ELGAPURELAB Ultra Mk 2 Operator ManualCopyright NoteThe information contained in this document is the property of ELGALabWater, a division of VWS (UK) Ltd. and is supplied without liabilityfor errors or omissions.No part of this document may be reproduced or used except asauthorized by contract or other written permission from VWS (UK) Ltd.The copyright and all restrictions on reproduction and use apply to allmedia in which this information may be placed.VWS (UK) Ltd. pursue a policy of continual product improvement andreserve the right to alter without notice the specification, design, priceor conditions of supply of any product or service. VWS (UK) Ltd. 2007All rights reserved.Publication ref: MANU38390Version 3 11/08ELGA LabWater is a trading name of VWS (UK) Ltd.ELGA and PURELAB are registered trademarks.PURELAB Ultra Mk 2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage i

PURELAB Ultra Mk2 Operator ManualELGATABLE OF CONTENTS1.2.3.Product Range . 11.2Use of this Manual . 11.3Customer Support . 1HEALTH AND SAFETY NOTES . 22.1Electricity. 22.2Ultraviolet Light . 22.3Sanitization Chemicals. 22.4Control of Substances Hazardous to Health(COSHH). 2PRODUCT AND PROCESS DESCRIPTION . 33.1Product Description. 33.2Process Description . 43.3Technical Specification . 9CONTROLS . 125.INSTALLATION INSTRUCTIONS . 137.Downloaded from www.Manualslib.com manuals search engine1.14.6.Page iiINTRODUCTION. 15.1Unpacking the PURELAB Ultra . 135.2Positioning the PURELAB Ultra . 135.3Connecting the PURELAB Ultra . 155.4Initial Controller Set up. 175.5Setting Up Password / Resetting ReplacementTimers . 225.6Initial Start Up . 245.7POU Filter Installation (LC134) - OptionalAccessory . 26OPERATION . 276.1Intermittent or Continuous Recirculation . 276.2Manual Dispense . 286.3Profile Dispense . 286.4Dispense Porfiling . 286.5Automatic Calibration . 296.6Water Supply Flush. 30MAINTENANCE . 317.1Replacing the Purification Packs . 317.2Replacing the Ultra-Microfilter (LC109). 337.3Replacing the Ultrafilter (LC151). 357.4Replacing the Ultraviolet Lamp (LC105 andLC118) . 377.5Replacing the Point-Of-Use Filter (LC134) . 387.6Cleaning the Inlet Strainer (External) . 39PURELAB Ultra Mk 2 Version 3 11/08

ELGAPURELAB Ultra Mk 2 Operator Manual8.SANITIZATION PROCEDURES . 408.1CT1 Sanitization Tablet - Safety Information. 429.TROUBLE SHOOTING . 4310.CONSUMABLES AND ACCESSORIES . 4411.KEY TO CONTROL PANEL . 4511.1 Icons . 4511.2 Alarm Conditions . 4611.3 Replacement Timers . 4611.4 Quality Alarms . 4712.WARRANTY/CONDITIONS OF SALE . 4813.USEFUL CONTACT DETAILS . 50PURELAB Ultra Mk 2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage iii

Downloaded from www.Manualslib.com manuals search engine

ELGAPURELAB Ultra Mk 2 Operator Manual1. INTRODUCTION1.1Product RangeThis operator manual has been prepared for the PURELAB Ultraproduct models:PURELAB Ultra Mk 2 ScientificPURELAB Ultra Mk2 BiosciencePURELAB Ultra Mk2 GeneticPURELAB Ultra Mk2 IonicPURELAB Ultra Mk2 AnalyticPURELAB Ultra1.2Use of this ManualThis manual contains full details on installation, commissioning andoperation of the PURELAB Ultra unit. If the instructions in thishandbook are not followed then the performance of this productand/or the safety of the user may be compromised.1.3Customer SupportService support and consumable items are available from your localsupplier or distributor. Refer to customer service contact details shownat the end of this publication.PURELAB Ultra Mk2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage 1

