specifications8753ET/ESNetwork AnalyzersHP 8753ET and 8753ES Network Analyzers30 kHz to 3 or 6 GHzThis document describes the performanceand features of the following products:HP 8753ES S-parameter vector network analyzerHP 8753ES Option 011 vector network analyzer without test setHP 8753ET Transmission/reflection vector network analyzerFor more information about these analyzers, please see the following documents:HP 8753ET and 8753ES Network Analyzers BrochureHP 8753ET and 8753ES Network Analyzers Configuration GuideHP literature number5968-5159E5968-5158E

Table of contentsDefinitions and test conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3System performance summaries . . . . . . . . . . . . . . . . . . . . . . . . . . . .4Specifications and characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . .10Measurement throughput summary . . . . . . . . . . . . . . . . . . . . . . . . .21Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21Test set specifications for Option 011 . . . . . . . . . . . . . . . . . . . . . . . .22Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .242

Definitions and test conditionsSpecifications describe the instrument's warrantedperformance after a half-hour warm-up and over thetemperature range of 25 C 5 C, unless otherwisestated. Specifications for frequencies above 3 GHz donot apply to instruments with Option 075 (75-ohmimpedance).Supplemental characteristics are typical but non-warranted performance parameters. These are denoted as“typical,” “nominal,” or “approximate.”The measurement uncertainty curves and measurement port characteristics given for HP 8753ESsystems also apply to the HP 8753ES with Options 006and 011 and the HP 85047A test set (50-ohm), or theHP 8753ES Option 011 with an HP 85046B test set(75-ohm).Dynamic rangeSystem dynamic range is calculated as the differencebetween the receiver noise floor and the lesser ofeither the source maximum output or the receivermaximum input level. System dynamic range applies totransmission measurements only, since reflection measurements are limited by directivity.Noise floor is specified as the mean of the noise traceover frequency. Noise floor is measured with the testports terminated in loads, full two-port error correction for the HP 8753ES and enhanced-response errorcorrection for the HP 8753ET (with 16 averages usedduring isolation), 10 Hz IF bandwidth (BW), maximumtest port power, and no averaging during the measurement.Measurement uncertaintyMeasurement uncertainty curves utilize a Root SumSquare (RSS) model for the contribution of randomerrors such as noise, typical connector repeatabilities,and test set switching; this is combined with a worstcase model for the contributions of dynamic accuracyand residual systematic errors.Curves show the worst-case magnitude and phaseuncertainty for reflection and transmission measurements, after a full two-port error correction for theHP 8753ES and enhanced-response error correctionfor the HP 8753ET (including isolation with an averaging factor of 16) using the specified cal kit, with 10 HzIF bandwidth (BW) and no averaging.Measurement port characteristicsCharacteristics show the residual system uncertaintiesfor uncorrected performance and after accuracyenhancement using full two-port error correction forthe HP 8753ES and enhanced-response error correction for the HP 8753ET. These characteristics applyfor an environmental temperature of 25 5 C, withless than 1 C deviation from the calibration temperature. HP 8753ET Option 004 may degrade transmissionsource match as much as 2 dB, resulting in up to 0.05dB additional uncertainty in transmission tracking.Corrected performance indicates residual error aftercalibration. It is determined by the quality of calibration standards, system repeatability, stability, andnoise.Uncorrected performance indicates intrinsic errorswithout calibration correction applied. This is relatedto the ultimate stability of a calibration.3

System performance summariesMeasurement port characteristicsHP 8753ES (50-ohm systems)7-mm test portsFrequency rangeCorrected 30 kHz-300 kHz2 300 kHz-1.3 GHz 1.3 GHz-3 GHzDirectivity55 dB55 dB51 dBSource match55 dB51 dB49 dBLoad match55 dB55 dB51 dBReflection (0.001 dB (0.001 dB (0.005 dBtracking 0.02 dB/ C) 0.01 dB/ C) 0.02 dB/ C)Transmission (0.008 dB (0.006 dB (0.009 dBtracking 0.02 dB/ C) 0.01 dB/ C) 0.02 dB/ C)UncorrectedDirectivity20 dB*35 dB30 dBSource match18 dB**16 dB16 dBLoad match18 dB**18 dB16 dBReflection 2.5 dB 1 dB 1 dBtrackingTransmission 2.5 dB 1 dB 1 dBtrackingCrosstalk90 dB***100 dB100 dB*15 dB, 30 kHz to 50 kHz**10 dB, 30 kHz to 50 kHz***60dB, 30 kHz to 50 kHzConfigurationNetwork analyzerHP 8753ESStandard and Option 006HP 85031BHP 11857DCalibration kitTest-port cablesSystem dynamic rangedB130 kHz to 50 kHz50 kHz to 300 kHz300 kHz to 16 MHz16 MHz to 3 GHzOption 0143 GHz to 6 GHzOption 0147090 dB1100 dB110 dB108 dB105 dB103 dBReflection uncertaintyMagnitudePhaseTransmission uncertaintyMagnitude1.2.PhaseTypical below 300 kHz.Typical performance.43 GHz- 6 GHz46 dB43 dB46 dB (0.020 dB 0.03 dB/ C) (0.021 dB 0.03 dB/ C)25 dB14 dB14 dB 1.5 dB 1.5 dB90 dB

