DataSheet - F00.psgsm

Transcription

DataSheetSHS1000 Series Handheld Digital OscilloscopeSH1102/SHS1062Features & Benefits SHS1000 Series have 2 channels; provide functions as Oscilloscope, Multimeter and Recorder (TrendPlot andwaveform Recorder).SHS1000 Series with patent IsolatedChannel technology provide isolation from ground and isolation betweenchannelsCATII1000V and CATIII600V between two channels references,between channels reference and earth groundCATII600V and CATIII300V between channels reference and Multimeter input referenceCATII300V and CATIII150V input directCATII1000V and CATIII600V input with 10:1 probe5.7 inch TFT color LCD display100MHz Bandwidth, 1GS/s real-time sampling per channel, up to 50GSa/s equivalent sampling rate, 2Mptsmemory depth6000 counts Multimeter, provides measurements of DCV, ACV, DCI, ACI, Resistance, Diode, Capacitance,ContinuitySupport Scope TrendPlot, Meter TrendPlot and Scope RecorderAutomatic and manual trigger modesTrigger types: Edge, Pulse, Video,Slope and Alternative32 automatic waveform measurements,3 cursor measure modesDigital Filter functions:Low pass filter,High pass filter, Band pass filter, Band limit filter

Math functions: , - , ,FFT operationsMultiple Language User InterfaceStandard setup interface: USB Device,USB HostUSB storage update, PC communication and PictBridge print are availableRechargeable battery and battery charger / line power adapter includedApplications Power electronics test, such as Switch mode power supply, Inverter, Converter and Lighting electronics.Wind power, PV power and other new energy equipment testAutomotive electronic, electric vehicles testIndustrial Power systems strong power testElectrical industrial site commissioning and testField testApplications from microelectronic circuits to power electronics, in fields floating measurements or locale sitemeasurements neededEducationIntroductionSHS1000 series are dual isolated channel handhold oscilloscopes with patent IsolatedChannel technology.SHS1000 series integrate functions as Oscilloscope, Multimeter and Recorder.SHS1000 Series provides isolation from ground and isolation between channels allowing you to take floatingmeasurements without worrying about damaging circuitry.100 MHz Bandwidths, 1GS/s real-time sampling per channel, up to 50GSa/s equivalent sampling rate, 2Mptsmemory depth. Support Scope TrendPlot, Meter TrendPlot and Scope Waveform Record, record length up to7Mpts. 5.7 inch TFT color LCD display. Support USB storage and internal memory. Battery included, handholdavailable, convenient for outdoor measurement.Isolated input,make measurements in security

Patent IsolatedChannel technology used in SHS1000 series oscilloscopes, dual channel, and 100MHzbandwidth.CATII300V and CATIII150V maximum BNC input voltage direct, CATII1000V and CATIII600V with standard10:1 probe.CATII1000V and CATIII600V maximum voltage between two channels references. CATII600V and CATIII300V maximum voltage between Multimeter input reference and the ground.High-performance oscilloscope The SHS1000 series channels are isolated from each other; real-time sampling rate is up to 1GSa/s perchannel, equivalent sampling rate up to 50GSa/s 2Mpts memory depth Dynamic and broad input voltage range, 5mV/div 100V/div direct input Math functions: , - , ,and FFT Digital Filter functions:Low pass filter,High pass filter, Band pass filter, Band limit filter 32 types of automatic waveform measurements,3 cursor measure modes Automatic and manual trigger modesTrigger types: Edge, Pulse, Video,Slope and Alternative Support EasyScope software Standard SPCI command collections, support telecommuting Multiple Language User Interface, support Multilingual help system onlineFFT32 types of measurementsLong Memory

TrendPlot Scope TrendPlot records scope measurement data in scan mode, 800K points capacity, more than 24hours recording timeMeter TrendPlot records multimeter measurement data, 1.2M points recording depth, at 0.5GSa/s,recording time 8120 hoursRecording results export available, convenient for father analysisTwo kinds of display mode, ‘ALL’ and ‘NORMAL’; support zoom and cursorSupport recording real timeScope TrendPlotMeter TrendPlotScope Recorder Recording scope waveform continually in scan modeSupport recording, replay and zoom function7M points memory depth,18 hours recording time4GB in USB storage mode, 3000hours recording timeMultimeter 6000 counts high performance MultimeterProviding measurements of DCV, ACV, DCI, ACI, Resistance, Diode, Capacitance, ContinuityDCV measurementACI measurement