PURELAB Ultra Mk2 Operator ManualELGA2. HEALTH AND SAFETY NOTESPURELAB Ultra products have been designed to be safe, however, itis important that personnel working on these units understand anypotential dangers. All safety information detailed in this handbook ishighlighted as WARNING and CAUTION instructions. These are usedas follows:Mains power socketFuseWARNING!WARNINGS ARE GIVEN WHERE FAILINGTO OBSERVE THE INSTRUCTION COULDRESULT IN INJURY OR DEATH TOPERSONS.CAUTION!Cautions are given where failure toobserve the instruction could result indamage to the equipment, associatedequipment and processes.ON/OFFswitchMains power lead2.1Mains Power SupplyElectricityIt is essential that the electrical supply to the PURELAB Ultra isisolated before any items are changed or maintenance workperformed.The ON/OFF switch is located at the left-hand side of the unit. Themains power lead is located just behind the ON /OFF switch.WARNING!THIS APPLIANCE MUST BE EARTHED.The main water supply should be isolated and residual pressurereleased prior to removal of any Purification Packs or carrying outwork on the unit.Switching off the electrical supply will isolate the source of mainswater pressure.2.2Ultraviolet LightWARNING!UNDER NO CIRCUMSTANCES SHOULDTHE LAMP BE CONNECTED ANDACTIVATEDWHENOUTSIDETHEHOUSING.The Genetic, Analytic and Ionic units are fitted with an ultraviolet lamp.The UV lamp is enclosed in a stainless steel chamber ensuring thatthe operator will not be exposed to UV light.2.3Sanitization ChemicalsDuring the sanitization cycle a CT1 sanitization tablet is used andrelevant safety information is included in this handbook. A safety datasheet conforming to COSHH regulations is also provided with thetablets and should be read before a tablet is used.2.4UV LampControl of Substances Hazardous to Health (COSHH)Material safety data sheets covering the various replaceablePurification Packs are available upon request. Contact your localsupplier or distributor.Ultra with UV LampPage 2Downloaded from www.Manualslib.com manuals search enginePURELAB Ultra Mk 2 Version 3 11/08

ELGAPURELAB Ultra Mk 2 Operator Manual3. PRODUCT AND PROCESSDESCRIPTION3.1Product DescriptionThe PURELAB Ultra water purification unit has been specificallydesigned to provide a supply of ultrapure water with very low levelsof impurities for laboratory, medical and industrial applications.The PURELAB Ultra can be bench, under bench or wall mountedwith an optional wall mounting kit. A range of accessories is availableto complement the unit (see Section 10 - Consumables andAccessories, for details).Control PanelDispense NozzleDispense ControllerDoorPURELAB UltraThe PURELAB Ultra range of water purification units has beendesigned to provide the ultimate in high purity water for laboratoryapplications. The products need to be fed with pre-treated water,typically from a reverse osmosis supply. They can be operated directlyfrom a ringmain feed as point of use polishers, with or without localbuffer storage, using a Docking Vessel or suitable reservoir.Alternatively, they can be supplied from a local pre-treatment/storagesystem, for example a PURELAB Prima and Docking Vesselcombination.PURELAB Ultra Mk2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage 3

PURELAB Ultra Mk2 Operator ManualELGA3.2Process Water QualitySensorOutlet WaterQuality SensorNRVSolenoid ValveSolenoidValveFeedwater Inlet(Pre-purified Water)Dispense ControllerPumpPrimaryPurificationPackSolenoid ValvePolishingPurificationPack0.2µm Point-of-use Filter(Optional)Purified WaterOutletDrainPre-treated water enters via an inlet solenoid, and is pumpedthrough the primary purification pack, an intermediate waterquality sensor, a polishing purification pack and finaltemperature and water quality sensors before beingdispensed or re-circulated through a non-return valve back tothe pump inlet.The first purification pack removes most of the impuritiesfrom the water. The intermediate water quality sensormeasures the resistivity of the water from the first pack andinitiates an alarm when the pack needs to be replaced.Any remaining ionic and organic impurities are removed bythe second polishing purification pack. Product waterresistivity and temperature are measured before dispense.The water within the unit is re-circulated through thepurification technologies to maintain purity. To minimize heatbuild up the recirculation is at reduced flow rate and can beset to be intermittent (10 minutes every hour). Similarly,intermittent recirculation can be used overnight in 'sleepmode'.An optional POU filter is available if required to protect theoutlet from bacterial contamination.Page 4Downloaded from www.Manualslib.com manuals search enginePURELAB Ultra Mk 2 Version 3 11/08

