Invacare Platinum MobilePOC1–100B, POC1–100B-CenThis manual MUST be given to the user of the product.BEFORE using this product, read this manual and save for future reference.Oxygen ConcentratorUser Manual

2016 Invacare CorporationAll rights reserved. Republication, duplication or modification in whole or in part is prohibited withoutprior written permission from Invacare. Trademarks are identified by and . All trademarks areowned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.Platinum is a registered trademark in the U.S.A., Europe and Australia.Making Life’s Experiences Possible is a registered trademark in the U.S.A.Pine-sol, Lestoil and Clorox are registered trademarks of The Clorox Company.Dawn is a registered trademark of The Proctor and Gamble Company.

Contents1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.1 Quick Start Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.2 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.3 Indications For Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.4 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.5 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.1 Control Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.2 Input/Output Connections . . . . . . . . . . . . . . . . . . . . .1.6.3 Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.4 Power Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.5 Carry Bag. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.7 Optional Accessories and Replacement Parts. . . . . . . . . . .559111111121213131314142 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.1 Label Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.2 General Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.3 Radio Frequency Interference . . . . . . . . . . . . . . . . . . . . . .2.4 Traveling with the Concentrator. . . . . . . . . . . . . . . . . . . .15151620213 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.2 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.3 Powering the Oxygen Concentrator . . . . . . . . . . . . . . . . .3.3.1 Installing the Battery(s) . . . . . . . . . . . . . . . . . . . . . . . .3.3.2 Charging the Battery(s) . . . . . . . . . . . . . . . . . . . . . . . .3.3.3 Reading the Battery Gauge for Installed andUninstalled Battery(s) . . . . . . . . . . . . . . . . . . . . . . . .3.3.4 Removing the Battery(s) . . . . . . . . . . . . . . . . . . . . . . .3.3.5 External AC Power. . . . . . . . . . . . . . . . . . . . . . . . . . .3.3.6 External DC Power . . . . . . . . . . . . . . . . . . . . . . . . . .232323232324252627273.4 Carry Bag Setup. . . . . . . . . . . . . . . . . . . .3.4.1 Installing the Carry Bag . . . . . . . . . . . .3.4.2 Removing the Carry Bag . . . . . . . . . . .3.4.3 Installing Shoulder Straps. . . . . . . . . . .3.5 Connecting/Positioning the Nasal Cannula .28282929304 Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.1 Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.2 Concentrator Operating Position . . . . . . . . . . . . . . . . . . .4.3 Using the Carry Bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.4 Turning On the Concentrator . . . . . . . . . . . . . . . . . . . . .4.5 Turning Off the Concentrator . . . . . . . . . . . . . . . . . . . . .4.6 Warm Up Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.7 Breathing with the Use of the Concentrator . . . . . . . . . . .4.8 Reading the Display Screen. . . . . . . . . . . . . . . . . . . . . . . .4.9 Adjusting the Pulse Flow Setting . . . . . . . . . . . . . . . . . . . .4.10 Battery Life and Management . . . . . . . . . . . . . . . . . . . . .32323232333334343435365 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.2 Service Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.3 Cleaning the Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.4 Cleaning the Air Intake Filter Screen . . . . . . . . . . . . . . . . .5.5 Cleaning the Carry Bag . . . . . . . . . . . . . . . . . . . . . . . . . .5.6 Cleaning and Disinfection Between Patients . . . . . . . . . . . .5.7 Viewing Hour Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.8 Preventative Maintenance Checklist . . . . . . . . . . . . . . . . .3737373738383940416 After Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6.1 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6.3 Wear and Tear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .434343447 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 457.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 457.2 Alarm Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

7.2.17.2.2Operational Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . 47Shut Down Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . 558 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8.1 Technical Description . . . . . . . . . . . . . . . . . . . . . . . . . . .8.2 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8.3 Regulatory Listing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8.4 Electromagnetic Compliance (EMC) . . . . . . . . . . . . . . . . .60606064659 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 699.1 North American Limited Warranty . . . . . . . . . . . . . . . . . . 69