System performance summariesMeasurement port characteristicsHP 8753ES (50-ohm systems)Type-N test portsFrequency rangeCorrected30 kHz-300 kHz2 300 kHz-1.3 GHz 1.3 GHz-3 GHzDirectivity50 dB50 dB47 dBSource match 49 dB42 dB36 dBLoad match50 dB50 dB47 dBReflection (0.005 dB (0.009 dB (0.019 dBtracking 0.02 dB/ C) 0.01 dB/ C) 0.02 dB/ C)Transmission (0.014 dB (0.013 dB (0.026 dBtracking 0.02 dB/ C) 0.01 dB/ C) 0.02 dB/ C)ConfigurationNetwork analyzerHP 8753ESStandard and Option 006HP 85032BHP 11857DCalibration kitTest-port cablesSystem dynamic range30 kHz to 50 kHz50 kHz to 300 kHz300 kHz to 16 MHz16 MHz to 3 GHzOption 0143 GHz to 6 GHzOption 01470 dB190 dB1100 dB110 dB108 dB105 dB103 dBReflection uncertaintyMagnitudePhaseTransmission uncertaintyMagnitudePhase1. Typical below 300 kHz.2. Typical performance.53 GHz-6 GHz40 dB31 dB40 dB (0.070 dB 0.03 dB/ C) (0.065 dB 0.03 dB/ C)

System performance summariesHP 8753ES (50-ohm systems)3.5-mm test portsConfigurationNetwork analyzerHP 8753ESStandard and Option 006HP 85033DHP 11857DCalibration kitTest-port cablesSystem dynamic rangeMeasurement port characteristicsFrequency rangeCorrected 30 kHz-300 kHz2 300 kHz-1.3 GHz 1.3 GHz-3 GHz 3 GHz-6 GHzDirectivity49 dB46 dB44 dB38 dBSource match 49 dB44 dB41 dB37 dBLoad match49 dB46 dB44 dB38 dBReflection (0.010 dB (0.005 dB (0.007 dB (0.009 dBtracking 0.02/ C) 0.01/ C) 0.02/ C) 0.03/ C)Transmission (0.016 dB (0.014 dB (0.022 dB (0.048 dBtracking 0.02/ C) 0.01/ C) 0.02/ C) 0.03/ C)70 dB190 dB1100 dB110 dB108 dB105 dB103 dB30 kHz to 50 kHz50 kHz to 300 kHz300 kHz to 16 MHz16 MHz to 3 GHzOption 0143 GHz to 6 GHzOption 014Reflection uncertaintyMagnitudePhaseTransmission uncertaintyMagnitudePhase1. Typical below 300 kHz.2. Typical performance.6

System performance summariesMeasurement port characteristicsHP 8753ES (75-ohm systems)Type-N test portsFrequency rangeCorrected30 kHz - 300 kHz2300 kHz - 1.3 GHz 1.3 GHz - 3 GHzDirectivity48 dB48 dB43 dBSource match47 dB41 dB35 dBLoad match48 dB48 dB43 dBReflection (0.004 dB (0.010 dB (0.019 dBtracking 0.02 dB/ C) 0.01 dB/ C) 0.02 dB/ C)Transmission 0.018 dB (0.016 dB (0.033 dBtracking 0.02 dB/ C) 0.01 dB/ C) 0.02 dB/ C)Uncorrected2 30 kHz - 300 kHz 300 kHz - 1.3 GHz 1.3 GHz - 3 GHzDirectivity20 dB*35 dB30 dBSource match16 dB**16 dB16 dBLoad match15 dB**18 dB16 dBReflection 2.5 dB 1 dB 1 dBtrackingTransmission 2.5 dB 1 dB 1 dBtrackingCrosstalk90 dB***100 dB100 dBConfigurationNetwork analyzerCalibration kitTest-port cablesHP 8753ES Option 075HP 85036BHP 11857BSystem dynamic range68 dB190 dB196 dB106 dB30 kHz to 50 kHz50 kHz to 300 kHz300 kHz to 16 MHz16 MHz to 3 GHz* 15 dB, 30 to 50 kHzReflection uncertaintyMagnitude**10 dB, 30 to 50 kHz*** 60 dB, 30 to 50 kHzPhaseTransmission uncertaintyMagnitudePhase1. Typical below 300 kHz.2. Typical performance.7

