzle and hose lengths upon requestSniffer probe with Pass/Fail indication and Zero buttonLP 505, 5 m hose length, rigid 12 cm nozzleBypass optionWith European cableWith US cableInterfaceProfibusOrder number104411104758103718106688108880PT 445 432-T112535109951SNC1E1T1BG 449 208-TPT 445 411-TPT 445 413-T12164521

High performance – mobile – ASM 380Mobile unit optimized for rapid pump down and short response timeson large test objectsDry and mobileOur ASM 380 high-performance leak detector is fitted with anACP 40 dry backing pump combined with a powerful highvacuum pump. This makes it the ideal solution for maximumtesting sensitivity in analytics, laboratories, the solar panelindustry and the semiconductor industry. With the ASM 380,you will achieve extremely short pump down times evenwhere large volumes are involved. The slim design and compact size set up on a cart with large wheels and a low centerof gravity make this leak detector mobile and safe to handle.User-friendlyThe ASM 380 with its low noise level of only 53 dB(A) isunusually quiet for a leak detector of its class. The colordisplay panel with full 360 viewing can be read from anyposition. This display is also detachable and can be positionedwhere the user can see it using four strong magnetic clips.An integrated SD memory card makes it easy to save testdata and settings parameters. The ASM 380 has an integratedlockable toolbox for tools, spare parts and accessories.Thanks to its convenient bottle-holder, it is even possible toattach and carry a tracer gas bottle.CompatibleThe ASM 380 can likewise be operated with the wirelessremote control RC 500 WL. This enables the leak detector tobe operated even from a distance of up to 100 meters.Remote control RC 500 WLCustomer benefitsQQQQQQQ22Performance, sensitivity and mobility in thesmallest footprintHigh roughing capacity using particle-free,clean and dry ACP pumpClean room compatibleHigh crossover pressure for a rapid entryinto measurement modeOptional communication interfacesColor touch display with full 360 viewingoffering excellent visibilityIntuitive menus, easy to navigate andavailable in 9 languages

ApplicationsQQQQQQQQQQSemiconductor industryLarge area coatingSolar industryAcceleratorsVacuum components – feedthroughs, valves, bellows,expansion jointsLaser technologyUltra-pure media supplyElectronicsAeronauticsMedical technology23

TurbopumpenHighperformanceHiPace– mobile1200-2300– ASM 380Mobile unit optimized for rapid pump down and short response timeson large test ions in mmTechnical dataASM 380Flange (in)DN 40 ISO-KFTest methodsVacuum and sniffing leak detectionDetectable gases4Minimum detectable leak rate for He (sniffer leak detection)1 · 10-8 Pa m3/sMinimum detectable leak rate for He (vacuum leak detection)Pumping speed for He5 · 10-13 Pa m3/s7 l/sNoise level53 dB (A)Supply90-240 V, 50/60 HzPower consumption max.1,500 WPower consumption700 WMaximum inlet test pressure15 hPaStart-up time (20 C) without calibration 4.5 minStart-up time (20 C) with calibration 5.5 minI/O interfacesInterface1 digital input (start cycle); 2 digital outputs (test mode ok, heliumsignal above reject setpoint); 3 analog outputs (configurable: heliumsignal log, mantissa, exponent, inlet pressure)RS-232Operating temperature10 – 35 CBacking capacity35 m3/hWeight110 kg24He, 3He, H2

Order numbersLeak detectorOrder numberASM 380, universal voltagePSGB00E0MM9AASM 380, universal voltage, integrated BluetoothPSGB00E1MM9AASM 380, universal voltage, configurable I/O interface boardPSGB00E2MM9AASM 380, universal voltage, configurable I/O interface board,integrated WiFiASM 380, universal voltage, configurable I/O interface board,EthernetASM 380, universal voltage, configurable I/O interface board,integrated BluetoothPSGB00E3MM9APSGB00E4MM9APSGB00E5MM9AOrder numbers for accessoriesAccessoriesOrder numberBottle holder for ASM 380118444Inlet filter, 20 μm, DN 40/40 ISO-KF105842Remote control RC 500 WL, wirelessPT 445 432-TLocking clamp, DN 40 KF118801General accessories for our leak detectors can be found on page 46.25

