Data Sheet Version 12.00R&S ZVLVector Network AnalyzerSpecificationsZVL dat-sw en 5213-8150-22 v1200 cover.indd 123.01.2019 15:56:34

Version 12.00, January 2019CONTENTSDefinitions. 3Specifications . 4Measurement range . 4Measurement speed . 4Measurement accuracy . 5Effective system data. 6Test port . 7Test port output . 7Test port input . 7Additional front panel connectors . 8Rear panel connectors. 8Spectrum analysis option. 9General data. 13Ordering information . 15Recommended Calibration Kits for R&S ZVL . 16Power sensors supported by R&S ZVL with R&S FSL-K9 . 16Service options . 16Specifications apply under the following conditions: 30 minutes warm-up time at ambient temperature, specified environmentalconditions met, calibration cycle adhered to, and all internal automatic adjustments performed. Unless stated otherwise, specificationsapply to the two test ports and a nominal source power of –10 dBm.Rohde & Schwarz equipment is designed for reliable operation up to an altitude of 3000 m above sea level, and for transport up to analtitude of 4500 m above sea level.2Rohde & Schwarz R&S ZVL Vector Network Analyzer

Version 12.00, January 2019DefinitionsGeneralProduct data applies under the following conditions: Three hours storage at ambient temperature followed by 30 minutes warm-up operationSpecified environmental conditions metRecommended calibration interval adhered toAll internal automatic adjustments performed, if applicableSpecifications with limitsRepresent warranted product performance by means of a range of values for the specified parameter. These specifications aremarked with limiting symbols such as , , , , , or descriptions such as maximum, limit of, minimum. Compliance is ensured bytesting or is derived from the design. Test limits are narrowed by guard bands to take into account measurement uncertainties, driftand aging, if applicable.Specifications without limitsRepresent warranted product performance for the specified parameter. These specifications are not specially marked and representvalues with no or negligible deviations from the given value (e.g. dimensions or resolution of a setting parameter). Compliance isensured by design.Typical data (typ.)Characterizes product performance by means of representative information for the given parameter. When marked with , or as arange, it represents the performance met by approximately 80 % of the instruments at production time. Otherwise, it represents themean value.Nominal values (nom.)Characterize product performance by means of a representative value for the given parameter (e.g. nominal impedance). In contrast totypical data, a statistical evaluation does not take place and the parameter is not tested during production.Measured values (meas.)Characterize expected product performance by means of measurement results gained from individual samples.UncertaintiesRepresent limits of measurement uncertainty for a given measurand. Uncertainty is defined with a coverage factor of 2 and has beencalculated in line with the rules of the Guide to the Expression of Uncertainty in Measurement (GUM), taking into accountenvironmental conditions, aging, wear and tear.Device settings and GUI parameters are indicated as follows: “parameter: value”.Typical data as well as nominal and measured values are not warranted by Rohde & Schwarz.Rohde & Schwarz R&S ZVL Vector Network Analyzer3