System performance summariesHP 8753ES (75-ohm systems)Type-F test portsConfigurationNetwork analyzerCalibration kitTest-port cablesHP 8753E Option 075HP 85039BHP 11857BSystem dynamic range68 dB190 dB196 dB106 dB30 kHz to 50 kHz50 kHz to 300 kHz300 kHz to 16 MHz16 MHz to 3 GHzMeasurement port characteristicsData is shown for a Type-F female reflection port and aType-F male transmission port.Corrected2DirectivitySource matchLoad cy range30 kHz-300 kHz300 kHz-1.3 GHz 1.3 GHz-3 GHz45 dB45 dB40 dB40 dB40 dB30 dB45 dB45 dB40 dB (0.060 dB (0.060 dB (0.024 dB 0.02 dB/ C) 0.01 dB/ C) 0.02 dB/ C) (0.033 dB (0.019 dB (0.057 dB 0.02 dB/ C) 0.01 dB/ C) 0.02 dB/ C)Reflection uncertaintyMagnitudePhaseTransmission uncertaintyMagnitude1. Typical below 300 kHz.2. Typical performance.Phase8

System performance summariesHP 8753ET (50-ohm systems)Type-N test portsConfigurationNetwork analyzerCalibration kitTest port cableHP 8753ET Standard,Option 006 or Option 004HP 85032BHP part number 8120-4781System dynamic range300 kHz to 16 MHz16 MHz to 3 GHz3 GHz to 6 GHz100 dB110 dB105 dBMeasurement port characteristicsFrequency range (continued)Corrected300 kHz-1.3 GHz 1.3 GHz-3.0 GHz 3.0 GHz-6.0 GHzReflection measurements1Directivity50 dB47 dB40 dBSource match42 dB36 dB31 dBLoad matchOne-port cal24 dB19 dB16 dBEnhanced reflection cal 24 dB19 dB16 dBReflection tracking (0.009 dB (0.019 dB (0.07 dB 0.01 dB/ C) 0.02 dB/ C) 0.03 dB/ C)Transmission measurements2Source matchEnhanced-response cal 42 dB36 dB31 dBResponse-only cal325 dB20 dB14 dBTransmission trackingEnhanced-response cal (0.006 dB (0.018 dB (0.054 dB 0.01 dB/ C) 0.02 dB/ C) 0.03 dB/ C)Response-only cal (0.033 dB (0.1 dB (0.27 dB 0.01 dB/ C) 0.02 dB/ C) 0.03 dB/ C)Frequency rangeUncorrected300 kHz-1.3 GHz 1.3 GHz-3.0 GHz 3.0 GHz-6.0 GHzDirectivity30 dB24 dB19 dBSource match425 dB20 dB14 dBLoad match24 dB19 dB16 dBReflection tracking 1.0 dB 1.0 dB 2.0 dBTransmission tracking 1.5 dB 1.5 dB 2.5 dBCrosstalk100 dB100 dB90 dBTransmission uncertainty:(enhanced-response calibration)Transmission uncertainty:(response calibration)Reflection uncertainty(one-port calibration)1. One-port or enhanced-response calibration.2. Enhanced-response or response-only calibration.3. 2 dB less with Option 004.4. Option 004 may degrade uncorrected source match as much as 2 dB.9

