User Manual for Easy ECG MonitorInstructions to UserDear Users,Thank you very much for purchasing our product. Pleaseread the manual very carefully before using this device.These instructions describe the operating procedures tobe followed strictly. Failure to follow these instructionsmay cause measuring abnormality or device damage.This User Manual is published by Shenzhen CreativeIndustry Co., Ltd. (Creative) in English and Creativehas the ultimate right to explain the Manual. No part ofthis manual may be photocopied, reproduced ortranslated into another language without the prior writtenconsent. Creative reserves the right to improve andamend it at any time without prior notice.Version of This Manual: Ver2.0All rights reserved.3502-2430001I

User Manual for Easy ECG MonitorAnnouncement The contents contained in this manual are subject tochange without prior notice. Information furnished by Creative is believed to beaccurate and reliable. However, no responsibility isassumed by us for its use, or any infringements ofpatents or other rights of third parties that may resultfrom its use.Declaration of Conformity:Shenzhen Creative Industry Co., Ltd. herebydeclares that this device complies with thefollowing standards:IEC 60601-1IEC 60601-1-2and follows the provisions of the CouncilDirective MDD 93/42/EEC.II

User Manual for Easy ECG MonitorWARNINGS! Check the device to make sure that there is no visibledamage that may affect user’s safety andmeasurement performance. Stop using the device,when there is obvious damage. Do NOT make diagnosis oneself by the measurementand measurement results, always consult the doctor ifabnormal information is presented frequently. It is NOT designed or intended for medical diagnosis. Do NOT use the device in the bathroom or moistcircumstance. Do NOT use it if you carry a pacemaker. Do NOT operate in the environment where strongelectro-magnetic interference exists. Necessary servicing must be performed by qualifiedservice engineers ONLY. Clinical ECG examination is suggested for normalexamination of your heart condition. Connect/Disconnect data cable to the device gentlyand carefully. Do NOT wring or shake the data cableplug with force at any time, or the data interface portIII

User Manual for Easy ECG Monitormay be damaged. When the monitor is inoperable due to an overload orsaturation of any part of the amplifier, it will prompt“Contact?” to remind operator. Do not touch the metal electrodes or place theelectrodes of lead wire on patient for ECGmeasurement during USB connected to PC, otherwiseno predictable hazard will be caused by the ECGcombination leakage currents. Dispose of the expired device and its accessoryaccording to applicable local regulations.CAUTION: U.S.FEDERAL LAW RESTRICTS THIS DEVICETO SALE BY OR ON THE ORDER OF A PHYSICIAN.IV

User Manual for Easy ECG MonitorTable of Contents1 OVERVIEW. 11.1 APPEARANCE . 11.2 NAME AND MODEL . 31.3 STRUCTURE . 31.4 FEATURES . 31.5 SCOPE OF APPLICATION . 41.6 WORKING ENVIRONMENT . 52 INSTALLATION OF BATTERIES . 63 OPERATION . 73.1 CAUTIONS . 73.2 START TO MEASURE ECG . 73.3 OPERATION OF MAIN MENU . 184 TECHNICAL PARAMETERS . 354.1 ECG MONITORING . 354.2 DISPLAY . 364.3 POWER . 364.4 DATA MEMORY . 364.5 OVERALL DIMENSION AND WEIGHT . 374.6 CLASSIFICATION . 375 ACCESSORIES . 386 MAINTENANCE AND SERVICE . 39V

User Manual for Easy ECG Monitor6.1 MAINTENANCE . 396.2 CLEANING AND DISINFECTING INSTRUCTION . 407 TROUBLESHOOTING . 418 KEY OF SYMBOLS . 439 APPENDIX. 45VI

User Manual for Easy ECG Monitor1 Overview1.1 Appearance①(Power/Back button): Long time pressing (about2 seconds) powers on or off the device; short timepressing exits the operation, or back to upper levelmenu.1