100MHzRise Time 5.8ns 3.5nsSingle Channel: 1GSa/s,Single Channel: 1GSa/s,Double Channels: 500MSa/sDouble Channels: 500MSa/sEquivalent Sampling Rate50GSa/s50GSa/sMemory Depth2Mpts2MptsTime Base Range5 ns/ div – 50s/ div2.5ns/ div – 50s/ divScan Range100ms/ div 50s/ div100ms/ div 50s/ divVertical Sensitivity5mV/div – 100V/div(1-2-5 order)5mV/div – 100V/div(1-2-5 order)Vertical Resolution8 bits8 bitsTrigger TypesEdge, Pulse, Video, Slope, AlternativeEdge, Pulse, Video, Slope, AlternativeFrequency Counter6 bits6 bitsConnectionUSB Device, USB HostUSB Device, USB HostMath , -, * , /, FFT , -, * , /, FFTOscilloscope Trend Plot800K pointsReal Time Sampling RateMeterMaximum ResolutionDC VoltageAC VoltageDC CurrentAC CurrentResistanceCapacitance600060mV( 1% 15digit)600mV—1000V( 1% 5digit)60mV( 1% 15digit)600mV—750V( 1% 5digit)60.00mA—600mA( 1% 5digit)6A—10A( 1.5% 5digit)60.00mA—600mA( 1% 5digit)6A—10A( 1.5% 5digit)600Ω—60MΩ( 1% 5digit)40nF( 3% 10digit)400nF--400uF( 4% 5digit)Diode0—2VContinuity 50Ω Buzzer soundsMultimeter Trend Plot1.2M points

Technical SpecificationsOscilloscopeAcquisition SystemSampling TypesReal time, EquivalentSampling ModeSampling, Peak detection, AverageAverage Times4, 16, 32, 64, 128, 256Input SystemInput CouplingAC, DC, GNDInput Impedance1MΩ 2%, 18pf 3pfProbe Attenuation Factor10XProbe Attenuation FactorsSet(V)1X, 5X , 10X, 50X , 100X, 500X , 1000Xchannels from earth ground,Overvoltage CategoryMaximum VoltagebetweenCAT I&CAT II1000VrmsCAT III600VrmsOvervoltage CategoryMaximum VoltageCAT I&CAT II600VrmsCAT III300VrmsOvervoltage CategoryMaximum Voltage1x CAT I&CAT II300Vrms1x CAT III150Vrms10x CAT I&CAT II1000Vrms10x CAT III600VrmsVoltage portDC 1000V, AC 750VCurrent port(mA)AC 250V/10ACurrent port(A)AC putreference and the groundMax. input Voltage for BNCMax.inputVoltageMultimeter input portSingle Channel CommonMode Rejection, typicalChannel-to-ChannelIsolationfor 1000:1 50MHz 35dBHorizontal SystemReal time Sample RateSingle Channel :50Sa/s 1GSa/sDouble Channels: 500MSa/sEquivalent Sample Rate50GSa/sInteraction ModeLine, (Sinx)/xChannel ModeMemory DepthSingle ChannelSample Rate1Gsa/sNormal40kptsDeepnonsupportSingle Channel 500MSa/s20kpts2MptsDouble Channels 500MSa/s20kpts1Mpts

Display ModeMAIN, WINDOW, ZOOM, SCAN, X-YTime Base Accuracy 50ppm (measured over 1ms interval)Horizontal Scan Range2.5ns/div – 50s/div(SHS1000)Scan mode: 100ms/div 50s/div (1-2.5-5 order )Vertical SystemVertical SensitivityChannelVoltage5mV/div – 100V/div(1-2-5 order )OffsetRange5mV-200mV: 1.6V206mV-10V: 40V20.6V-100V: 400VVertical Resolution8 bitChannels2Analog Bandwidth100MHz (SHS1102)60MHz(SHS1062)Single Bandwidth100MHz (SHS1102)60MHz(SHS1062)Lower Frequency(AC-3dB) 10Hz (at input BNC)DC Gain Accuracy5mv/div-100v/div: 3%DC [3.0%X( reading offset ) 1% X offset 0.2div 2mV]MeasurementAccuracy 100mv/divDC Measurement Accuracy [3.0%X( reading offset ) 1% X offset 0.2div 100mV] 100mv/divRise Time 3.5ns (SHS1102) 5.8ns (SHS1062)Vertical Input CouplingAC, DC, GNDMath Operation , -, * , /, FFTFFTBandwidth LimiterWindow Mode: Hanning, Hamming, Blackman, RectangularSampling: 1024 points20MHz (-3dB)Trigger SystemTrigger TypesEdge, Pulse Width, Video, Slope, AlternativeTrigger SourceCH1, CH2Trigger ModesAuto, Normal, SingleTrigger CouplingTrigger Level RangeTrigger DisplacementAC, DC, LF rej, HF rejCH1, CH2: 6 divisions from center of screenPre-trigger: (Memory depth/(2*sampling)),Delay Trigger: 268.04divHoldoff Range100ns – 1.5sEdge TriggerEdge Type: Rising, Falling, Rising and FallingPulse Width TriggerVideo TriggerTrigger Modes: ( , , ) Positive Pulse Width, ( , , )Negative Pulse WidthPulse Width Range: 20ns – 10sSupport Signal Formats: PAL/SECAM, NTSCTrigger Condition : Odd Field, Even Field, All Lines, Line Num