HP 8753ES specificationsTest port input characteristics30 kHz to 3 GHzOption 00630 kHz to 6 GHzFrequency rangeTest-port output characteristics7Frequency characteristicsRange30 kHz to 3 GHzOption 006ResolutionStability9AccuracyOption 1D5Power range2Option 075/01430 kHz to 6 GHz1 Hz 7.5 ppm (0 to 55 C) 3 ppm/year 10 ppm at 25 C 5 C 0.05 ppm (0 to 55 C), 0.5 ppm/year-85 to 10 dBm-85 to 8 dBmAverage noise level2,8 3 GHz 3 GHz 3 GHz3 to 6 GHz3 to 6 GHz3 to 6 GHzMaximum input levelDamage levelImpedanceOption 075 -82 dBm (3 kHz BW) -102 dBm (10 Hz BW) -110 dBm (10 Hz BW) typical -77 dBm (3 kHz BW) -97 dBm (10 Hz BW) -105 dBm (10 Hz BW) typical10 dBm26 dBm or 35 VDC50 Ω (nominal)75 Ω (nominal)Frequency response2,525 dB (typically 31 dB)0.01 dB 1.0 dB-15 dBm to 5 dBm 0.2 dB5 dBm to 10 dBm6 0.5 dB9Impedance50 Ω (nominal)30 kHz to 3 GHz 16 dB RL ( 1.38 SWR)3 GHz to 6 GHz 14 dB RL ( 1.50 SWR)Option 07575 Ω (nominal)30 kHz to 3 GHz 16 dB RL ( 1.38 SWR)Power sweep rangeResolutionLevel accuracy1,2,5Level linearity1,2,5Spectral purity2nd harmonic3Option 0023rd harmonic4Option 002 -40 dBc at 0 dBm9 -50 dBc at -10 dBm9 -25 dBc at 10 dBm6 -40 dBc at 0 dBm9 -50 dBc at -10 dBm9 -25 dBc at 10 dBm6 -55 dBc at -10 dBm 1.0 dB 2.0 dBInternally generated harmonics (Option 002)2nd harmonic3 -15 dBc at 8 dBm3rd harmonic4 -30 dBc at 0 dBm9 -45 dBc at -15 dBm9 -30 dBc at 8 dBm -50 dBc at 0 dBm9 -50 dBc at -15 dBm9Harmonic measurement accuracy10300 kHz to 3 GHz3 GHz to 6 GHzNonharmonic spurious9Mixer related -30 dBc at 10 dBm61.2.3.4.5.6.7.8.9.10.11.300 kHz to 3 GHz3 GHz to 6 GHz 1.5 dB 3 dB (Option 006)Harmonic measurement dynamic range-40 dBc (output -10 dBm, input -15 dBm)9Frequency offset mode11Frequency range300 kHz to 3 GHzOption 006300 kHz to 6 GHzR-channel input requirementsPower level300 kHz to 3 GHz 0 to -35 dBm3 GHz to 6 GHz 0 to -30 dBmLO spectral purity (typical)Maximum -25 dBcspurious inputResidual FM 20 kHzLO frequency accuracy9 1 MHz at nominal frequencyRelative to 0 dBm output power.Typical below 300 kHz.16 MHz to 3 GHz.16 MHz to 2 GHz.Typical from 2 to 3 GHz for instruments with Option 075. 8 dBm maximum with Option 075, or Option 014.Test performed on port 1 only.Instruments with Option 075 are degraded 2 dB.Typical performance.25 5 C.The HP 8753ES source characteristics and measurement accuracy in this mode are dependent on the stability of the externalLO source. The RF source tracks the LO to maintain a stable IF signal at the R-channel receiver input. Degradation in accuracyis negligible when using an HP 8642A/B, 8656B, or E4432B RF signal generator as the LO source.10