Version 12.00, January 2019SpecificationsMeasurement rangeFrequency rangeR&S ZVL3 and R&S ZVL3-75R&S ZVL6R&S ZVL139 kHz to 3 GHz9 kHz to 6 GHz9 kHz to 13.6 GHz(overrange 5 kHz to 15 GHz)Reference frequency, internal, nominalaging per year1 10–6temperature drift 0 C to 50 C1 10–6 with optional R&S FSL-B4 OCXO reference frequencyaging per year1 10–7temperature drift 0 C to 50 C1 10–7Frequency resolution1 HzNumber of measurement pointsdefault value201user-selectable2 to 4001Measurement bandwidths1/2/5 steps10 Hz to 500 kHzDynamic range of the R&S ZVL3,9 kHz to 1 MHz 75 dB, typ. 85 dBR&S ZVL6, and R&S ZVL3-751 MHz to 7 MHz 85 dB, typ. 100 dB7 MHz to 50 MHz 105 dB, typ. 120 dB50 MHz to 3 GHz 115 dB, typ. 123 dB(R&S ZVL3 and R&S ZVL6)50 MHz to 3 GHz (R&S ZVL3-75 only) 110 dB, typ. 120 dB3 GHz to 5 GHz (R&S ZVL6 only) 115 dB, typ. 123 dB5 GHz to 6 GHz (R&S ZVL6 only) 100 dB, typ. 110 dBDynamic range of the R&S ZVL139 kHz to 300 kHz 75 dB, typ. 85 dB300 kHz to 20 MHz 90 dB, typ. 100 dB20 MHz to 10 GHz 100 dB, typ. 105 dB10 GHz to 13.6 GHz 95 dB, typ. 100 dBThe dynamic range is measured as the difference between the –10 dBm source power and the rms value of the data trace of thetransmission magnitude. This magnitude is produced by noise and crosstalk with the test ports short-circuited and the stepattenuators set to 0 dB. This specification applies without system error correction and at 10 Hz measurement bandwidth (filter type:normal) in the temperature range from 18 C to 28 C. The dynamic range is also affected by receiver inherent spurious atparticular frequencies.Measurement speedMeasurement timeR&S ZVL3 and R&S ZVL6R&S ZVL3, R&S ZVL6,and R&S ZVL3-75R&S ZVL13Data transfer time4for S21 with 1.1 GHz center frequency, 200 MHz span, 201 measurements points, anddisplay switched onwith normalization calibrationand 100 kHz measurement bandwidth 50 msand 10 kHz measurement bandwidth 70 mswith full two-port calibrationand 100 kHz measurement bandwidth 60 msand 10 kHz measurement bandwidth 90 mswith full two-port calibrationand 100 kHz measurement bandwidth 100 msand 10 kHz measurement bandwidth 130 msfor 201 measurements pointsvia VX11 over 100 Mbit/s LAN 2.1 msvia RSIB over 100 Mbit/s LAN 1.5 msvia optional R&S FSL-B10 GPIB 4.7 msinterfaceRohde & Schwarz R&S ZVL Vector Network Analyzer

Version 12.00, January 2019Measurement accuracyThis data is valid between 18 C and 28 C, provided the temperature has not varied by more than 1 K after calibration. Validity ofthe data is conditional on the use of a suitable calibration kit. This calibration kit is used to achieve the effective system data specifiedbelow. Frequency points, measurement bandwidth, and sweep time have to be identical for measurement and calibration (nointerpolation allowed).Accuracy of transmissionmeasurements9 kHz to 50 MHzfor 0 dB to –40 dB 0.2 dB or 2 50 MHz to 3 GHzfor 0 dB to –50 dB 0.2 dB or 2 for –50 dB to –70 dB 0.3 dB or 3 3 GHz to 6 GHz (R&S ZVL6 and R&S ZVL13)for 0 dB to –50 dB 0.2 dB or 2 for –50 dB to –70 dB 0.3 dB or 3 6 GHz to 13.6 GHz (R&S ZVL13 only)for 0 dB to –50 dB 0.3 dB or 3 for –50 dB to –70 dB 0.5 dB or 5 Specifications are based on a matched DUT, a measurement bandwidth of 10 Hz (filter type: normal), a step attenuation of 10 dB(default value), and a nominal source power of 0 dBm.Accuracy of reflection measurements9 kHz to 3 GHzfor 0 dB to –15 dB 0.4 dB or 3 for –15 dB to –25 dB 1 dB or 6 for –25 dB to –35 dB 3 dB or 20 9 kHz to 6 GHz (R&S ZVL6 and R&S ZVL13)for 0 dB to –15 dB 0.4 dB or 3 for –15 dB to –25 dB 1 dB or 6 for –25 dB to –35 dB 3 dB or 20 6 GHz to 13.6 GHz (R&S ZVL13 only)for 0 dB to –15 dB 0.4 dB or 3 for –15 dB to –25 dB 1 dB or 6 for –25 dB to –35 dB 3 dB or 20 Specifications are based on a matched DUT, a measurement bandwidth of 10 Hz (filter type: normal), a step attenuation of 10 dB,and a nominal source power of 0 dBm.Trace stabilityTrace noise of S21 (RMS) above 10 MHzat –10 dBm source power, 0 dB transmission, 0 dB step attenuation,and 2 kHz measurement bandwidth (filter type: normal)R&S ZVL3, R&S ZVL6, 0.005 dBand R&S ZVL3-75R&S ZVL13 0.005 dB 0.09 , typ. 0.05 Rohde & Schwarz R&S ZVL Vector Network Analyzer5

