Agilent 4263B LCR Meter 100 Hz To 100 KHz

Transcription

Agilent4263B LCR Meter 100 Hz to 100 kHzTechnical OverviewIntroductionThe Agilent Technologies LCR metermakes fast measurements on components. It is optimized for applicationsthat require precision and versatility.The instrument's performance rangesfrom general bench-top impedancemeasurements to complex transformer, coil and electrolytic capacitormeasurements. The LCR meter offersfast, reliable, and versatile testing ata low cost.Satisfy your needs for.Fast system test throughput Maximize testing with rapid 25 msmeasurements Minimize user intervention withpass/fail testing Communicate results with displayand GPIB Automate testing with built-inhandler interfaceFault-free results Test with confidence using contactcheck function Remove parasitics with error correction Get the best data with 0.1% basicaccuracy Eliminate trigger timing errorswith trigger delay functionVersatile measurements Select from 11 impedanceparameters Add three complex transformerparameters with Option 4263B-001 Set signal level with 5mVrmsresolution Monitor actual ac voltage andcurrent levels Pick from many test fixtures andaccessories Save and recall up to tenmeasurement setups

Display:LCD with back-light.Displays measurement values,instrument states, and comparator resultsFour-terminal pair:Reduces errors dueto cable extensionI&V monitor:Measurement parameter:Select the desired test parameter Monitor actualac voltage andcurrent levelsCable:Extend front panel cablelength from 0 to 4 metersSave/recall:Store and retrieve upto 10 measurement statesTrigger mode:Choose internal,external, or manual triggerComparator:Select values for HIGH,IN, and LOW testingContact check:Verify reliability of testconnectionsFrequency:Select one of five test frequenciesOpen/short/load:Correct for test fixture and cableerrorsKey Parameters and SpecificationsTest frequencies:100 Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHzOption 4263B-002 adds 20kHzAC test signal levels:20m–1Vrms, 5mVrms stepsBasic accuracy:0.1%Bias:1.5 and 2.0 VdcImpedance parameters: Z , R, X, Y , G, B, C, L, D, Q, uError correction:Open, short, and loadOption 4263B-001 adds transformermeasurement functions: turns-ratio,mutual-inductance and dc-resistanceBuilt-in system features:GPIB and handler interfacesCable length settings:0, 1, 2, 4 metersMeasurement time (typical):25 ms at best conditionsContact check time (typical):5 ms per measurement2

High-quality resultsSystem features for test automation See five digits of data Make precise measurements with0.1% basic accuracy Select from 11 impedanceparameters Verify device performance atsimulated operating conditions Monitor actual test signal voltageand current levels Maximize accuracy with errorcorrection Use performance specified with0, 1, 2, and 4 meter cables Test device contact failure withcontact check function Automate testing with GPIBinterface Reduce ground-loops with isolatedhandler interface Continue testing after ac powerloss with continuous memory Perform pass/fail testing withcomparator function(High/In/Low)Evaluate transformers and coilswith Option 4263B-001 Measure turns-ratio, mutualinductance and dc-resistance Easily make connections with16060A transformer test fixture Measure parameter responses withvariable signal levelsSimplify transformer testing.Make reliable impedance measurements.Make electrolytic capacitormeasurementsThe 4263B LCR meter is designed forautomated applications. Obtain versatile testing with alarge capacitance range Keep costs down with built-in dcbias source Protect your investment: highenergy protection on terminals Increase test throughput with fastsystem measurements Make reliable handler measurements with contact check functionQuickly evaluate electrolytic capacitors.3

SpecificationsMeasurement accuracyFigure 1. Conversion diagram4Table 1. Measurement accuracy ( % of reading)