General1 General1.1Quick Start GuideIMPORTANT!– The Quick Start Guide is for reference ONLY. It isimperative to read the manual it its ENTIRETY forfull instructions.DANGER!Risk of Death, Injury, from Electric ShockTo reduce the risk of burns, electrocution, death orinjury to persons:– DO NOT disassemble. Refer servicing to qualifiedservice personnel. There are no user serviceableparts.WARNING!Risk of Injury or DamageUse of this product outside of the intended use andspecifications has not been tested and may lead toproduct damage, loss of product function, or injury.– DO NOT use this product in any way other thandescribed in the specifications and intended usesections of this manual.1194969-A-00DANGER!Risk of Death, Injury Or Damage From FireTextiles, oil or petroleum substances, grease, greasysubstances and other combustibles are easily ignitedand burn with great intensity in oxygen enriched air andwhen in contact with oxygen under pressure. Smokingduring oxygen therapy is dangerous and is likely to resultin burns or death. To avoid fire, death, injury or damage:– DO NOT SMOKE while using this device. DO NOTuse near OPEN FLAME or IGNITION SOURCES.– NO SMOKING signs should be prominently displayed.– Avoid creation of any spark near oxygen equipment.This includes sparks from static electricity created byany type of friction.– Use only oxygen compatible water-based lotionsor salves before and during oxygen therapy. Toverify, refer to the lotion/salve container for oxygencompatible water-based statement. If necessary,contact the manufacturer. DO NOT use anylubricants on concentrator unless recommended byInvacare.– Keep the oxygen tubing, cord, AC adapter, andconcentrator out from under such items as blankets,bed coverings, chair cushions, clothing, and away fromheated or hot surfaces including space heaters, stoves,and similar electrical appliances.– Make sure concentrator is off when not in use.5

Invacare Platinum MobileDANGER!Risk Of Death, Injury Or DamageImproper use of the product may cause death, injury ordamage. This section contains important information forthe safe operation and use of this product.– DO NOT use this product or any available optionalequipment without first completely reading andunderstanding these instructions and any additionalinstructional material such as user manuals, servicemanuals or instruction sheets supplied with thisproduct or optional equipment.– If you are unable to understand the warnings, cautionsor instructions, contact a health care professional,dealer or technical personnel before attempting touse this equipment.– Check ALL external components and carton fordamage. In case of damage, or if the product is notworking correctly, contact a technician or Invacarefor repair.– This product is intended to be used by adults onlyafter reading and understanding the instructions andwarnings of this user manual.– THE INFORMATION IN THIS DOCUMENT ISSUBJECT TO CHANGE WITHOUT NOTICE.61194969-A-00

General1.Open the battery compartment door. With battery fuel gaugelabel facing toward the front of the oxygen concentrator, slowlylower battery A straight down into one of the two slots in thebattery compartment.Refer to Installing the Battery(s) in Setup.2.Connect the AC Power Adapter to the external power inputconnector B. Plug the AC cord into the electrical outlet.Refer to External AC Power in Setup.Using the concentrator for the first time requiresthe battery(s) to be charged. Refer to Charging theBattery(s) in Setup.3.Connect the nasal cannula to the concentrator oxygen outletport on your concentrator C. Oxygen will only be deliveredwhile you are breathing through the nasal cannula connectedto this device.Refer to Connecting/Positioning the Nasal Cannula in Setup.1194969-A-007

Invacare Platinum Mobile4.Turn on the concentrator by pressing the On/Off Button D untilthe control panel indicators light up.Refer to Turning on the Concentrator in Usage.5.Adjust the pulse flow setting to the one prescribed by yourclinical professional. The pulse flow setting can be changed bypressing the flow selection button E. Place the nasal cannulaover your ears and position the prongs in your nose as instructedby your health care provider.Refer to Adjusting the Pulse Flow Setting in Usage.81194969-A-00

General1.2SymbolsSignal words are used in this manual and apply to hazards or unsafepractices which could result in personal injury or property damage.Refer to the information below for definitions of the signal words.DANGER!– Danger indicates an imminently hazardous situationwhich, if not avoided, will result in death or seriousinjury.WARNING!– Warning indicates a potentially hazardous situationwhich, if not avoided, could result in death or seriousinjury.CAUTION!– Caution indicates a potentially hazardous situationwhich, if not avoided, may result in property damageor minor injury or both.IMPORTANT!– Important indicates a hazardous situation that couldresult in damage to property if it is not avoided.Gives useful tips, recommendations and information forefficient, trouble-free use.Refer to the following table for the meaning of symbols marked onthe equipment and/or packaging.General Warning SignThe background color inside the triangle isyellow on product labels.Follow Instructions for UseThe color of the symbol background is blueon product labels.No open flameThe color of the circle with diagonal bar isred on product labels.DO NOT SmokeThe color of the circle with diagonal bar isred on product labels.Consult Instructions for UseThis symbol is located on the cannulapacking.Direct CurrentType BF Applied PartClass II Equipment1194969-A-009