User Manual for Easy ECG Monitor②(Navigation key): The main operating buttongroup for this device. Press the four directions’ key forfunction selection or parameter setting; press the leftnavigation key for turning on/off the LCD backlightquickly.③Battery cover: Insert/take out the battery byopening/replacing the cover.④(Confirmation key): Press this key for theselection at different levels of menu.⑤ Metal electrode I, II: When measuring, bothelectrodes should be held in the right hand.⑥ Metal electrode III: When measuring, it should beheld in the other hand or on the chest or on the left leg.⑦ LCD screen: Display ECG waveform and differentparameters.⑧ Data interface port: Transmit data memories to thecomputer by this port (optional).⑨ Lead wire socket: the lead wire connector shall beinserted into this socket if the measurement of ECG isdone by lead wires,⑩ Hanging hole: for mounting the hanging cord.2

User Manual for Easy ECG Monitor1.2 Name and ModelName: Easy ECG MonitorModel: PC-80A1.3 StructureThe Easy ECG Monitor comprises main board, LCD,electrodes etc.1.4 FeaturesTwo input connections for ECG signal: Embedded metal electrodes Lead wire connector (optional).Two output modes for ECG data: LCD display Serial output interface.The features are described in detail as follows: Measure ECG easily at any time. ECG waveform and measurement results aredisplayed clearly on the LCD screen. 8 types of arrhythmia can be analyzed andpresented. 30-second records each ECG measurement.3

User Manual for Easy ECG Monitor 24 records can be reviewed or sent to PC by aserial port. Shut down automatically while no key is pressedfor about 35 seconds. Data records can be locked to avoid inadvertentdeletion. Two new AAA size batteries can be used for morethan 500 times measurement; the battery powerstatus is indicated on the screen. Small in size and light in weight to carry. Easy to operate, especially suitable for old people.1.5 Scope of ApplicationThis Easy ECG Monitor is intended for measuring andrecording the ECG signal and average heart rate of adultpatient. It is applicable for use in clinics and home, andconvenient to operate by the patients themselves.This device is not an ECG monitor used in clinicalinstitution or hospital, but is used for spot-check purposeONLY. It can not be used to substitute the normal ECGexamination or real time monitoring. The measurementresults are useful reference for doctor, but do not makedirect diagnostic or analytical decision based on theinformation provided by this device.4

User Manual for Easy ECG Monitor1.6 Working EnvironmentOperating temperature: 5 C 40 COperating humidity:30% 80%Atmospheric pressure: 70 106kPa5

User Manual for Easy ECG Monitor2 Installation of batteries1. Open the battery cover (see the figure above).2. Insert two AAA size batteries directly.3. Close the battery cover. Two new AAA size batteries can be used for morethan 500 times measurement. The battery power status is indicated on the screen.Do NOT insert batteries with their polaritiesreversed. Dispose of the used batteries according toapplicable local regulations.6

User Manual for Easy ECG Monitor3 Operation3.1 CautionsIn order to get a more credible measuring result, youshould pay attention to the following advice Do NOT engage in vigorous exercise prior totesting. Keep you relax while testing. Do NOT move and talk, try to keep as still aspossible during testing.3.2 Start to measure ECG(1) Press the (Power/Back) button (Lasting for 2seconds) to turn on the device.After boot-strap, the device starts to initialize, anddisplays the following pictures.(1)7

User Manual for Easy ECG Monitor(2)(3)(2) After initialization, the device will enter into apreparing phase; this phase will last 6 seconds (seethe following figure), during this period the devicewill judge whether the placed position is befitting.Please adjust your hand-hold position to obtain goodECG signal if the detected waveform is not smooth.8

User Manual for Easy ECG Monitor(3) During the preparing phase, the user can use 4methods to measure ECG: Measure through both hands, name as palmmeasurement. Measure through right hand and chest, name as chestmeasurement. Measure through right hand and left leg, name as legmeasurement. Measure through ECG lead wire, name as lead wiremeasurement.Note: when you use one of methods to measure, if youcan’t obtain enough intensity of ECG signal to analyze,please try to use another measuring method.9