ELGAPURELAB Ultra Mk 2 Operator Quality SensorOutlet WaterQuality SensorNRVFeedwater Inlet(Pre-purified aryPurificationPackSolenoid ValveDrainPolishingPurificationPackSolenoid ValveDispense ControllerFlushSolenoidValveDrain0.2µm Point-of-use Filter(Optional)Purified WaterOutletPre-treated water enters via an inlet solenoid and is pumpedthrough the primary purification pack, an intermediate waterquality sensor, a polishing purification pack, an ultrafilter andfinal temperature and water quality sensors before beingdispensed or re-circulated through a non-return valve back tothe pump inlet.The first purification pack removes most of the impuritiesfrom the water. The intermediate water quality sensormeasures the resistivity of the water from the first pack andinitiates an alarm when the pack needs to be replaced.Any remaining ionic and organic impurities are removed bythe second polishing purification pack; the ultrafilter removespyrogens, bacteria and other microbial impurities as well asparticles. Product water resistivity and temperature aremeasured before dispense.The water within the unit is re-circulated through thepurification technologies to maintain purity. To minimize heatbuild up the recirculation is at reduced flow rate and can beset to be intermittent (10 minutes every hour). Similarly,intermittent recirculation can be used overnight in 'sleepmode'.An optional POU filter is available if required to protect theoutlet from bacterial contamination.PURELAB Ultra Mk2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage 5

PURELAB Ultra Mk2 Operator t utlet WaterQuality SensorNRVFeedwater Inlet(Pre-purified aryPurificationPackSolenoid ValveDrainUV LampSolenoid ValveDispense lveDrain0.2µm Point-of-use Filter(Optional)Purified WaterOutletPre-treated water enters via an inlet solenoid, and is pumpedthrough the primary purification pack and an intermediatewater quality sensor. The water then flows through a UVchamber, further resistivity and temperature sensors, apolishing purification pack, an ultrafilter and final temperatureand water quality sensors before being dispensed or recirculated through a non-return valve back to the pump inlet.The first purification pack removes most of the impuritiesfrom the water. The intermediate water quality sensormeasures the resistivity of the water from the first pack andinitiates an alarm when the pack needs to be replaced.Purified water flows directly through the UV chamber where itis exposed to intense UV radiation at wavelengths of 254nmand 185nm to provide continuous bacterial control and thephoto-oxidation of residual organic impurities. The secondtemperature compensated water quality sensor provides datafor TOC monitoring.Any remaining ionic and organic impurities are removed bythe second polishing purification pack; the ultrafilter removespyrogens, bacteria and other microbial impurities as well asparticles. Final water resistivity and temperature aremeasured before dispense.The water within the unit is re-circulated through thepurification technologies to maintain purity. To reduce heatbuild up the recirculation is at reduced flow rate and can beset to be intermittent (10 minutes every hour). Similarly,intermittent recirculation can be used overnight in 'sleepmode'.An optional POU filter is available if required to protect theoutlet from bacterial contamination.Page 6Downloaded from www.Manualslib.com manuals search enginePURELAB Ultra Mk 2 Version 3 11/08