Measurement conditions1. Warm-up time: 15 min.2. Ambient temperature: 23 5 C3. Test signal voltage: 1 Vrms4. Test cable length: 0 meter5. Open and short corrections performed6. Measurement time: Medium or Long(Other test condition data is available in theoperation manual.)For Z , Y , L, C, R, X, G, and B accuracy(Ae), refer to Table 1. Table 1 equations yieldaccuracy based on frequency and DUTcharacteristic impedance (Zm). Zm isfrom Figure 1, Conversion Diagram.Other SpecificationsMeasurement parameters/rangesParameterRange Z , R, X Y , G, BCLDQu 1 mΩ to 100 MΩ10 nS to 1000S1 pF to 1 F10 nH to 100 kH0.0001 to 9.99990.1 to 9999.9-180 to 180 -999.99% to 999.99%Option 4263B-001: DC resistance 1 mΩ to100 MΩMutual inductance 1 µH to 100 H(typical)D accuracy(De) Ae/100Turns-ratio 0.9 to 200 (typical)Q accuracy(Qe) where(Qm x De 1)(Qm)2xDe1-/ (QmxDe)u accuracy (ue) 0.573 x AeAe Accuracy of Z , Y , L, C, R, X, G,and BDe D accuracyDm Measured value of DQe Q accuracyQm Measured value of Que u phase angle accuracyZm DUT impedance at test frequency inHertzMeasurement conditions and functionsTest frequency: 100 Hz, 120 Hz,1 kHz, 10 kHz, 100 kHz. (Option 4263B-002adds 20 kHz.)AC test signal level: 20 m - 1 Vrms,5 mVrms stepsOther instrument functionsTest signal level monitor:Voltage, currentError Correction: Open, Short, LoadComparator: HIGH, IN, and LOW for eachdisplayed parameterSave/recall: 10 instrument states fromnon-volatile memoryFront-end Protection:Vmax 8/C@ Vmax 250 VVmax 2/C@ Vmax 1000 VC in FaradsHandler interface: Negative logic and isolated.Signals are HIGH/IN/LOW, No-Contact,EOM, Index, Alarm, Keylock, Ext. Trigger.GPIB interface: Instrument control, TALKonly mode for LISTEN-only printers usingGPIB or Centronics/GPIB converterPhysical characteristicsPower: 90-132 Vac or 198-264 Vac.47-66 Hz. 45 VA typical.Bias:Internal: 1.5 and 2.0 VdcExternal: 0 to 3.0 VdcOperating temperature: 0 to 45 CRanging: Auto and HoldDimensions: 320 (W) x 100 (H) x300 (H) mmTrigger: Internal, Manual, and ExternalWeight: 4.5 kg (typical)Trigger delay time: 0 to 9999 ms in1 ms stepsTest cable lengths:0, 1 meter @ f 100 kHz2 meter @ f 10 kHz (20 kHz)4 meter @ f 1 kHzMeasurement time:SHORTMEDIUM25ms65msLONG500ms5

Test Fixtures/Accessories for the Agilent 4263B16060A transformer test fixtureAllows fast connections totransformers16065C external bias adapterFor external dc bias of DUT. Vmax 40 Vdc.16089C Kelvin IC clip leadsIC package clip. 1 meter length.16064B LED display/trigger boxDisplays comparator status.1.5 meter cable. External trigger.16089A Kelvin clip leadsLarge clip. 1 meter length.16034G Test fixtureFor SMD components.16089B Kelvin clip leadsMedium clip. 1 meter length.Component dimensions (L x W):0.6 mm x 0.3 mm to 5.0. mm x 1.6 mm16089D Alligator clip leadsFour clips. 1 meter length.6