User Manual for Easy ECG MonitorI Palm measurementHold the device with right hand. Make sure that the palmand fingers touch the metal electrode I and II reliably.Press the electrode III close against the center of the leftpalm. The ECG signal detected by Palm measurement isequivalent to Lead I ECG signal.Improper operating method:A.Waving your both hands at will.B.Both hands loose contact with electrodes duringmeasurement.C.Not palm contacts the electrode.10

User Manual for Easy ECG MonitorII Chest measurementIf palm measurement is not ideal, please adopt chestmeasurement. Hold the device with the right hand. Makesure that the palm and fingers touch the metal electrode Iand II reliably. Place the electrode III on bare skin about5cm away below the left nipple. The ECG signaldetected by Chest measurement is similar to Lead VECG signal.Improper operating methodA. The contact place of electrode III deviates from leftchest.B. The holding gesture is incorrect.C. The electrode can not contact bare skin completely.(For example place the electrode on clothing.)11

User Manual for Easy ECG MonitorIII Leg measurementIf palm measurement is not ideal and it is not convenientto use chest measurement, the leg measurement isrecommended. When measuring, hold the device withthe right hand. Make sure that the palm and fingerstouch the metal electrode I and II reliably. Place theelectrode III on the skin 10cm away above the left footankle. The ECG signal detected by Leg measurement isequivalent to Lead II ECG signal.Improper operating method:A. You moved your foot.B. Wrong holding gesture.C. The electrode III can not contact bare skin completely.(For example, place the electrode III on the sleeve oftrousers or socks.)12

User Manual for Easy ECG MonitorIV Lead wire measurementFor the sake of obtaining high quality ECG, lead wiremeasurement can be used. Connect the lead and thedevice effectively by lead wire socket. Place theelectrodes and connect the lead wires as the figure belowto obtain the Lead II ECG signal; if you want to measureLead I and Lead III ECG signal, please connect the leadwires to the electrodes according to the following table.Improper operating method:A. Wrong in connecting lead port.B. Place the electrode in the wrong part of your body.13

User Manual for Easy ECG MonitorElectrodeNameLeadLead I Lead II Lead IIIElectrode LocationThe intersection betweenRARALAthe centerline of the rightclavicle and Rib 2.The intersection betweenLLLARAthe centerline of the leftclavicle and Rib 2.Between the left edge of theLALLLLbreast bone and Rib 5ECG Leads Configuration and Electrodes LocationTable(4) ECG waveform displayAfter the preparation, it enters the phase of ECG. Thefollowing is the detected ECG waveform.14

User Manual for Easy ECG MonitorOn-screen display16:54:01X2:ECGwaveformamplitudescaleCurrent timeThe electrodes contact well.Half size of the nominal scaleX1/2Nominal scaleTwice the size ofthe nominal scaleHeart beat indicator. It flashes whileheart beats.Status of battery powerScale bar for 1mV amplitudeX1X2ECG waveformStatus bar of measurement progress.Note: In order to obtain appropriate waveform display,the amplitude scale of ECG waveform is determined atthe beginning of the measurement automatically ( 2, 1/2, and 1).15

User Manual for Easy ECG Monitor3.2.1 The measuring time lasts about 30 seconds.Then enter the analysis phase in turn.Note: Refer to the period of result description forseveral detailed analysis.16

User Manual for Easy ECG Monitor3.2.2 The analysis result screen can last for 6 seconds.Then will store data automatically and enter into themain menu. To press the power/back button can storethe current measuring result and enter the mainmenu screen too.3.2.3 The system stores 24 pieces of record at most.When the memory is full, this iconwill appear onthe main menu screen.If the system detects that the record is full whenstarting the device, it will automatically delete a pieceof the oldest records for users to measure.17