ELGAPURELAB Ultra Mk 2 Operator ManualIonicTemperatureSensorOutlet WaterQualitySensorIntermediateWater QualitySensorNRVUMFSolenoidValveSolenoid ValveFeedwater Inlet(Pre-purified Water)PumpPrimaryPurificationPackSolenoid ValveUV LampDispense ControllerPolishingPurificationPack0.2µm Point-of-use Filter(Optional)Purified WaterOutletDrainPre-treated water enters via an inlet solenoid and is pumpedthrough the primary purification pack and an intermediatewater quality sensor. The water then flows through a UVchamber, a further resistivity and temperature sensor, apolishing purification pack, an ultra microfilter and finaltemperature and water quality sensors before beingdispensed or re-circulated through a non-return valve back tothe pump inlet.The first purification pack removes most of the impuritiesfrom the water. The intermediate water quality sensormeasures the resistivity of the water from the first pack andinitiates an alarm when the pack needs to be replaced.Purified water flows directly through the UV chamber where itis exposed to intense UV radiation at a wavelength of 254nmto provide continuous bacterial control.Any remaining ionic and organic impurities are removed bythe second polishing purification pack; the ultra microfilterremoves bacterial impurities and particles. Product waterresistivity and temperature are measured before dispense.The water within the unit is re-circulated through thepurification technologies to maintain purity. To reduce heatbuild up the recirculation is at reduced flow rate and can beset to be intermittent (10 minutes every hour). Similarly,intermittent recirculation can be used overnight in 'sleepmode'.An optional POU filter is available if required to protect theoutlet from bacterial contamination.PURELAB Ultra Mk2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage 7

PURELAB Ultra Mk2 Operator ManualELGAAnalyticIntermediateTemperaturePost SensorOutlet WaterQualitySensorNRVSolenoidValveFeedwater Inlet(Pre-purified Water)UMFPumpSolenoid ValvePrimaryPurificationPackUV LampPolishingPurificationPackSolenoid ValveDispense Controller0.2µm Point-of-use Filter(Optional)Purified WaterOutletDrainPre-treated water enters via an inlet solenoid and is pumpedthrough the primary purification pack and an intermediatewater quality sensor. The water then flows through a UVchamber, further resistivity and temperature sensors, apolishing purification pack, an ultra microfilter and finaltemperature and water quality sensors before beingdispensed or re-circulated through a non-return valve back tothe pump inlet.The first purification pack removes most of the impuritiesfrom the water. The intermediate water quality sensormeasures the resistivity of the water from the first pack andinitiates analarm when the pack needs to be replaced.Purified water flows directly through the UV chamber where itis exposed to intense UV radiation at wavelengths of 254nmand 185nm to provide continuous bacterial control and thephoto-oxidation of residual organic impurities. The secondtemperature compensated water quality sensor provides datafor TOC monitoring.Any remaining ionic and organic impurities are removed bythe second polishing purification pack; the ultra microfilterremoves bacterial impurities and particles. Product waterresistivity and temperature are measured before dispense.The water within the unit is re-circulated through thepurification technologies to maintain purity. To reduce heatbuild up the recirculation is at reduced flow rate and can beset to be intermittent (10 minutes every hour). Similarly,intermittent recirculation can be used overnight in 'sleepmode'.An optional POU filter is available if required to protect theoutlet from bacterial contamination.Page 8Downloaded from www.Manualslib.com manuals search enginePURELAB Ultra Mk 2 Version 3 11/08

ELGAPURELAB Ultra Mk 2 Operator Manual3.3Technical SpecificationThe Technical Specifications for the PURELAB Ultra are as follows:Feedwater*ParameterLimitsSource - Originally from potable supply,then pre-treatedPreferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled.Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion.Fouling Index (max)1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feedsService Deionisation (SDI) - MΩ-cm1 MΩ-cm minimum resistivity at exhaustionReverse Osmosis (RO) - µS/cmRecommended 30 µS/cmFree Chlorine0.05 ppm max.TOCRecommended 50 ppb max.Carbon Dioxide30 ppm max.Silica2 ppm max.ParticulatesFiltration down to 0.2 micron advisable to protect internal and/or point of use filtersTemperature1-40 C - Recommended 10-15 CFlowrate (maximum requirement)130 l/hrDrain requirements (gravity fall with air gap).Maximum during serviceUp to 2 l/minFeedwater Pressure0.7 bar (10 psi), maximum, 0.07 bar (1 psi) minimum* Contact technical support for advice on feedwaters outside the range listed.Note:Different system configurations are available for different feedwater sources. See system set up.DimensionsHeight490mm (19.3")Width410mm (16.2")Depth365mm (14.4")WeightScientific14.0kg (30.8 lb)Bioscience14.5kg (32.0 lb)Ionic14.5kg (32.0 lb)Analytic15.0kg (33.1 lb)Genetic15.0kg (33.1 lb)ConnectionsInlet-quick connect8mm (5/16") ODDrain8mm (5/16") ODFlush - Bioscience/ Genetic8mm (5/16") ODPositioningWall, bench or under bench mountedEnvironmentClean dry indoor. Temp 5-40 CHumidity max 80% non-condensingPURELAB Ultra Mk2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage 9