Ordering information 1Test fixtures and accessoriesAgilent 4263B LCR MeterFurnished accessory: power cable16034E/G/H SMD component test fixtureOptions16043A/B4263B-0014263B-002Add N/M/DCR MeasurementFunctionAdd 20 kHz Test Frequency316065A200-Vdc external voltagebias fixture16065C40-Vdc external voltage biasadapter16089ALarge Kelvin clip leads3-terminal SMD test fixtureOptions16043A-ABAU.S. - English localization16089BMedium Kelvin clip leads16043A-ABJJapan - Japanese localization16089CKelvin IC clip leads16043B-ABAU.S. - English localization16089DAlligator clip leads16043B-ABJJapan - Japanese localization16089EKelvin clip leadsTest fixtureTest fixtures are not furnished as standard.Manual options24263B-AB0Taiwan - Chinese localization4263B-AB1Korea - Korean localization16044A 34263B-AB2China - Chinese localizationOptions4263B-ABAU.S. - English localization16044A-ABAU.S. - English localization4263B-ABDGermany - German localization16044A-ABJJapan - Japanese localization4263B-ABESpain - Spanish localization4263B-ABFFrance - French localization16047A/E 4Axial and radial test fixture4263B-ABJJapan - Japanese localization4263B-ABZItaly - Italian localizationOptions16047E-ABAU.S. - English localization16047E-ABJJapan - Japanese localization16334ASMD tweezer test fixture16048A0.94-meter/BNC test leads16048B0.94-meter/SMC test leads(Rack flange and handle kit are not compatible.)16048D1.89-meter/BNC test leadsCalibration certificate option16048E3.8-meter/BNC test leads4263B-1A716060ATransformer test fixture16064BLED display/trigger box4263B-0BW Add service manualCabinet options4263B-1CM Rackmount kit4263B-1CN Handle kit1234ISO 17025 compliant calibrationAccessories and options are priced individually.Manual is not furnished as standardMust specify one of language options (ABA or ABJ) for operation manual for shipment with product.Must specify one of language options (ABA or ABJ) for operation manual of 16047E for shipment with product.7

Agilent Email Updateswww.agilent.com/find/emailupdatesGet the latest information on the products andapplications you select.Agilent T&M Software and ConnectivityAgilent’s Test and Measurement software andconnectivity products, solutions and developernetwork allows you to take time out of connectingyour instruments to your computer with toolsbased on PC standards, so you can focus on yourtasks, not on your connections. Visitwww.agilent.com/find/connectivityfor more information.Agilent Technologies’ Test and MeasurementSupport, Services, and AssistanceAgilent Technologies aims to maximize the value youreceive, while minimizing your risk and problems. Westrive to ensure that you get the test and measurement capabilities you paid for and obtain the supportyou need. Our extensive support resources and services can help you choose the right Agilent productsfor your applications and apply them successfully.Every instrument and system we sell has a globalwarranty. Support is available for at least five yearsbeyond the production life of the product. Two concepts underlie Agilent’s overall support policy:“Our Promise” and “Your Advantage.”By internet, phone, or fax, get assistance with allyour test & measurement needsOur PromiseOur Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing newequipment, we will help you with product information,including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verifythat it works properly, help with product operation,and provide basic measurement assistance for theuse of specified capabilities, at no extra cost uponrequest. Many self-help tools are available.Europe:(tel) (31 20) 547 2323(fax) (31 20) 547 2390Your AdvantageYour Advantage means that Agilent offers a widerange of additional expert test and measurementservices, which you can purchase according to yourunique technical and business needs. Solve problemsefficiently and gain a competitive edge by contractingwith us for calibration, extra-cost upgrades, out-ofwarranty repairs, and onsite education and training,as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and techniciansworldwide can help you maximize your productivity,optimize the return on investment of your Agilentinstruments and systems, and obtain dependablemeasurement accuracy for the life of those products.Phone or FaxUnited States:(tel) 800 829 4444Canada:(tel) 877 894 4414(fax) 905 282 6495China:(tel) 800 810 0189(fax) 800 820 2816Japan:(tel) (81) 426 56 7832(fax) (81) 426 56 7840Korea:(tel) (82 2) 2004 5004(fax) (82 2) 2004 5115Latin America:(tel) (305) 269 7500(fax) (305) 269 7599Taiwan:(tel) 0800 047 866(fax) 0800 286 331Other Asia Pacific Countries:(tel) (65) 6375 8100(fax) (65) 6836 0252Email: tm asia@agilent.comOnline Assistancewww.agilent.com/find/assistProduct specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc., 2004, 2003, 2000, 1995Printed in USA, April 2, 20045964-6181E

Agilent 4263B LCR Meter 100 Hz to 100 kHz Technical Overview. 2 Key Parameters and Specifications Test frequencies: 100 Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHz Option 4263B-002 adds 20kHz AC test signal levels: 20m-1Vrms, 5mVrms steps Basic accuracy: 0.1% Impedance parameters: