ENIntelect Mobile 2ElectrotherapyUser Manual

2CONTENTSENFOREWORD 4POWERCORD 15INTENDED USER PROFILE 4STIM SET INCLUDES 15INTENDED ENVIRONMENT FOR USE 4US SET INCLUDES 15INTENDED USE 4HEAD TO CART FIXATION 16PRECAUTIONARY INSTRUCTIONS 4CONNECTING CABLES AND INSERTING PLUGS 17GENERAL TERMINOLOGY 5INITIAL RECEIPT 17SYSTEM SOFTWARE SYMBOLS 5IF UNIT SUPPLIED WITH OPTIONAL BATTERY 17DESCRIPTION OF DEVICE MARKINGS 6IFU DOWNLOAD 17VACUUM MODULE MARKINGS 6DEVICE CONNECTED TO THE MAINS 18DEVICE WORKING ON BATTERY 18ELECTROTHERAPY INDICATIONS 7INDICATIONS 7DEVICE LIGHT INDICATORS CONTRAINDICATIONS 7FRONT PANEL INDICATORS 19ADDITIONAL PRECAUTIONS 8ON/OFF BUTTON BLUE INDICATOR: 19ADVERSE EFFECTS 8PLAY/PAUSE BUTTON BLUE INDICATOR 19PRODUCT DESCRIPTION 9SYSTEM SPECIFICATIONS AND DIMENSIONS 20COMPONENTS 9POWER 20HEAD 9ELECTRO STIMULATION SPECIFICATIONS 20CART 9VACUUM SPECIFICATIONS 20BATTERY MODULE (OPTIONAL) 9GENERAL SYSTEM OPERATING AND STORAGEVACUUM MODULE (OPTIONAL) 9OPERATOR INTERFACE 10VACUUM MODULE OPERATOR INTERFACE 11INTELECT MOBILE 2 SET COMPONENTS 15HEAD 15LEADWIRES 15INTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL19TEMPERATURE 21WAVEFORMS 22ELECTROTHERAPY PATIENT PREPARATION ANDELECTRODE PLACEMENT 25VACUUM ELECTRODE PREPARATION (OPTIONAL) 26

3CONTENTSDEVICE USER INTERFACE SCREEN DESCRIPTION EN28REPLACEMENT ACCESSORIES 6628ELECTRODES 6631ELECTROTHERAPY ACCESSORIES 67PRINT SCREEN FUNCTION 32GENERAL ACCESSORIES 67HOME SCREEN 32BATTERY 67TREATMENT REVIEW SCREEN 33VACUUM ACCESSORIES 68GUIDELINES SCREEN 35CLEANING THE INTELECT MOBILE 2 69ELECTROTHERAPY OPERATION 36VACUUM MODULE CLEANING 69START TREATMENT 38CALIBRATION REQUIREMENTS 70VACUUM OPERATION 41DEVICE DISPOSAL 70SPS (SUGGESTED PARAMETER SETUP) 44INSTRUCTION FOR SOFTWARE UPGRADE 70TREATMENT DATA 48IFU DOWNLOAD 70CUSTOM PROTOCOLS 53INSTALLATION OF BATTERY 70SHORT CUTS 57REPLACEMENT BATTERY 70UNASSIGN SHORT CUT 59WARRANTY REPAIR/OUT OF WARRANTY REPAIR CLINICAL RESOURCES 60WARRANTY 72MODALITY/WAVEFORM DESCRIPTIONS 63ELECTROMAGNETIC COMPATIBILITY (EMC) SETTINGS TROUBLESHOOTING 65INTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL7173