PURELAB Ultra Mk2 Operator ManualELGAElectrical RequirementsMains input100 -240V ac, 50-60Hz all modelsSystem voltage24V dcPower consumption during recirculation60VAPower consumption during dispense75VAFuses2 x T6.3 AmpReservoir level connectionJack Plug 3.5mmNoise level during recirculation 40dBAUser InterfaceDisplayContinuous graphical quality displayGraphical flow schematic on screen with mimic displayBacklit display with Intuitive IconsAdjustable settingsIndicatorsAlarms-AudiovisualOutputsTimed dispenseSelectable dispense times (10 secs - 60 min)Profile dispenseSelectable dispense volumesRecirculation modeIntermittent or continuousFeed water typeSelectable (RO/DI/SDI)Display viewing angleAdjustable electronicallyWater quality unitsSelectable (MΩ-cm or µS/cm)Water quality alarmSelectable alarm setpointsDate / timeAdjustableAuto restart after power failureSelectable (On/Off)Audible alarmSelectable (On/Off)Temperature controlSelectable alarm setpointsNight serviceSelectable (On/Off)Product water qualityResistivity or conductivityIntermediate water qualityResistivity or conductivityTemperatureDegrees Centigrade or FahrenheitTotal Organic Carbon (TOC)*Concentration ppb C (µg/l C)Purification packReplacement dateUV lamp*Replacement dateFilter*Replacement datePurified water purityAbove set point alarmTemperature****Above set point alarmUV lamp operation*Imminent lamp failure or failure to startCalibrationCalibration alarm – calibration requiredPurification packChange reminderUV lamp*Change reminderFilter*Change reminderReservoir****Low level / level control disconnect alarmRS232 printer connectionRS232 remote display connectionVolt free contact-internalRemote dispensePage 10Downloaded from www.Manualslib.com manuals search enginePURELAB Ultra Mk 2 Version 3 11/08

ELGAPURELAB Ultra Mk 2 Operator ManualSafety FeaturesPower fail safeWater temperature alarmWater quality alarmPurification pack interlockUV current monitoringTimeout of manual dispenseLow operating voltage 24VDry contact alarm connectionVisual alarmsAudible alarmsOperational FeaturesLow noise levels – minimum intrusionContinuous calibration to traceable standardsDrop dispenseMultistage monitoringRestart on power interruptProfile dispenseTimed dispenceOptional printer kit to record operating parametersOptional remote displayPurified Water rate2.0 l/min max.InorganicResistivity 18.2 MΩ-cm (0.055µS/cm)TOC3-10 ppb**3-10 ppb**Bacteria 1 cfu/ml*** 1 cfu/10 ml, *** 1 cfu/ml (without POU fitted)1-2 ppb**Bacterial EndotoxinspHParticles3-10 ppb** 0.001 Eu/ml 0.001 Eu/mlEffectively neutral0.2 µm***RNase / DNaseCapacity (LC182)3-10 ppb**Genetic0.05 µm-UltrafiltrationRemovedLiters at 18.2 MΩ-cm 80,000 / (µS/cm at 2.3 x ppm CO2 )*On some models only.**Dependant on feedwater - recommended feed 50 ppb TOC.*******With POU filter fitted.Switches process off.Purification PacksLabpure S1 LC182 recommended for most applications. Refer to Section 10 for alternative cartridges.As part of our policy of continual improvement we reserve the right toalter the specifications given in this document.PURELAB Ultra Mk2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage 11