User Manual for Easy ECG Monitor3.2.4 During the measuring preparation, if poorcontact with the electrodes, the screen displays thefigure as below. Press the power/back button (or waitfor 6 seconds) to enter the main menu screen.3.3 Operation of Main MenuOn the measuring preparation screen or on the analysisresult screen, press the power/back button to enter themain menu screen as figure (1). Meanwhile, press thekeyor of Navigation key to select correspondingtarget (figure (2), (3), (4)).18

User Manual for Easy ECG Monitor(1)(2)(3)19

User Manual for Easy ECG Monitor(4)- the display of battery power.- indicates low voltage, please replace thebatteries with new ones or it will be off automatically.3.3.1 ECG measurementOn the main menu screen, press the key or ofNavigation key to select ECG measurement and pressto measure ECG. Refer to 3.2 for details.3.3.2 The function of waveform review1. On the main menu screen, press the keyor ofNavigation key to select Waveform Review and pressto enter the record menu screen as the figurebelow (The order of the record is that the latest onedisplays in the front of the record).20

User Manual for Easy ECG Monitor2. On the waveform menu screen, press the keyorof Navigation key to select corresponding record andpressto view the corresponding waveform. Pressthe keyorof Navigation key to review thecurrent complete waveform. Press the power/backbutton to return to the previous menu.21

User Manual for Easy ECG MonitorOn-screen displayWaveformX2:ECGwaveformamplitudescaleHR 78Half size of thenominal scaleX1Nominal scaleTwice the size of theX2nominal scaleHeart RateScale bar for 1mV amplitudeX1/2detected ECG waveformStatus bar of measurementprogress.3.3.3 Reviewing the detected result1. On the main menu screen, press the keyor ofNavigation key to select Analysis Result and pressto enter the result menu screen.22

User Manual for Easy ECG Monitor2. On the result menu screen press the keyor ofNavigation key to select the result description andpressto review the corresponding resultintroduction. The following is the detailed description:2.1 No apparent irregularity.2.2 Fast beat.23

User Manual for Easy ECG Monitor2.3 Slow beat.2.4 When find suspicious arrhythmia waveform,through analyzing the measurement results, theconclusions “Suspected early beat”, “Suspectedbigeminy”, “Suspected trigeminy”, “ Suspectedshort run”, “Suspected missing beat” and “May bethe others arrhythmia” are obtained. The followingfigure describes the suspicious early beat.Note:Always consult your doctor for the measurementresults and analysis. Do not attempt self-diagnosis.24

User Manual for Easy ECG Monitor2.5 If the detected signal is too weak, the followingdescription can be seen. Please try again.2.6 On the result description screen press thepower/back button to return to the previous menu.Messages Table12No apparent irregularitySuspected fast beat34Suspected slow beatSuspected early beat5Suspected missing beat67Suspected bigeminySuspected trigeminy8May be the others arrhythmia9Poor signal, try again.Note: DO NOT provide the information for directdiagnostic or analytical decision, but used for referenceONLY.25

User Manual for Easy ECG Monitor3.3.4 Deletion Operation1. On the main menu screen press the keyNavigation key to select delete and pressthe deletion menu screen.or ofto enter2. On the deletion menu screen press the keyorof Navigation key to select a piece of record andpress , the following figure can be seen.26

User Manual for Easy ECG Monitor3. According to the figure above, press, thecorresponding record is deleted. Press any other keycan cancel deleting.4. When pitching on any record on the deletion Menuscreen, press , this record is locked, and can’t bedeleted.5. Unlocking. Choose the locked record and presslock is opened and the icon disappears., the6. Press the power/back button to return to the previousmenu.27

User Manual for Easy ECG Monitor3.3.5 System SettingsOn the main menu screen, press the keyorofNavigation key to select System Settings and presstoenter the system settings screen (1). Then press the keyor to pitch on setting option.(1)(2)(3)28