Invacare Platinum MobilePower On/OffDO NOT use oil or greaseData Output TerminalKeep dryDO NOT dispose of in household wasteProtected against solid foreign objects of 12.5 mmdiameter and greater.ManufacturerProtected against vertically falling water drops whenenclosure tilted up to 15 .Date of ManufactureProtected against solid foreign objects of 12.5 mmdiameter and greater.Protected against vertically falling water drops.RecycleTransport and Storage TemperatureSerial NumberTransport and Storage HumidityDO NOT Reuse (Single Patient Use Only)This symbol refers to the cannula which isfor single patient use only.Electrical Safety Agency Certification Mark74103Latex FreeIndoor Use ONLYThis symbol is included on the AC poweradapter.101194969-A-00

General1.3Indications For UseThe Invacare Platinum Mobile Oxygen Concentrator is intendedto provide supplemental oxygen to patients with respiratorydisorders. The device can be used in a home, institution, vehicle, orother environments outside the home.1.4Intended UseThe Invacare Platinum Mobile Oxygen Concentrator is intendedto provide supplemental oxygen to patients with respiratorydisorders. The Invacare Platinum Mobile Oxygen Concentratorcan be used in a home, institution, vehicle, or other environmentsoutside the home. The device is not intended to be life-supporting orlife-sustaining.WARNING!Risk of Injury or DamageUse of this product outside of the intended use andspecifications has not been tested and may lead toproduct damage, loss of product function, or personalinjury.– DO NOT use this product in any way other thandescribed in the specifications and intended usesections of this k of InjuryThis product is to be used as an oxygen supplement andis not intended to be life-supporting or life-sustaining.ONLY use this product if the patient is capable ofspontaneous breath, able to inhale and exhale withoutthe use of a machine– DO NOT use in parallel or series with other oxygenconcentrators or oxygen therapy devices.WARNING!Risk of Minor Injury or DiscomfortThe conserving, or pulse dose, oxygen delivery techniqueused by this device is contraindicated in persons whosebreathing during normal resting would be unable totrigger the device.– Proper device triggering, setup and operation mustbe confirmed by an experienced clinician or otherrespiratory professional.– Not for pediatric use.– Not for use by tracheotomized patients.11

Invacare Platinum Mobile1.61.6.1FeaturesControl PanelAudio Off ButtonDPress to mute the audible signal (beep) for both alarmsand status indications. The blue indicator light next tothe button illuminates when activated. Audio off can beactivated during or prior to an alarm condition occurring.Display ScreenEDisplays information about the operating status of theconcentrator.Alarm IndicatorFThe yellow indicator light at the center of the triangularsymbol illuminates to indicate abnormal operatingconditions. Refer to Alarm Conditions in Troubleshootingfor more details.Battery GaugeAOn/Off ButtonGPress for one second to turn concentrator on or off.Flow Selection ButtonBCPress to adjust the pulsed oxygen flow to your prescribedsetting. The selected flow setting is shown on the displayscreen. Refer to Adjusting the Pulse Flow Setting in Usagefor more details.Display Backlight ButtonPress once to illuminate the display screen for 5 seconds.Indicates the charge status of the battery(s). Refer toReading the Battery Gauge for Installed and UninstalledBattery(s) in Setup for more details.External Power IndicatorHIThe orange indicator light next to the plug symbolilluminates when an external power source is connectedto the concentrator.Battery Compartment DoorUsed to access the removable battery(s).NotShown12Audible Signal (Beep)Indicates a change in operating status or a conditionwhich requires the operator’s attention.1194969-A-00