PURELAB Ultra Mk2 Operator ManualELGA4. CONTROLSProcess buttonThe PURELAB Ultra operates with a tactile membrane control panel,which has a graphics display window and four multi-purpose controlbuttons.Details of how to use the controls will be given in the appropriatesections.The PURELAB Ultra control panel has a range of control icons asfollows:ButtonIconPROCESSLeft handControl buttonCentreControl buttonFunctionTurns the processon/offRight handControl buttonESCAPE from anyfeatureControl PanelLEFTMenuScrollM1Profile nseM2RIGHTProfile dispensememoryMute AlarmPrinterM3Profile dispensememoryIntermittentrecirculationPage 12Downloaded from www.Manualslib.com manuals search enginePURELAB Ultra Mk 2 Version 3 11/08

ELGAPURELAB Ultra Mk 2 Operator Manual5. INSTALLATION INSTRUCTIONS5.1Unpacking the PURELAB UltraThe following items are supplied with your PURELAB Ultra:5.21.PURELAB Ultra unit.2.Sanitization by-pass blocks LA638 (2 off, fitted in unit).3.6 metres of 8mm (5/16") O/D tube.4.1 off Pack of CT1 tablets.5.1 off Strainer Assembly.6.Operator Manual in English.7.Multilingual Operator Manual on CD.8.Mains Lead.9.Reservoir Level Connecting Lead.Positioning the PURELAB UltraBefore installation and operation of the PURELAB Ultra unit, pleaseread and observe the following points:EnvironmentThe unit should be installed on a flat, level surface, in aclean, dry environment. The unit can also be wall mountedagainst a vertical wall capable of supporting the weight (forthis we recommend the use of the wall mounting kit Part No.LA610 on substantial brick/concrete walls or LA622 for studpartition walls).WARNING!IF THE UNIT IS TO BE WALL MOUNTED,ENSURE IT IS MOUNTED USING THECORRECT WALL MOUNTING KIT ANDTHAT THE WALL IS CAPABLE OFSUPPORTING THE OPERATING WEIGHTOF THE SYSTEM. ALWAYS CAREFULLYFOLLOW THE INSTRUCTIONS INCLUDEDIN THE KIT.Note:Refer to Specifications for unit weights(Section 3.4 - Technical Specification).Rear Mounting PointsThe unit is designed to operate safely under the following conditions: Indoor Use. Altitude up to 2000m. Temperature Range 5 - 40 C. Maximum Relative Humidity 80% at 31 C decreasing linearlyto 50% at 40 C, non-condensating.Unit Rear Mounting PointsThe unit is in Installation Category II, Pollution Degree 2, as SafetyStandard IEC1010-1.PURELAB Ultra Mk2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage 13