4INTRODUCTIONENFOREWORDINTENDED USEThis manual is intended for users of Intelect Mobile 2 . ItThe Intelect Mobile 2 device will be used to deliver a varietycontains general information on operation, precautionaryof modalities to the patient, Ultrasound and electricalpractices, and maintenance.stimulation delivered either as stand alone therapies or inIn order to maximize use, efficiency, and the life of thecombination.system, please read this manual thoroughly and becomefamiliar with the controls, as well as the accessories beforeoperating the system.PRECAUTIONARY INSTRUCTIONSIn addition to the above information, this manual containsThe precautionary instructions found in this section andcare and installation instructions for the optional Cart andthroughout this manual are indicated by specific symbols.Vacuum module for the users of the Intelect Mobile 2.Understand these symbols and their definitions before operating this equipment. The definition of these symbolsBefore administering any treatment to a patient, theare as follows:users of this equipment should read, understand, andfollow the information contained in this manual for eachmode of treatment available, as well as the indications,contraindications, cautions, warnings, and dangers. Consultother resources for additional information regarding theapplication of electrotherapy and ultrasound.INTENDED USER PROFILECAUTIONText with a “CAUTION” indicator explains possible safetyinfractions that have potential to cause minor or moderateinjury or damage to the equipment.WARNINGText with a “WARNING” indicator explains possible safetyThe intended user of this device is a licensed medicalinfractions that will potentially cause serious injury andprofessional. The user should be able to:equipment damage. Read and understand the operator’s manual,warnings, cautions and dangers. Sense auditory and visual signals. Read and understand indications andcontraindications of the deviceINTENDED ENVIRONMENT FOR USEThe device is intended to be operated in the clinic andremote treatment locations. The intended clinical conditionsfor use are a typical clinic setting including chiropracticclinics, physical therapist clinics, athletic training rooms orother rehabilitation settings. The patients home will also bea frequent use setting where the clinician treats the patientin his/her own home environment.INTELECT MOBILE 2 ELECTROTHERAPY USER MANUALDANGERText with a “DANGER” indicator will explain possible safetyinfractions that are imminently hazardous situations thatwould result in death or serious injury.NOTE: Throughout this manual, “NOTE” indicatorsprovide helpful information regarding the particular areaof function being described.

5INTRODUCTIONENGENERAL TERMINOLOGYThe following are definitions for the terminology used throughout this manual . Study these terms to become familiar withthem for ease of system operation and control functionality of the Intelect Mobile 2 .SYSTEM SOFTWARE SYMBOLSHomeRun againBack to previous screenExitSettingsExportIndicates a USB Flash Drive is InsertedImportIndicates Battery LevelDeleteIndicates more content can be viewed by swipingverticallyIndicates more content can be viewed by swipinghorizontallyDelete allStop treatmentIndicates more content can be viewed by scrollingStimClose window / exit full screenShortcutConfirmSPS (Suggested Parameter Setup)Save DataCustom ProtocolsEditTreatment DataGuidelines / Assign toClinical ResourcesPain informationINTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL

6INTRODUCTIONENDESCRIPTION OF DEVICE MARKINGSThe markings on the unit are assurance of its conformityto the highest applicable standards of medical equipmentsafety and electromagnetic compatibility and conform toISO 7010 and ISO15-223-1 One or more of the followingmarkings may appear on the device:Refer to Instructional Manual BookletAtmospheric Pressure RangeWarning, Caution, or DangerTest agencyElectrical Type BF EquipmentCE Mark of Conformity with notified body numberStimAlternating currentPlayClass II equipmentPauseIP21ON/OFFRadio frequency equipmentManufacturerWEEE Directive conformityDate of manufactureShelf lifeCatalogue numberBatch numberSerial numberMDFragile, handle with careThis end upKeep dryTemperature RangeRelative Humidity RangeVACUUM MODULE MARKINGSChannel 2 device connection to vacuum /- Channel 1 polarity /- Channel 2 PolarityINTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL1Channel 1 device connection to vacuum12211

7INDICATIONS FOR USEENELECTROTHERAPY INDICATIONSINDICATIONSCONTRAINDICATIONSTENSThe Intelect Mobile 2 should NOT be used under the Symptomatic relief of chronic pain Management of post-operative painfollowing conditions: Do not use for symptomatic local pain relief unlessetiology is established or unless a pain syndrome hasNMES Muscle re-education Increasing local blood supply Relaxation of muscle spasms Maintaining/Increasing range of motionbeen diagnosed. Do not use when cancerous lesions are present in thetreatment area. Do not apply stimulation over swollen, infected,inflamed areas or skin eruptions (e.g., phlebitis,thrombophlebitis, varicose veins, etc.). Do not use when patient is suspected or known tohave infectious disease and/or disease where it isadvisable, for general medical purposes, to suppressheat or fevers. Do not place electrode placements to the carotidsinus region (anterior neck) or transcerebrally(through the head). Do not use on pregnant women. Safety has notbeen established for the use of therapeutic electricalstimulation during pregnancy. Do not use Intelect Mobile 2 on patients who haveor have had implantable neurostimulating cardiacdemand pacemakers, ICD, or other implantableelectronic devices. Do not use Intelect Mobile 2 on patients with bodyworn electro mechanical medical devices, i.e. insulinpump. Do not use this system in an MRI or CT environment.The Intelect Mobile 2, its components, andaccessories are not to be present in an MRI or CTenvironment. Do not apply stimulation transthoraticorally or on thechest, the introduction of electrical current into theheart may cause cardiac arrhythmia Do not apply stimulation over the anterior neck ormouth. Severe spasm of the laryngeal and pharyngealmuscles may occur and the contractions may bestrong enough to close the airway or cause difficultyin breathing.INTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL

8INDICATIONS FOR USEENELECTROTHERAPY INDICATIONS (CONTINUED)ADDITIONAL PRECAUTIONS Use caution for patients with suspected or diagnosedheart problems. Use caution for patients with suspected or diagnosedepilepsy. Use caution in the presence of the following:» When there is a tendency to hemorrhagefollowing acute trauma or fractureADVERSE EFFECTS Skin irritation and burns beneath the electrodeshave been reported with the use of powered musclestimulators. The irritation can usually be reducedby using an alternative conductive medium or analternative electrode placement Potential adverse effects with TENS are skin irritationand electrode burns» Following recent surgical procedures whenmuscle contraction may disrupt the healingprocess» Over a menstruating or pregnant uterus» Over areas of the skin that lack normalsensation Powered muscle stimulators should be used onlywith the lead wires and electrodes recommended foruse by the manufacturer. With TENS waveforms, isolated cases of skin irritationmay occur at the site of electrode placementfollowing long-term application. The effective management of pain by TENSwaveforms is highly dependent upon patientselection by a person qualified in pain management.INTELECT MOBILE 2 ELECTROTHERAPY USER MANUALNote: 1. Skin irritation and burns beneath the electrodes can bereduced or avoided by using appropriate electrode size and ensuringoptimal contact quality. If any question arises as to the properelectrode size, consult a licensed practitioner prior to therapy session.2.Some people, with very sensitive skin, may experience rednessunder the electrodes after a session. Generally, this redness is totallyharmless and usually disappears after 10 to 20 minutes. However, donot start another stimulation session on the same area if the rednessis still visible

9DEVICE DESCRIPTIONPRODUCT DESCRIPTIONENBATTERY MODULE (optional)The Intelect Mobile 2 Electrotherapy is an electrotherapysystem used with or without an optional Cart, allowing forthe inclusion of a Vacuum module. This equipment is tobe used only under the prescription and supervision of alicensed medical practitioner.COMPONENTSThroughout these instructions the terms “left” and“right”referring to the machine sides are from the perspective of aBattery is an 18V 3350mAh LiIon rechargeable batteryuser standing in front of the unit.The components of the Intelect Mobile 2 Electrotherapy areshown below.HEADCARTINTELECT MOBILE 2 ELECTROTHERAPY USER MANUALVACUUM MODULE (OPTIONAL)

10DEVICE DESCRIPTIONENOPERATOR INTERFACE4.The Intelect Mobile 2 Electrotherapy Operator InterfaceON/OFF switch (only active whenconnected to the mains)contains all the functions and controls necessary foroperator access to all operator utilities, modalities, and5.Mains power connectorparameters for modification and system set up. Color6.Battery coverDisplay and touch screen7.USB Flash Drive Port8.Magnetic fixation to the cartVacuum cover1.Adjustment dial9.2.Play/pause button10. Device handle3.“On/Off“ button. Press and hold (2 sec) the button toswitch OFF the device.HandleColour Touch Screen“On/Off“ buttonPlay / pause buttonAdjustment dialBattery coverVacuum coverLead wire connectors forelectro stimulationON / OFF switch onlyactive on mains powerUSB port forsoftware upgradeMains powerconnectorINTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL

11DEVICE DESCRIPTIONENVACUUM MODULE OPERATORINTERFACEResevoirChannel 1 device connection to vacuum /- Channel 1 polarityChannel 2 device connection to vacuum1INTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL2 /- Channel 2 Polarity