User Manual for Easy ECG Monitor3.3.5.1 Date/Time Setting1. On the system settings screen, press the keyorof Navigation key to select Date/Time Setting andpressto enter time setup screen (see the followingfigure).2. On the time setting screen, press the keyortomove the cursor for selecting setting option, and thenpress the keyor to set numerical value.3. Press the power/back button to complete the settingand return to the previous menu.29

User Manual for Easy ECG Monitor3.3.5.2 Beeper Setting1. On the system settings screen press the key or ofNavigation key to select Beeper and pressto enterBeeper Setting screen (see the following figure).2. On the beeper setting screen press the key or ofNavigation key, the Beeper Status ON/OFF will beset.3. Press the power/back button to complete saving andreturn to the previous menu.4. Beeper Setting indicates the sound of pressingkeystroke and the sound of heartbeat whenmeasuring.30

User Manual for Easy ECG Monitor3.3.5.3 LCD Setting1. On the system settings screen, press the key orof Navigation key to select Contrast and presstoset the contrast (see the following figure).2. Press the key or of Navigation key, the optionsof 01 to 07 are provided.3. Press the power/back button to complete and save thesetup. At the same time, it returns to the previousmenu.31

User Manual for Easy ECG Monitor3.3.5.4 LCD Backlight Setting1. On the system settings screen press the key or ofNavigation key to choose Backlight Setting andpressto enter Backlight settings screen (see thefollowing figure).2. Press the keybacklight on/off.orof Navigation key to set3. Press the power/back button to complete saving andreturn to the previous menu.4. On the others screen (except Waveform screen andSetting screen) short press the right key of Navigationkey is able to turn on/off the backlight quickly.32

User Manual for Easy ECG Monitor3.3.5.5 Software Version1. On the system settings screen press the keyNavigation key to select Version and pressSoftware Version screen.or ofto enter2. Press the power/back button to return to the previousmenu.3.3.6 Upload ECG Data1. On the system settings screen press the key or ofNavigation key to select Upload ECG data andpressto enter into Upload ECG data screen.33

User Manual for Easy ECG MonitorOn upload data screen, connect the monitor to PC withappropriative data cable, all the records of waveformrecall list and waveform description list can be uploadedto PC for saving after entering into ECG Viewer System.Refer to ECG Viewer Manager User Manual for details.Notes: During uploading, the keys “ ” and power keyare effective. After upload, turn off the device andrestart it again to start measurement. Keep the monitor and data cable still duringuploading data. Unplug the data cable plugs on thehalfway of uploading data is not recommended. If the battery power is not enough, please changethe batteries, or uploading operation will beinterrupted.This function is optional, and needs software support(ECG Viewer Manager System).In addition, anappropriative data cable supplied by Creative is required.34

User Manual for Easy ECG Monitor4 Technical Parameters4.1 ECG Monitoring The number of channels: single (differential input). Electrodes: 3 embedded metal electrodes or using3 adhesive ECG electrodes by connection to thelead wire. Body measurement: the detecting ECG signal canbe the following lead selection:Lead I (between right hand and left hand);Chest Lead (between right hand and chest);Lead II (between right hand and left leg). ECG bandwidth: 0.5Hz 40Hz Internal noise level: 30μVp-p Heart rate measuring range: 30bpm 240bpm Heart rate measuring accuracy: 2bpm or 2%,whichever is greater Display sensitivity: 5.0mm/mV 10% Common-mode rejection ratio (CMRR): 60dB Waveform sweeping speed: 20mm/s 10% Input loop current: 0.1μA Input impedance: 5MΩ35

User Manual for Easy ECG Monitor4.2 Display 128 64dot-matrix single color LCD Display dimension: 49mm 25mm Display of waveform: 128 48 pixels Backlight: LED backlight.4.3 PowerPower: 2 AAA alkaline batteriesSupply voltage range: 2.7 3.3VDCMax average working current: 50mA (backlight off)Auto power-off: 35s 5s (without operation)Low battery indication: 2.5VDC 0.2VDC4.4 Data Memory Data memory of up to about 30 seconds of ECGfor every piece of record. Ability to Store 24 pieces of record at most. Able to transfer data records to the computer bydata interface port. (Optional)36