Test port input characteristics (continued)External source mode1 (CW time sweep only)Frequency range300 kHz to 6 GHzR-channel input requirements (typical)Power level0 to -25 dBmSpectral purityMaximum -30 dBcspurious inputResidual FM 20 kHzTypical settling500 ms (automatic)timeReference levelRangeResolutionStability230 kHz to 3 GHz3 GHz to 6 GHz 500 dB0.001 dB0.02 dB/ C0.04 dB/ CPhase characteristicsDynamic accuracyRange(10 Hz IF BW) 180 (300 kHz to 3GHz)50 ms (manual)0.1% (automatic)Frequency readoutaccuracyInput frequency accuracy requirement2Manual-0.5 to 5 MHzAccuracy(see magnitude and phasecharacteristics)Magnitude characteristicsDynamic accuracy(10 Hz IF BW)(300 kHz to 3GHz)(3 to 6GHz)(3 to 6GHz)Display resolutionMarker resolution3Trace noise0.01 / division0.01 ( 5 dBm at test-port, ratio measurement, 3 kHz BW)300 kHz to 3 GHz 0.038 rms3 GHz to 6 GHz 0.070 rmsDisplay resolution 0.001 dB/divisionMarker resolution3 0.001 dBTrace noise( 5 dBm at test-port, ratio measurement, 3 kHz BW)300 kHz to 3 GHz 0.006 dB rms3 GHz to 6 GHz 0.010 dB rmsReference levelRangeResolutionStability30 kHz to 3 GHz3 GHz to 6 GHzPolar characteristicsRangeReference-180 to 180 0.01 0.05 / C0.20 / C10 x 10-12 to 1000 unitsfull scale 500 units1. See the HP 8753ES descriptions and options for a functional description. Measurement accuracy is dependenton the stability of the input signal.2. Typical performance.3. Marker resolution for magnitude; phase and delay is dependent upon measured value. Resolution is limited tofive digits.11

HP 8753ES Option 011 specificationsTest port input characteristics Option 011300 kHz to 3 GHzOption 00630 kHz to 6 GHzFrequency rangeTest port output characteristicsFrequency characteristicsRange300 kHz to 3 GHzOption 00630 kHz to 6 GHzResolution1 HzStabilitytypically 7.5 ppm 0 to 55 Ctypically 3 ppm/yearOption 1D5 typically 0.05 ppm 0 to 55 Ctypically 0.5 ppm/yearAccuracy 10 ppm at 25 C 5 CPower range-5 to 20 dBmOption 006-5 to 18 dBmResolution0.01 dBLevel accuracy1,2 1.0 dBLevel linearity1,2,5 0.25 dB, -5 to 15 dBm 0.5 dB, 15 to 20 dBmImpedance50 Ω nominal300 kHz to 3 GHz2 16 dB RL ( 1.38 SWR)300 kHz to 6 GHz 14 dB RL ( 1.50 SWR)Spectral purity2nd harmonic3Option 0023rd harmonic4Option 002dBm6 -40 dBc at 10 -50 dBc at 0 dBm6 -25 dBc at max power -40 dBc at 10 dBm6 -50 dBc at 0 dBm6 -25 dBc at max powerNonharmonic spurious6Mixer related -30 dBc at max power -55 dBc at -10 dBmAverage noise level250 kHz to 3 GHz50 kHz to 3 GHz50 kHz to 3 GHz3 to 6 GHz3 to 6 GHz3 to 6 GHz -90 dBm (3 kHz BW) -110 dBm (10 Hz BW) -120 dBm (10 Hz BW) typical -85 dBm (3 kHz BW) -105 dBm (10 Hz BW) -115 dBm (10 Hz BW) typicalMaximum input level0 dBmDamage level20 dBm or 25 VDCImpedance:50 ohms nominal300 kHz to 2 MHz 20 dB RL2 MHz to 1.3 GHz 24 dB RL1.3 GHz to 3 GHz 19 dB RLOption 0063 GHz to 6 GHz 15 dB RL6Frequency response300 kHz to 3 GHz3 GHz to 6 GHz 1.0 dB, 2.0 dBHarmonics (Option 002)2nd harmonic3 -15 dBc at 0 dBm3rd harmonic4 -30 dBc at -10 dBm6 -45 dBc at -30 dBm6 -30 dBc at 0 dBm -50 dBc at -10 dBm6 -50 dBc at -30 dBm6Harmonic measurement accuracy16 MHz to 3 GHz3 GHz to 6 GHz Option 006 1.5 dB 3 dBHarmonic measurement dynamic range-40 dBc6(output -10 dBm,input -15 dBm)1.2.3.4.5. 10 dBm output power for the HP 8753ES Option 011.Typical below 300 kHz.16 MHz to 3 GHz.16 MHz to 2 GHz.For HP 8753ES Option 011 and Option 006, linearity is specified for the ranges of -5 to 13 dBmand 13 to 18 dBm.6. Typical performance.12