12GENERAL WARNINGS AND PRECAUTIONSENCAUTION This unit should be operated at 5 C to 40 C and 15% to 90% Relative Humidity. The unit should be transported andstored at -20 C to 60 C and 10% to 90% Relative Humidity. Use of parts or materials other than DJO’s can degrade minimum safety. Connect to this unit only items and equipment that have been specified in this IFU as part of the ME SYSTEM or that havebeen specified as being compatible with the ME SYSTEM. DO NOT disassemble, modify, or remodel the unit or accessories. This may cause unit damage, malfunction, electricalshock, fire, or personal injury. DO NOT permit foreign materials, liquids or cleaning agents to enter the unit, including, but not limited to, inflammables,water, and metallic objects, to prevent unit damage, malfunction, electrical shock, fire, or personal injury. Before each use, inspect Applicator cables, STIM cables and associated connectors. Before each use, inspect Vacuum Electrode Cups and Lead Hoses for cracks and damage which may not allow thevacuum to properly secure the electrodes. Caution should always be exercised with current densities more than 2mA/cm². There are no user-serviceable parts inside the unit. If a malfunction occurs, discontinue use immediately and consultdealer for repair service. In case of device unused with battery embedded, it is recommended to connect the device at least once every 4 monthsto allow battery recharge. For waveforms with a DC component:» Do not shave electrodes application area» Warn the patient that tingling sensation under electrodes is normal and it is not linked to burn risk.» Rinse thoroughly treatment area with tap water immediately after the treatmentINTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL

13GENERAL WARNINGS AND PRECAUTIONSENWARNING This device should be used only under the continued supervision of a physician or licensed practitioner. Contaminated sponges, electrodes, leadwires, and gel can lead to infection. Use of electrode on multiple patients can lead to infection. Do not apply electro stimulation treatment during bath, shower, sauna,. DO NOT operate the Intelect Mobile 2 within the vicinity or environment of an ultrasonic diathermy system. DO NOT operate the Intelect Mobile 2 within the vicinity or environment of any microware and RF shortwavediathermy system. DO NOT operate this unit in an environment where other devices are being used that intentionally radiateelectromagnetic energy in an unshielded manner. Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when electricalstimulation is in use. Simultaneous connection of a PATIENT to a high frequency surgical ME EQUIPMENT may result in burns at the site of theSTIMULATOR electrodes and possible damage to the STIMULATOR. Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the IntelectMobile 2, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipmentcould result. Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipmentcould result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and resultin improper operation. Battery replacement by inadequately trained personnel could result in fire or explosion. Please read carefully the batteryreplacement instructions in the Mobile 2 IFU before attempting to replace the battery. Device is designed to comply with electromagnetic safety standards. This equipment generates, uses, and can radiateradio frequency energy and, if not installed and used in accordance with instructions for use, may cause harmfulinterference to other devices in the vicinity. Harmful interference to other devices can be determined by turning thisequipment on and off. Try to correct the interference using one or more of the following:» Reorient or relocate the receiving device» Increase the separation between the equipment» Connect the equipment to an outlet on a different circuit from that to which the other device(s) are connected» Consult your authorized DJO dealer for help. Disconnect the system from the power source before attempting any maintenance, installation, removal, or replacementprocedures to prevent electrical shock and possible damage to system. The Intelect Mobile 2 may be susceptible to Electro-Static Discharge (ESD) at greater than 6 kV when first graspingthe Ultrasound applicator. In the event of such a discharge, the Intelect Mobile 2 may display a permanent error. TheIntelect Mobile 2 will terminate all active outputs (stim, ultrasound,), automatically place the unit in a safe state.INTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL

14GENERAL WARNINGS AND PRECAUTIONSENDANGER DO NOT connect the unit to an electrical supply without first verifying that the power supply is the correct voltage.Incorrect voltage may cause unit damage, malfunction, electrical shock, fire, or personal injury. Your unit was constructedto operate only on the electrical voltage specified on the Voltage Rating and Serial Number Plate. Contact your DJOdealer if the unit is not properly rated. Device is not designed to be used in oxygen rich environment, Explosion hazard if the device is used in the presence offlammable anesthetic mixture with air, oxygen, or nitrous oxide.INTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL

15DETAIL DEVICE DESCRIPTIONINTELECT MOBILE 2 SET COMPONENTSENSET INCLUDES:The components of the Intelect Mobile 2 set are shown15-0133below.79967Carbon electrodes70010STIM lead wiresINTELECT MOBILE 215-1200Intelect Mobile 2 Ultrasound INTL Set EU Plug6522055Chattanooga straps15-1201Intelect Mobile 2 Ultrasound INTL Set All Plug42198Electrodes gel15-1202Intelect Mobile 2 Stim INTL Set EU Plug15-0144/46/47Power cord15-1203Intelect Mobile 2 Stim INTL Set All Plug13-1604Printed Quick Start Guide15-1204Intelect Mobile 2 INTL Set EU Plug15-01425 CM2 Ultrasound Applicator15-1205Intelect Mobile 2 INTL Set All Plug4248Ultrasound Gel Bottle15-1140USB DriveSTIM SET INCLUDES:HEADLEADWIRESThe available leadwires are shown below. If the user ordersa Mobile 2 Stim or Mobile 2 , the box will include the blueand green leadwiresPOWERCORD15-0144Wall Power Cable 2m Black EU15-0146Wall Power Cable 2m Black UK15-0147Power Cable 2m Black AUSINTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL15-0132INTELECT MOBILE 2 STIM79967Carbon electrodes70010STIM lead wires6522055Chattanooga straps42198Electrodes gel15-0144/46/47Power cord13-1604Printed Quick Start Guide15-1140USB DriveUS SET INCLUDES:15-0131INTELECT MOBILE 2 ULTRASOUND15-0144/46/47Power cord13-1604Printed Quick Start Guide15-01425 CM2 Ultrasound Applicator4248Ultrasound Gel Bottle15-1140USB Drive

16SETUP INSTRUCTIONSHEAD TO CART FIXATIONThe optional Therapy System Cart, is designed for use withthe Intelect Mobile 2 only and allows the user to easilytransport the System from patient to patient within theclinic as well as store all necessary accessories, supplies, andapplicators used for the various modalities of the System.The fixation of the head to the cart is magnetic.Remove the Intelect Mobile 2 device and cart from theshipping carton. Visually inspect for damage. Report anydamage to the carrier immediately.To assemble the Mobile 2 Head to the Cart, follow thesesteps:1. Insert device front bottom on the cart lip2. Release device back gently on the cart. Magnets will helpto position the device correctly on the cart top.INTELECT MOBILE 2 ELECTROTHERAPY USER MANUALEN

17SETUP INSTRUCTIONSCONNECTING CABLES AND INSERTINGPLUGSEN1. The Initialisation screen below will be shown for a fewseconds whilst the device starts.When inserting the plugs, be sure to align the flat side of theplug with the flat side of the slot and push in gently. This isto avoid bending the pins in the plug.Insert cable into the appropriate connector prior to startingtherapy.INITIAL RECEIPTRemove all packaging2. The first setup screen will be dispalyed after this allowingIF UNIT SUPPLIED WITH OPTIONALBATTERYthe user to set language, device name and time.After unpacking Intelect Mobile 2 to fit the battery followthe following steps1. Unscrew the battery cover from the base of the device byremoving the 2 screws see below2. Remove the battery cover3. Plug the battery into the battery connector on the device4. Insert the battery into its location5. Replace the 2 screws to close the battery cover3. Click on "Continue" button to go to home screenIFU DOWNLOAD1. Go to the Chattanooga websitewww.chattanoogarehab.com2. Go to Intelect Mobile 2 product tab3. Complete the registration form to be informed about newproduct software version availability and IFU updates.4. Go to documents tab5. Click on the latest version of your Intelect Mobile 2 devicePOWERING UP THE DEVICEFirst time use always use mains power even if batteryconnected. Insert the power cord into the back of the unit,insert the plug into a power outlet, do not position theIntelect Mobile 2 in such a way that makes it difficult todisconnect from the mains power.Switch device on with ON/OFF switch switch on the back ofthe unitINTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL(, US or STIM) User manual to downloadNota: a pdf viewer is required to display IFU

18SETUP INSTRUCTIONSDEVICE CONNECTED TO THE MAINS1. Plug the Power cord into the back of device. Plug the otherend of the cord into an electrical outlet.NOTE: The Power Cord may be unplugged from the back ofthe unit in an emergency situation.2. Turn on the ON/OFF switch located on the back of thedevice.3. Select desired function on the Home ScreenENSTOP TREATMENT AND TURN OFF THEDEVICEPress Play/pause button to pause treatment then press stopon touch screen. If device is on mains power press the on/off button on the front panel then turn off the switch on theback of the unit.If device is working on battery follow the above procedurebut to switch off only press the on/off button on the frontpanelPLAY/PAUSE buttonDEVICE WORKING ON BATTERY1. Press the ON/OFF button on the LCD Front panel, asshown below2. Select desired function on the Home Screen (shownbelow).ON/OFF ButtonINTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL

19SETUP INSTRUCTIONSDEVICE LIGHT INDICATORSIntelect Mobile 2 ELECTROTHERAPY has several lightindicators:FRONT PANEL INDICATORS:1. Colors: Dark blue indicator around ElectrostimulationChannel 1 Green indicator around Electrostimulation Channel 22. Behaviour: Steady when modality is selected and output is notactive Flashing when output is active Quickly flashing when treatment is interrupted anduser action is requestedON/OFF BUTTON BLUE INDICATOR: steady ON from device connection to the mains Flashing while powering ON/OFFPLAY/PAUSE BUTTON BLUE INDICATOR: It flashes when user can start/resume a treatment.Otherwise, steady.INTELECT MOBILE 2 ELECTROTHERAPY USER MANUALEN

20SETUP INSTRUCTIONSENCOMMUNICATIONSTo prepare for communication with the INTELECT mobile2press the settings button and scroll down the screen and pressthe Data transfer button to begin the Bluetooth connection.1. You should now see the screen Waiting for connection4. When connected in the App we will see the device namewhilst the device discovers the computer to be paired withand type and the treatment data/ protocols available forexport to the computer2. If there is one or more active channels on the devicedelivering treatment an error message will be displayed5. Treatment data will be displayed by identifier, selecting ansaying Active channel please stop before data transferitem displays the treatmnet data associated with thaty iD3. The device will give a numerical key ;to complete pairing6. Custom protocols will be listed by name and can bewith computer put the key into the computer to completeexported to the computer and imported from the computerconnection. The device screen will indicate Connected andto the device. Please note treatment data can only bedata transfer via te Windows 10 App can begin.exported and not imported.INTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL

21SYSTEMENSYSTEM SPECIFICATIONS AND DIMENSIONSWidthDepthHeightWeight (no battery)25.5cm35.5cm15cm2.9kgCart (Safe working load 6.5kg)48cm (MAX)52cm (MAX)96cm10.1kgCart with vacuum48cm (MAX)52cm (MAX)96cm11.5kgDevice on cart--111 cm-Intelect Mobile 2 Head UnitSTIMCart configurationsPOWERInputVACUUM SPECIFICATIONS100 - 240 V AC, 1.0 to 0.42 A, 50/60 HzElectrical ClassMode of OperationCLASS IIContinuousNote: Mains isolation is achieved by use of the double pole switchlocated on the rear panel.Electrical Type (Degree of Protection)Electrotherapy.TYPE BFElectrotherapy Vacuum.TYPE BFPowerInput20-25 Vdc, maximum peak current 4AElectrical TypeTYPE BFGeneral characteristicsVacuum Range.0 to 600 mbar maximum /- 5%Vacuum ModesContinuous or Pulsed Continuous.10 setting over vacuum range,60 mbar per setting, 10 mbar to 10 mbar per settingELECTRO STIMULATION SPECIFICATIONSOutput specifications are described for each waveform fromPulsed Modepages 24-26.Maximum Vacuum settings 2 to 10, 10mbar to -10mbarUnless otherwise specified, electrotherapy controls accuracyper settingis: 20 %.Minimum Vacuum settings in 1 to 9, 10mbar toLoad impedance: 500-1000 Ohm-10mbar per settingCC constant current, effect of load impedance on voltageHold Time in minimum & maximum vacuum settings,CV constant voltage, effect of load impedance on current0-20 seconds, in 1 second steps, /-0.5 secondsINTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL

22SYSTEMENGENERAL SYSTEM OPERATING ANDSTORAGE TEMPERATUREIPXX Rating for UnitRated to IP21IP2* Protection against fingers or other object not greaterOperating Conditionsthan 80mm in length and 12mm in diameterThe device will meet its requirement under the followingconditions:Temperature:Relative Humidity:Atmospheric Pressure:5 C to 40 C15% to 90%70kPa to 106kPa*1 Protection from vertically dripping waterIPXX Rating for US applicatorRated to IPX7IPX7 Protection from immersed in water (up to 1m depth)Transport and Storage ConditionsREDThe device will remain in proper condition under theRF transmitter/receiver characteristics:following conditions:- Frequency Band transmission:Temperature:Relative Humidity:Atmospheric Pressure:-20 C to 60 C10% to 90%50kPa to 106kPaTime required for the Intelect Mobile 2 to warm from theminimum storage temperature between uses until theIntelect Mobile 2 is ready for its INTENDED USE when theambient temperature is 20 C: 5hTime required for the Intelect Mobile 2 to cool from themaximum storage temperature between uses until theIntelect Mobile 2 is ready for its INTENDED USE when theambient temperature is 20 C: 5hINTELECT MOBILE 2 ELECTROTHERAPY USER MANUAL- Modulation type:- Data rate:2400–2483.5 MHzGFSKup to 2Mbps 500kHz deviation at 2Mbps- Effective radiated power: 6dBm

23SYSTEMENWAVEFORMSAdvice on size and type of electrodes to be used is given indevice GUI treatment guidelines.CC: Constant CurrentCV: Constant VoltageTENS- Asymmetrical BiphasicIFC (Interferential) Traditional (4 Pole)Advice on size and type of electrodes to be used is given in device GUIAdvice on size and type of electrodes to be used is given in device GUI"treatment guidelines" feature"treatment guidelines" featureInterferential Current is a medium frequency waveform. Current isThe Asymmetrical Biphasic waveform has a short pulse duration. It isdistributed through two channels (four electrodes). The currents crosscapable of strong stimulation of the nerve fibers in the skin as well as ofeach other in the body at the area requiring treatment. The two currentsmuscle tissue. This waveform is often used in TENS devices.Because ofinterfere with each other at this crossing point, resulting in a modulationits short pulse, the patient typically tolerates the current well, even atof the intensity (the current intensity increases and decreases at arelatively high intensities.regular frequency).Output ModeOutput ModeAvailable on ChannelElectrodes1&2,Treatment Time1-60 MinutesMode SelectionCCOutput IntensityBeat FrequencyCarrier FrequencyCycle TimeSweep TimeSweep Low Beat FrequencySweep High Beat FrequencyScan PercentageIRMSDC component0-100 mA (CC)1-200 Hz2000-10,000 HzContinuous or User DefinedOutput IntensityAvailable on ChannelElectrodes0-140 mA (CC) 0-140 V (CV)1,2Treatment Time (Stim)1-60 minutesTreatment Time (Combo)1-30 minutesMode Selection (Stim)Mode Selection (Combo)Amplitude ModulationBurst FrequencyCC or CVCV0% (off) to 100% on 10% steps0-10 bps14 secCycle TimeContinuous or User Defined1-199 HzFrequency1-200 pps2-200 HzStatic, 40%, 100%, Manual0-78mANoFrequencySweepPhase DurationSweep time14 secSweep Low Frequency1-199 ppsSweep High Frequency2-200 ppsIRMSDC componentINTELECT MOBILE 2 ELECTROTHERAPY USER MANUALOn/Off30-400 µsec0-50mANo

24SYSTEMENWAVEFORMS (CONTINUED)TENS - HANTENS- Symmetrical BiphasicAdvice on size and type of electrodes to be used is given in device GUIAdvice on size and type of electrodes to be used is given in device GUI"treatment guidelines" feature"treatment guidelines" featureThe Symmetrical Biphasic waveform has a short pulse duration and isThe HAN Waveform provides optimal parameters with a preciselycapable of strong stimulation of nerve fibers in the skin and in muscle.controlled sequence of Dense-and-Disperse (DD) modes ofThis waveform is often used in portable muscle stimulation units, andstimulation where a burst of 8 pulses at 80Hz is alternating withsome TENS devices.continuous stimulation (no burst), each lasting for 3 seconds. ThisOutput ModeAvailable on ChannelTreatment Time (Stim)Treatment Time (Combo)Mode Selection (Stim)Mode Selection (Combo)Output IntensityAmplitude ModulationBurst Freque

Do not use Intelect Mobile 2 on patients who have or have had implantable neurostimulating cardiac demand pacemakers, ICD, or other implantable electronic devices. Do not use Intelect Mobile 2 on patients with body worn electro mechanical medical devices, i.e. insulin pump. Do not use this system in an MRI or CT environment.