Slope TriggerAlternative Trigger( , , ) Positive slope, ( , , ) Negative slopeTime: 20ns-10sCH1 Trigger Type: Edge, Pulse, Video, SlopeCH2 Trigger Type: Edge, Pulse, Video, SlopeX-Y ModeX-Pole Input /Y-Pole InputSample FrequencyChannel 1 (CH1) / Channel 2 (CH2)XY mode has a breakthrough that trad oscilloscopes restrict sampling rate at 1MSa/s and supports5KSa/s 500MSa/s:Measure SystemAuto Measure(32 Types)Cursor MeasureVpp, Vmax, Vmin, Vamp, Vtop, Vbase, Vavg, Mean, Crms, Vrms, ROVShoot, FOVShoot, RPREShoot,FPREShoot, Rise time, Fall time, Freq, Period, Wid, -Wid, Dut, -Dut, BWid, Phase, FRR, FRF,FFR, FFF, LRR, LRF, LFR, LFFManual mode, Track mode and Auto modeControl Panel FunctionAuto SetSave/RecallAuto adjusting the Vertical, Horizontal system and Trigger PositionSupport 2 group referenced waveforms, 20 group setups,10 group captured waveforms internalstorage/recall function and USB flash driver storage function.Hard Ware Frequency CounterReading Resolution1HzRangeDC Couple, 10Hz to MAX BandwidthSignal TypesSatisfying all Trigger signals(Except Pulse width trigger and Video Trigger)MultimeterMaximum Resolution6000 countsMeasure FunctionDCV, ACV, DCI, ACI, Resistance, Diode, Capacitance, ContinuityMax Input VoltageAC(Vrms): 750V (AC frequency :20Hz 1kHz)DC :1000VMax Input CurrentAC (Vrms) : 10A (AC frequency :20Hz 1kHz)DC : 10AImpedance10MΩ

RecorderScope TrendPlotDisplayAll, NormalRecord Size800K points, more than 24 hoursRecord Channel2 channelsCursor, ZoomYesManual ModeYesMeter TrendPlotDisplayAll, NormalRecord Size1.2M pointsRecord Channel1 channelCursor, ZoomYesManual ModeYesScope RecordFunctionRecord scope waveforms, Replay recorded waveformsAcquisition ModeScan ModeTimeRecord mode: recording timeReplay mode: replay timeViewer: full screen, split screenSetsRecord mode: continuous, singleReplay mode: point, frameSave mode: Internal memoryViewer: split screenDefaultRecord mode: continuousReplay mode: pointSave mode: Internal memoryTotal: 7M pointsSingle channel: 7M points single channelRecord SizeDouble channels: 3.5M points per channelAt different time base, get max record time , e.g. time base 100ms, each point counts 0.04ms, TotalTime 7000000*0.04ms 4.6minRecord ManualStart, Pause, Stop, ContinueReplay ManualStart, Pause, Stop, Continue, Previous, Next,

Generic SpecificationDisplay SystemDisplay Mode5.7 inch TFT color LCDResolution320 horizontal by 234 vertical pixelsDisplay Color24 (Typicalstate)150:1300nitWaveform Display Range8 x 12 divWaveform Display ModePoint, VectorPersistOff, 1 sec, 2 sec, 5 sec, InfiniteMenu Display2 sec, 5 sec, 10 sec, 20 sec, InfiniteScreen-SaverOff, 1min, 2min, 5min, 10min, 15min, 30min, 1hour, 2hour, 5hourSkinClassical, Modern, Tradition, SuccinctWaveform InterpolationSin(x)/x, LinearColor modelNormal , InvertLanguageSimplified Chinese, Traditional Chinese, English, Arabic, French, German, Russian, Spanish,Portuguese,Japanese, Korean, ItalianPowerLine Power AdapterInput voltage100V-240V 50/60HzOutput voltage9V 4ABattery7.4VDC, 5000mAh, persisting about 4 hoursCharge timeAbout 4 hoursEnvironmentsTemperatureOperating0 45 Storage-20 70 CoolingInternal fan usedHumidity85%RH, 40 63.2mmheight53.3mm1.5Kg

Type Selections:NAME:SHS1000 series Handheld Digital OscilloscopeTYPE:SHS1102 100MHzSHS1062 60MHzStandard accessories:A 9V, 3A, power adapterTwo special 10:1, CATII 1000V, CATIII 600V, 100MHz oscilloscope probesTwo test leads for multimeterA USB data transmitting cableUser ManualA service warranting cardA CD of Easyscope used for PC controlContact usIf you need any technical support or other help, you can contact SIGLENT service departmentdirectly.SIGLENT TECHNOLOGIES CO.,LTDAdd: 3/F, Building 4, Antongda industry area,3rd LiuXan Rd, Baona District, ShenZhen, China.Tel: 0086 755 3661 5186E-mail:sales@siglent.comTo get latest product and service information please log in http://www.siglent.com

Support EasyScope software Standard SPCI command collections, support telecommuting Multiple Language User Interface, support Multilingual help system online