User Manual for Easy ECG Monitor4.5 Overall Dimension and WeightDimensions: (L)115 (W)58 (H)18 mmWeight: 100g (including 2 AAA batteries)4.6 Classification The type of protection against electric shock:Internally powered equipment The degree of protection against electric shock:Type B applied part The degree of protection against harmfulingress of liquid:Ordinary equipment without protection againstingress of water. Electro-Magnetic Compatibility:Group I, Class B37

User Manual for Easy ECG Monitor5 AccessoriesA lanyardTwo batteries (AAA)A pouchA User ManualA Data Cable (optional)An ECG lead wire (optional)Note: The accessories are subject to change. Detaileditems and quantity see the Packing List.38

User Manual for Easy ECG Monitor6 Maintenance and Service6.1 Maintenance Please change the batteries, when low-voltage iconappears. It is recommended to use soft cloth with medicalalcohol to wipe the metal electrode, when themeasuring signal is too weak. Please wipe the surface of device with clean clothor let it dry in air. Please take out the batteries if the device will notbe used for a long time. Storage and transportationAmbient temperature:-20 to 60ºCRelative humidity:10% 95%Atmospheric pressure:50 107.4kPaThis monitor should be transported by land (vehicle orrailway) or air in accordance with the contractual terms.Do not hit or drop it with force and protect it fromsunlight and rain.39

User Manual for Easy ECG MonitorNotes Keep it away from corrosive substances, explosivesubstances, high/low temperature and moisture. If the device gets wet or has water steam, pleasestop operating it. When it is taken from cold environment to warmand humid environment, please DO NOT use itimmediately. DO NOT operate the button on front panel withsharp materials. DO NOT immerge the device into liquid. Do NOTspray any liquid on it directly. The life of this device is 5 years. In order to ensureits long service life, please pay attention to the useof maintenance.6.2 Cleaning and Disinfecting InstructionSurface-clean sensor with a soft gauze by saturating witha solution such as 75% isopropyl alcohol, if low-leveldisinfection is required, use a 1:10 bleach solution. Thensurface-clean with a damp cloth and dry with a piece ofcloth.Caution: Do not sterilize by irradiation steam, or ethyleneoxide.40

User Manual for Easy ECG Monitor7 Troubleshooting TroubleThe device can not be turned on.Possible Reason1. The batteries are drainedor almost drained.2. The batteries are notinserted properly.3. The device is broken out.Solution1. Change batteries.2. Reinstallbatteries.3. Please contactthe local servicecenter. TroubleThe device is failure in measuring heart rate.Possible Reason1. Electrode and bodydoes not contact well.2. Moved whenmeasuring.3. Electromagneticinterference.4. Weak signal.Solution1. Place the electrodecorrectly.2. When measuring,please keep quiet andavoid moving.3. Keep away frominterference source.4. Change a place (chest)to measure.41

User Manual for Easy ECG Monitor TroubleECG waveform excursion or strong irrelevant waveformPossible ReasonSolution1. Clean skin with soap and waterand eliminate skin crumb and1. Skin is dry or oily.grease to make the skin wet and2. Electrode andgreaseless.body does not2. Press the electrode with acontact tightly.certain force.3. Muscle strain.3. Keep hand relax whilemeasuring.42

User Manual for Easy ECG Monitor8 Key of SymbolsSymbolDescriptionType B applied partLead wire socketWarning See User ManualHeart Rate(Unit: beats per minute)Battery IndicatorConfirmation buttonPower/Back buttonRecord lock iconElectrode contracts well.Memory is full.Data Interface PortSerial numberCE markManufacturer(including address)43