General1.6.2Input/Output ConnectionsTo clean the air intake filter, refer to Cleaning the Air Intake FilterScreen in Maintenance.To ensure the concentrator is properly installed in the carry bag,refer to Installing the Carry Bag in Setup.1.6.4Power OptionsWARNING!Risk of Injury or DamageTo avoid injury or damage which will void warranty:– Use only Invacare specified power supplies.AOxygen Outlet PortUsed to connect the nasal cannula to the concentrator.BExternal Power Input ConnectorUsed to connect an external power source to theconcentrator.CUSB PortUsed for service only.1.6.3FiltersAir enters the concentrator through an air intake filter located on thecarry bag. This filter prevents hair and other large particles in the airfrom entering the unit. Before you operate the concentrator, makesure this filter is clean and dry, and the concentrator is properlyinstalled in the carry bag.1194969-A-00Battery(s): Up to two rechargeable batteries can be installed in thebattery compartment of the concentrator. When fully charged, asingle battery supplies power for up to five hours (up to ten hoursfor two batteries). Visible and audible alarm signals occur whenthe battery power is getting low. Refer to Alarm Conditions inTroubleshooting and Charging the Battery(s) in Setup.AC Power Adapter: An AC power adapter allows theconcentrator to be connected to a 100–240 volt 50–60 hertz outlet.Use of the AC power adapter will allow the concentrator to operateand simultaneously recharge the installed battery. Refer to Chargingthe Battery(s) in Setup.DC Power Cable: A DC power cable allows the concentrator tobe connected to an automobile’s (boat, motor home, etc.) 12-voltDC outlet. Use of the DC power cable will allow the concentratorto operate and simultaneously recharge an installed battery(s). Referto Charging the Battery(s) in Setup.13

Invacare Platinum Mobile1.6.5Carry BagThe carry bag provides a convenient means to hold the concentratorwhile performing daily activities.It can be configured as a messenger bag or backpack using theshoulder straps provided. Refer to Installing the Carry Bag in Setup.1.7 Optional Accessories and ReplacementPartsThe following optional accessories are available from Invacare: Additional batteries, model number: POC1–110External battery charger with power supply: POC1–115The following replacement parts are available from Invacare: Battery, model number: POC1–110Carry Bag, model number: POC1–150DC power cable, model number: POC1–140AC power adapter with power cord: POC1–130These parts are base models. Please contact Invacare oryour provider for country specific models.141194969-A-00

Safety2 Safety2.1Label LocationsFollow Ins truc tionsFor Us eDO NOTS MOKENO OP ENFLAMERe c yc leTyp e BFApplie d Pa rtCla s s IIEquipm e n t1187525 A 07In va c a re Co rp o ra tio nwww .invacare.comUSAOne Invacare W a yElyria, Ohio USA44035-2125Te l: 440-329-6000800-333-6900Te c h n ic a l S e rvic e sTe l: 440-329-6593800-832-4707Platinum MobileOxygen ConcentratorModel # POC1-100BManufacturer:Invacare Corporation2101 E. Lake Mary BlvdSanford, FL 32773Te l: 407-321-5630O2 % 87% - 95.6%Co n fo rm s to AAMI S TD ES 60601-1, IS O S TDS 80601-2-67, 80601-2-69, IEC S TDS 60601-1-8, 60601-1-11, 60601-1-6Ce rtifie d to CS A S TD C22.2 No . 60601-174103DANGER - RIS K OF FIRENO S MOKING - Ke e p ALL s o u rc e s o f ig n itio n o u t o f th e a re a in wh ic h th is p ro d u c t is lo c a te d a n d a wa y fro m a re a swh e re o xyg e n is b e in g d e live re d .Te xtile s , o ils a n d o th e r c o m b u s tib le s a re e a s ily ig n ite d a n d b u rn with g re a t in te n s ity in o xyg e n e n ric h e d a ir.Th e b a tte ry u s e d in th is d e vic e m a y p re s e n t a ris k o f fire . DO NOT d is a s s e m b le , in c in e ra te , s h o rt te rm in a ls o r h e a ta b o ve 60 C (140 F).DANGER - RIS K OF ELECTRIC S HOCKDO NOT d is a s s e m b le . Re fe r s e rvic in g to q u a lifie d s e rvic e p e rs o n n e l. No u s e r s e rvic e a b le p a rts in s id e .WARNING - RIS K OF INJ URYTh e d e vic e is n o t in te n d e d to b e life -s u p p o rtin g o r life -s u s ta in in g .CAUTION - DO NOT a d ju s t th e flo w s e ttin g u n le s s a c h a n g e h a s b e e n p re s c rib e d b y yo u r p h ys ic ia n o r th e ra p is t.P ro te c te d a g a in s t s o lid fo re ig n o b je c ts o f 12.5m m d ia m e te r a n d g re a te r.P ro te c te d a g a in s t ve rtic a lly fa llin g wa te r d ro p s wh e n e n c lo s u re tilte d u p to 15 .Fo r p re s c rip tio n u s e ONLY.DO NOT u s e o il o r g re a s e .Th e m a n u fa c tu re r o f th is P OC h a s d e te rm in e d th is d e vic e c o n fo rm s to a ll a p p lic a b le FAA re q u ire m e n ts fo r P OCcar ria g e a n d u s e o n b o a rd a irc ra ft.The label is located on the back of the concentrator. The serial number is located on the right side of label shown here.1194969-A-0015

Invacare Platinum Mobile2.2General GuidelinesIn order to ensure the safe installation, assembly and operation of theconcentrator these instructions MUST be followed.DANGER!Risk Of Death, Injury Or DamageImproper use of the product may cause death, injury ordamage. This section contains important information forthe safe operation and use of this product.– DO NOT use this product or any available optionalequipment without first completely reading andunderstanding these instructions and any additionalinstructional material such as user manuals, servicemanuals or instruction sheets supplied with thisproduct or optional equipment.– If you are unable to understand the warnings, cautionsor instructions, contact a health care professional,dealer or technical personnel before attempting touse this equipment.– Check ALL external components and carton fordamage. In case of damage, or if the product is notworking correctly, contact a technician or Invacarefor repair.– This product is intended to be used by adults onlyafter reading and understanding the instructions andwarnings of this user manual.– THE INFORMATION IN THIS DOCUMENT ISSUBJECT TO CHANGE WITHOUT NOTICE.16DANGER!Risk of Death, Injury Or Damage From FireTextiles, oil or petroleum substances, grease, greasysubstances and other combustibles are easily ignitedand burn with great intensity in oxygen enriched air andwhen in contact with oxygen under pressure. Smokingduring oxygen therapy is dangerous and is likely to resultin burns or death. To avoid fire, death, injury or damage:– DO NOT SMOKE while using this device. DO NOTuse near OPEN FLAME or IGNITION SOURCES.– NO SMOKING signs should be prominently displayed.– Keep all open flames, matches, lighted cigarettes,electronic cigarettes or other sources of ignition atleast 10 ft (3 m) away from this concentrator or anyoxygen carrying accessories such as cannulas or tanks.– If you disregard these warnings about the severehazard of oxygen use while you continue to smoke,you must always turn off the concentrator, removethe cannula and then wait ten minutes before smokingor leave the room where either the concentrator orany oxygen carrying accessories such as cannulas ortanks are located.– DO NOT SMOKE while using the concentrator.CAUTION!Federal (statutory) law restricts this device to sale byor on the order of a medical practitioner licensed by agovernmental agency where he/she practices.– ONLY a licensed medical practitioner may order thepurchase or use of this device.1194969-A-00

SafetyDANGER!Risk of Death, Injury Or Damage From FireTextiles, oil or petroleum substances, grease, greasysubstances and other combustibles are easily ignitedand burn with great intensity in oxygen enriched air andwhen in contact with oxygen under pressure. To avoidfire, death, injury or damage:– Avoid creation of any spark near oxygen equipment.This includes sparks from static electricity created byany type of friction.– Use only oxygen compatible water-based lotionsor salves before and during oxygen therapy. Toverify, refer to the lotion/salve container for oxygencompatible water-based statement. If necessary,contact the manufacturer. DO NOT use anylubricants on concentrator unless recommended byInvacare.– Keep the oxygen tubing, cord, AC adapter, andconcentrator out from under such items as blankets,bed coverings, chair cushions, clothing, and away fromheated or hot surfaces including space heaters, stoves,and similar electrical appliances.– Make sure concentrator is off when not in use.1194969-A-00CAUTION!Risk of Minor Injury or DiscomfortIt is important to plan ahead for travel and othersituations in which you may not have access to additionaloxygen or power supplies. To prevent injury fromoxygen deprivation:– Carry supplemental batteries with you.– Stow the concentrator properly during travel. Forproper storage instructions when travelling refer tothe travel guide.WARNING!Risk of Injury or DamageTo prevent injury or damage from misuse:– NEVER leave concentrator unattended whenconnected to power.– Make sure concentrator is off when not in use.– Outdoor use MUST be conducted with internalbattery power only.WARNING!Risk Of Injury Or Damage– Invacare products are specifically designed andmanufactured for use in conjunction with Invacareprovided or Invacare specified accessories. Any otheraccessories have not been tested by Invacare and arenot recommended for use with Invacare products.– No modification of this equipment is allowed.17