PURELAB Ultra Mk2 Operator ManualMains power socketFuseON/OFFswitchMains power leadELGAElectricalThe units can be connected universally to any electricalsupply in the range of 100 - 240V and 50 - 60Hz. The mainslead is supplied with a molded plug on one end and a moldedconnector to the unit on the other. The unit should beconnected to an earth. The unit includes a battery which willrequire changing on a periodic basis, typically every 3-5years. (Contact local Service provider)WARNING!Electrical ConnectionsDANGER OF EXPLOSION IF BATTERY ISINCORRECTLY REPLACED. REPLACEONLY WITH THE SAME OR EQUIVALENTTYPERECOMMENDEDBYTHEMANUFACTURER (3 VOLTS, LITHIUMCR2032). DISPOSE OF USED BATTERIESACCORDING TO THE MANUFACTURER’SINSTRUCTIONS.Drain/FlushA semi-rigid flexible connection to a sink or suitable draincapable of at least 2.0 l/min is required. The drain pointshould have a gravity fall below the level of the unit and anyconnections direct to drain should have an air-break devicefitted. On the Genetic & Bioscience units a separate flush lineto drain is required. The drain and flush drain lines should runindividually to a convenient drain point.Feed WaterThe feedwater should be potable water, pre-purified usingreverse osmosis, deionisation, or distillation. If using a supplyother than reverse osmosis treated, it is stronglyrecommended that a 0.2 micron membrane filter is installedprior to remove colloidal impurities. The feedwater shouldenter the PURELAB Ultra via an 8mm (5/16") O/D semi rigidtube, and should be in the temperature range 1 to 40 C (34to 104 F). To ensure optimum performance it isrecommended that the feedwater temperature be between 10to 15 C.CAUTION!Operating with feed water temperaturesoutside the range from 1 to 40 C (33.8 to104 F) will cause damage to the PURELABUltra unit.For pressurized feeds, the minimum direct inlet pressure is0.07 bar (1.0 psi) and maximum inlet pressure is 0.7 bar(10 psi). Feedwater pressures up to 4 bar (60 psi) must bereduced using a pressure regulating valve (Part No. LA652).CAUTION!High feedwater pressures which mayexceed 4bar (60 psi) must be reducedusing a pressure regulator with built inpressure relief (Part No. LA575).Failure to install the correct pressureregulator will cause damage to thePURELAB Ultra unit.Reservoir feeds to the PURELAB Ultra unit should bepositioned at the same height, or above the unit, to provide apositive flooded inlet pressure of approximately 0.07 bar(1 psi).Page 14Downloaded from www.Manualslib.com manuals search enginePURELAB Ultra Mk 2 Version 3 11/08

ELGAPURELAB Ultra Mk 2 Operator ManualConnectorColletBlankingtransit plug5.3Connecting the PURELAB UltraOnce the PURELAB Ultra unit has been positioned either on a wall oron a bench, it should be connected as follows: Mains water inlet tubePull plug Drain tube - to drain Flush tube (Bioscience and Genetic) - to drainDepresscolletClean squarecut endTube 8mm ODStep 1 - Remove Transit Plugs1. PUSH in collet on connector.2. PULL out transit plug.3. CUT a clean square end on an 8 mm OD semi rigiddrain tube.4. PUSH tube into connector.Simply push in tube to attachTube in secured positionFitting tubesStep 2 - Connect Water Inlet1. Connect 8 mm tube from the PURELAB Ultra inlet tothe inlet strainer.2. Connect the inlet strainer to the pre-purified watersupply.Direction ofwater flowCollarCollarMesh filterStrainerInletFeed Water InletStep 3 - Connect Drains1. Connect 8 mm tube from the drain and UF drain (iffitted) and direct to a suitable drainage point.CAUTION!The drain and flush lines should allow agravity fall to drain with no restrictions.UF DrainDrainInletInlet and Drain ConnectionPURELAB Ultra Mk2 Version 3 11/08Downloaded from www.Manualslib.com manuals search enginePage 15

PURELAB Ultra Mk2 Operator ManualELGAMains power socketFuseON/OFFswitchStep 4 - Connect Electrical Supply1. PLUG mains power lead into the mains power socketon the left hand side of the PURELAB Ultra unit.2. PLUG mains power lead into mains socket.Mains power leadElectrical ConnectionsRS232ReservoirlevelconnectionStep 5 - Resevoir Level Connection1. INSERT jack plugs into the level control socket locatedat rear of unit and reservoir.Note:LevelcontrolIf wall mounting the unit it is recommendedthat the jack plug should be inserted prior tolocating the unit on the wall mounting bracket.Reservoir Level ConnectionPage 16Downloaded from www.Manualslib.com manuals search enginePURELAB Ultra Mk 2 Version 3 11/08

ELGAPURELAB Ultra Mk 2 O

ELGA PURELAB Ultra Mk 2 Operator Manual PURELAB Ultra Mk2 Version 3 11/08 Page 3 3. PRODUCT AND PROCESS DESCRIPTION 3.1 Product Description The PURELAB Ultra water purification unit has been specifically designed to provide a supply of ultrapure water with very low levels of impurities for laboratory, medical and industrial applications.