HP 8753ES Option 011 specificationsTest port input characteristics (continued)Frequency offset mode3Frequency rangeOption 006300 kHz to 3 GHz300 kHz to 6 GHzR-channel input requirementsPower level300 kHz to 3 GHz3 GHz to 6 GHz0 to -35 dBm0 to -30 dBmLO spectral purityMaximum spurious -25 dBcinputResidual FM 20 kHzLO frequency accuracy 1 MHz at nominalfrequency(CW time sweep only)300 kHz to 6 GHzExternal source mode4Frequency rangeR-channel input requirements1Power level0 to -25 dBmSpectral purityMaximum spurious -30 dBcinputResidual FM 20 kHzTypical settling time1 500 ms (automatic)Reference levelRangeResolutionStability230 kHz to 3 GHz3 GHz to 6 GHzRangeDisplay resolutionMarker resolution5Trace noise230 kHz to 3 GHz3 GHz to 6 GHzReference levelRangeResolutionStability30 kHz to 3 GHz3 GHz to 6 GHzPolar characteristicsRangeReference 500 dB0.001 dB0.02 dB/ C0.04 dB/ C 180 0.01 / division0.01 0.038 rms 0.070 rms( 5 dBm at test port, ratiomeasurement, 3 kHz BW)-180 to 180 0.01 0.05 / C0.20 / C10 x 10-12 to 1000 units fullscale 500 units50 ms (manual)Input frequency accuracy requirement1Manual:-0.5 to 5 MHzDisplay resolution0.001 dB/divisionMarker resolution50.001 dBTrace noise( 5 dBm at test-port, ratio measurement, 3 kHz BW)Magnitude300 kHz to 3 GHz3 GHz to 6 GHz 0.006 dB rms 0.010 dB rmsPhase300 kHz to 3 GHz3 GHz to 6 GHz 0.038 rms 0.070 rms1. Typical performance.2. Typical below 300 kHz.3. The HP 8753ES source characteristics and measurement accuracy in this mode are dependent on the stability of theexternal LO source. The RF source tracks the LO to maintain a stable IF signal at the R-channel receiver input.Degradation in accuracy is negligible when using an HP 8642A/B, 8656B, or E4432B RF signal generator as the LO source.4. See the HP 8753ES descriptions and options for a functional description. Measurement accuracy is dependent on thestability of the input signal.5. Marker resolution for magnitude, phase and delay is dependent upon measured value. Resolution is limited to five digits.13

HP 8753ET specificationsTest-port output characteristicsFrequency characteristicsRange300 kHz to 3 GHzOption 006 300 kHz to 6 GHzResolution1 HzStability4 7.5 ppm (0 to 55 C)Option 1D5AccuracyPower range:ResolutionLevel accuracy1Level linearity1Standard 3 ppm/year 0.05 ppm (0 to 55 C) 0.5 ppm/year 10 ppm at 25 C 5 C-20 to 5 dBm-85 to 10 dBm (with Option 004)0.01 dB 1.0 dB-20 to -15 dBm-15 to 0 dBm0 to 5 dBm 0.5 dB 0.2 dB 0.5 dBTest port input characteristicsFrequency range300 kHz to 3 GHzOption 006300 kHz to 6 GHzAverage noise level300 kHz to 3 GHz300 kHz to 3 GHz300 kHz to 3 GHz3 GHz to 6 GHz3 GHz to 6 GHz3 GHz to 6 GHzMaximum input levelDamage levelImpedanceFrequency response5300 kHz to 3 GHz3 GHz to 6 GHzImpedance300 kHz to 3 GHz3 GHz to 6 GHz 0.2 dB 0.5 dB50 Ω (nominal) 18 dB RL ( 1.28 SWR) 14 dB RL ( 1.50 SWR)Spectral purity2nd harmonic2Option 002 1.0 dB 2.0 dB3rd harmonic3 -30 dBc at 0 dBm4 -45 dBc at -15 dBm4 -30 dBc at 8 dBm -50 dBc at 0 dBm4 -50 dBc at -15 dBm4Harmonic measurement accuracy5300 kHz to 3 GHz3 GHz to 6 GHz 1.5 dB 3 dB (Option 006)Harmonic measurement dynamic range4 -40 dBc at 0 dBm4 -50 dBc at -10 dBm4 -25 dBc at maximum outputpower3rd harmonic3Option 002 10 dBm reflection port0 dBm transmission port 26 dBm or 35 VDC5

Network analyzer HP 8753E Option 075 Calibration kit HP 85039B Test-port cables HP 11857B System dynamic range 30 kHz to 50 kHz 68 dB1 50 kHz to 300 kHz 90 dB1 300 kHz to 16 MHz 96 dB 16 MHz to 3 GHz 106 dB Reflection