User Manual for Easy ECG MonitorDate of manufactureAuthorised representative in theEuropean communityDisposal of this device according toWEEE regulations44

User Manual for Easy ECG Monitor9 AppendixGeneral Knowledge of ECG1 Normal sinus rhythm: In sinus conditions, SA nodepaces the heart with the regularly rate and the normalrhythm. The heart rate is in the range of 60 to 100times per minute and the rhythm is regular. P wave isnormal and each one is followed by a QRS wave. P-Rinterval: 0.12 0.20s; QRS wave: 0.06 0.10s; Noectopic ECG activity. The display of “No apparentirregularity” may be regarded as Normal sinus rhythm.Symptom: Sinus rhythm, heart rate: 60 100bpmIndication: Normal2 Several abnormal ECG waveforms1) Tachycardia: The heartbeat becomes faster.According to the rhythm of the heart, the averagerhythm beats faster than 100 times a minute. Thedisplay of “fast beat” may be doubted as Tachycardia.Symptom: heart rate 100bpm45

User Manual for Easy ECG MonitorIndication: It may occur with the normal people whohave these physiology conditions: rage, fatigue,smoking, drinking too much wine, excessive coffeeand strong tea, etc.Pathology: anemia, Hyperthyroidism, blood hypoxia,myocarditis, hypokalemia, fever, influence of somemedication (such as atropine, epinephrine etc.).Suggestion: If it belongs to pathology condition,please go to hospital.2) Bradycardia: The heartbeat becomes more slowly.According to the rhythm of the heart, the averagerhythm is less than 60 beats per minute. The displayof “slow beat” may be suspected Bradycardia.Symptom: heart rate 60bpmIndication: It occurs when healthy people fall asleep,and it can be found in athlete (or those who lovedoing sports frequently), old people, or vagusexcitement person.Pathology: Sick sinus syndrome, Ischemic heartdisease,Cardiomyopathy, intracranial hypertension,increased hypokalemia, Low temperature, period ofconvalescence of acute infectious disease or after use46

User Manual for Easy ECG Monitorsome medicines such as digitalis.Suggestion:If it belongs to the pathology condition,please see a doctor.3) Premature beat:In a normal beat a QRS waveformappears prematurely and is followed by a comebackpause.What is premature beats?It is premature contraction.The normal heart beat is always regularly, and eachbeat interval is also symmetrically; Premature beat isout of this general regulation, the heart beatsprematurely and followed by a long time interval; thephenomena of a premature beat between two heartbeats is called inserting premature beat. The displayof “early beat” may be suspected Premature beat.According to different original positions,it can bedivided into:Premature atrial Contraction, PrematureNodal Contraction and Premature VentricularContraction (PVC), which needed to be judged byexperts.Symptom: The heart beat is arrhythmic, it occurs asthe phenomenon that the heart suddenly beats andthen stops for a while. Some people have palpitationor have no symptom.47

User Manual for Easy ECG MonitorIndication: The premature beat can be foundoccasionally in healthy people, with no distinctsymptom or sometimes with palpitation. This may becaused by fatigue, anxiety, insomnia, smoking toomuch, or drinking too much wine, coffee, strong teaetc. It can be recovered without treatment. However,if premature beat occurs frequently, continuously or inmulti-focus, it will indicate functional cardiovasculardisease. please see a doctor as soon as possible。Suggestion: The character of premature beat needsprofessional to confirm, so please save the temporalfigure in time. When seeing a doctor , you can recall itto doctor to judge the character of premature( Premature Atrial Contraction, Premature NodalContraction, Premature Ventricular Contraction ormulti-focus premature beat)and help cure.Testee’s Symptom: Normal heart beat is fol

User Manual for Easy ECG Monitor 7 7 3 Operation 3.1 Cautions In order to get a more credible measuring result, you should pay attention to the following advice Do NOT engage in vigorous exercise prior to testing. Keep you relax while testing. Do NOT move and talk, try to keep as still as