High performance – compact –ASM 182 T, ASM 182 TD The perfect combination of high performance and easy operation for exceptionalreliability even in the harshest industrial applicationsPowerful and compactThe high-performance compact leak detectors in the ASM 182series are available in oil-lubricated and dry versions. Bothversions are equipped with a high backing capacity. The smallfootprint of these leak detectors enables them to be easilyintegrated into production lines. Our powerful ASM 182 canbe used for leak detection on parts of the most varied sizesand volumes. Whether used in the automotive industry,the aeronautic and aerospace or in semiconductor industry,these leak detectors fit the bill with their high performanceand reliability.Oil sealed version:ASM 182 TThe ASM 182 T leak detector has a rotary vane pump with apumping speed of 20 m³/h. This large backing pump, togetherwith a powerful turbopump, enables the ASM 182 T toachieve the fastest cycle times in its performance class. It ismost impressive when large volume test parts are concerned.Thanks to the fold-out feature of the front and rear housingsides the leak detector can be easily maintained.Dry version:ASM 182 TD The dry ASM 182 TD combines a particle-free backing pumpwith extreme long-term stability and a pumping speed of25 m³/h, a molecular pump for high pumping speed in themedium vacuum range and a powerful turbopump in the highvacuum range. The test pressure is attained rapidly for largeor strongly outgassing test objects. Thanks to ultra-smallbackground signals, the ASM 182 TD reliably performseven in the most demanding environments.Customer benefitsASM 182 TQQQCustomer benefitsASM 182 TD QQQ26Short cycle times thanks to the largest backing pumpin its classLow helium background even with large test partsEase of service thanks to the fold-out feature of the frontand rear housing sidesThe test pressure is attained rapidly even with large orstrongly outgassing test specimens.Short response times and test cycles thanks to high heliumpumping speedUltra-small background signals and rapid signal reduction –for demanding test jobs

ApplicationsQQQQQQQQQLeak detection systemsAutomotiveAeronautic and aerospace industryLaser technologySensorsVacuum components – feedthroughs, valves, bellows,expansion jointsElectronicsMechanical components for petrochemistryHeat exchangers27

High performance – compact –ASM 182 T, ASM 182 TD The perfect combination of high performance and easy operation for exceptionalreliability even in the harshest industrial applicationsDimensions170Flange(in)DNDN4040Inlet Port421421444227408434594Dimensions in mmTechnische DatenTechnical dataASM 182 TASM 182 TD Flange (in)DN 40 ISO-KFDN 40 ISO-KFTest methodsVacuum and sniffingleak detectionVacuum and sniffingleak detectionDetectable gases44Minimum detectable leak rate for He (sniffing leak detection)HeHe-831 · 10 Pa m /s-1331 · 10-8 Pa m3/s5 · 10-13 Pa m3/sMinimum detectable leak rate for He (vacuum leak detection)5 · 10Pumping speed for He4.4 l/sNoise level54 dB (A)65 dB (A)Supply200-240 V, 50/60 Hz200-240 V, 50/60 HzPower consumption max.1,000 W1,500 WMaximum inlet test pressure6 hPa6 hPaBacking pump capacity20 m3/h25 m3/hStart-up time (20 C) without calibration3 min 4 minStart-up time (20 C) with calibration4.5 min 5.5 minI/O interfacesDigital inputs (start, vent, calibration ); digital outputs (testmodes, cycle in progress, heliumsignal above reject setpoint );analog outputs (helium signal log,inlet pressure)Digital inputs (start, vent, calibration ); digital outputs (testmodes, cycle in progress, heliumsignal above reject setpoint );analog outputs (helium signal log,inlet pressure)InterfaceRS-232RS-232Pa m /s4.4 l/sOperating temperature10 – 40 C10 – 40 CWeight76 kg88 kg28

Order number matrixASM 182 T, ASM 182 TD Order numberabcdefghij0Leak detectoraASM 182 TD2ASM 182 TD without cartD1ASM 182 TD with cartD7Detectable gasesbHelium (4He)03 masses (4He, 3He, H2)3Seals for vacuum module and analyzer cellcElastomerRMetalMControl paneldStandardSGraphic color touchscreenTDetection optioneASM 182 T (not available)0ASM 182 TD standardSASM 182 TD with gas line optionGTest chamberfnone0Small test chamber1Medium test chamber2Large test chamber3CovergPaintedPUCT (stainless steel covers and frame),only for ASM 182 TD 0-130 V, 50/60 Hz7220-240 V, 50/60 Hz8Cable and plug ut plug729

High performance – console –ASM 192 T, ASM 192 T2D Ultra sensitive detection limits on large-volume components,ergonomic unit for standing operatorsErgonomic console unitsHigh-performance leak detectors in the ASM 192 series provide high reliability even in harsh environments. The consoleunits have an ergonomic work height for standing operatorsand the remote control is fixed to a freely movable arm forconvenient use. The work surface is grounded and provides agenerous surface for placing large test objects, useful smallparts and installation components on.These console leak detectors benefit of an easy menunavigation combined with a voice synthesizer for deliveringimportant status messages. As with all other Pfeiffer

The MiniTest 300 helium leak detector weighs only a sensa-tional 5 kg. It is the first vacuum leak detector on the market which works with a quartz window sensor. Thanks to this innovative technology, it stands out due to its lightweight design and small footprint. The uncomplicated operation and an optional wireless remote