Invacare Platinum MobileCAUTION!Risk of Minor Injury or DiscomfortA change in altitude may affect total oxygen available toyou. To prevent oxygen deprivation:– Consult your physician before traveling to higher orlower altitudes to determine if your flow settingsshould be changed.DANGER!Risk of Death, Injury, from Electric ShockTo reduce the risk of burns, electrocution, death orinjury to persons:– DO NOT disassemble. Refer servicing to qualifiedservice personnel. There are no user serviceableparts.– Avoid using while bathing. If continuous usageis required by the physician's prescription, theconcentrator must be located in another room atleast 7 ft (2.5 m) from the bath.– DO NOT come in contact with the concentratorwhile wet.– DO NOT place or store concentrator where it candrop into water or other liquid.– DO NOT reach for concentrator that has fallen intowater. Unplug IMMEDIATELY.– DO NOT block access to power outlet needed tounplug the AC power cord.– DO NOT use frayed or damaged power cords.– DO NOT use AC power adapter if its housing iscracked or separated.18WARNING!Risk of Injury or DeathTo prevent injury or death from product misuse:– Closely supervise when this concentrator is used byimpaired individuals or near children and/or impairedindividuals.– Monitor patients using this device who are unable tohear or see alarms or communicate discomfort.CAUTION!Risk of Minor Injury or DiscomfortWhile Invacare strives to produce the best oxygenconcentrator in the market today, this oxygenconcentrator can fail to produce oxygen due to powerfailure or device malfunction.– ALWAYS have a backup source of oxygen readilyavailable.– In the event the concentrator fails to produce oxygen,the concentrator will briefly alarm signaling the patientto switch to their backup source of oxygen. Refer toTroubleshooting for more detail.– Invacare recommends keeping at least one batteryinstalled in the concentrator even when operatingfrom an external power source.1194969-A-00

SafetyWARNING!Risk of Death, Injury or DamageTo prevent injury or damage from cord misuse:– DO NOT move or relocate concentrator by pullingon the cord.– DO NOT use extension cords with AC poweradapters provided.– Properly store and position electrical cords and/ortubing to prevent a tripping and strangulation hazards.WARNING!Risk of Injury or DamageTo prevent injury or damage during use:– If you feel ill or uncomfortable, or if the concentratordoes not signal an oxygen pulse and you are unableto hear and/or feel the oxygen pulse, consultyour equipment provider and/or your physicianIMMEDIATELY.– For optimum performance, Invacare recommends thateach use of the concentrator be a minimum of 30minutes. Shorter periods of operation may reducemaximum product life.– The concentrator should be used in an uprightposition.– The concentrator cannot be used in conjunction withPAP, Bi-Level, mechanical ventilator or other suchdevices.1194969-A-00CAUTION!Risk of Minor Injury, Discomfort or Damage– Use of this device at an altitude above 10,000 ft (3048m) or outside a temperature of 41 F to 104 F (5 C to40 C) or a relative humidity above 90% is expectedto adversely effect the flowrate and the percentage ofoxygen and consequently the quality of therapy.WARNING!Risk of Injury or DamageAs a safety feature, this appliance may have a polarizedplug (one blade is wider than the other). To avoid injuryor damage from electrical shock:– This plug will fit in a polarized outlet only one way.If the plug does not fit fully in the outlet, reversethe plug. If it still does not fit, contact a qualifiedelectrician.– DO NOT alter the plug in any way.– DO NOT use if the cord or plug are damaged. Call aqualified electrician for repair.19

Invacare Platinum Mobile20CAUTION!Risk of DamageTo prevent damage from temperature variations:– DO NOT operate in temperatures below 41 F (5 C)or above 104 F (40 C).– When your automobile is turned off, disconnectthe car accessory power cord and remove theconcentrator from the automobile. NEVER allowthe concentrator to be stored in a very hot orcold automobile or in other similar, high or low,temperature environments. Refer to Specifications inTechnical Data.2.3Radio Frequency InterferenceWARNING!Risk of Injury or DamageTo prevent damage from liquid ingress:– If the concentrator is not working properly, if ithas been dropped or damaged, or dropped intowater, call equipment provider/qualified technician forexamination and repair.– NEVER drop or insert any object or

The Invacare Platinum Mobile Oxygen Concentrator is intended to provide supplemental oxygen to patients with respiratory disorders. The device can be used in a home, institution, vehicle, or other environments outside the home. 1.4 Intended Use The Invacare Platinum Mobile Oxygen Concentrator is intended