Version 12.00, January 2019Effective system dataThis data is valid between 18 C and 28 C, provided the temperature has not varied by more than 1 K after calibration. The data isbased on a measurement bandwidth of 10 Hz (filter type: normal), step attenuation of 0 dB, nominal source power of 0 dBm andsystem error calibration by means of a suitable calibration kit. Frequency points, measurement bandwidth, and sweep time have to beidentical for measurement and calibration (no interpolation allowed).DirectivitySource matchReflection trackingLoad matchTransmission tracking69 kHz to 100 kHz100 kHz to 1 MHz1 MHz to 3 GHz3 GHz to 6 GHz(R&S ZVL6 and R&S ZVL13)6 GHz to 13.6 GHz (R&S ZVL13)9 kHz to 100 kHz100 kHz to 1 MHz1 MHz to 3 GHz3 GHz to 6 GHz(R&S ZVL6 and R&S ZVL13)6 GHz to 13.6 GHz (R&S ZVL13)9 kHz to 100 kHz100 kHz to 1 MHz1 MHz to 3 GHz3 GHz to 6 GHz(R&S ZVL6 and R&S ZVL13)6 GHz to 13.6 GHz (R&S ZVL13)9 kHz to 100 kHz100 kHz to 1 MHz1 MHz to 3 GHz3 GHz to 6 GHz(R&S ZVL6 and R&S ZVL13)6 GHz to 13.6 GHz (R&S ZVL13)9 kHz to 100 kHz100 kHz to 1 MHz1 MHz to 3 GHz3 GHz to 6 GHz(R&S ZVL6 and R&S ZVL13)6 GHz to 13.6 GHz (R&S ZVL13 only)Rohde & Schwarz R&S ZVL Vector Network Analyzer 30 dB, typ. 35 dB 40 dB, typ. 46 dB 46 dB, typ. 50 dB 40 dB, typ. 50 dB 40 dB, typ. 50 dB 30 dB, typ. 35 dB 40 dB, typ. 46 dB 40 dB, typ. 46 dB 36 dB, typ. 40 dB 36 dB, typ. 40 dB 0.15 dB, typ. 0.1 dB 0.1 dB, typ. 0.05 dB 0.1 dB, typ. 0.05 dB 0.1 dB, typ. 0.05 dB 0.1 dB, typ. 0.05 dB 30 dB, typ. 35 dB 40 dB, typ. 46 dB 46 dB, typ. 50 dB 40 dB, typ. 46 dB 40 dB, typ. 46 dB 0.15 dB, typ. 0.05 dB 0.1 dB, typ. 0.05 dB 0.1 dB, typ. 0.05 dB 0.1 dB, typ. 0.05 dB 0.12 dB, typ. 0.05 dB

Version 12.00, January 2019Test portSpecifications apply to the two test ports, i.e. PORT 1 and PORT 2.ImpedanceR&S ZVL3, R&S ZVL6, and R&S ZVL13R&S ZVL3-75Connector50 Ω75 ΩN, femaleTest port outputSpecifications apply to the two test ports, i.e. PORT 1 and PORT 2.Source matchPower rangePower accuracy of the R&S ZVL3 andR&S ZVL6Power accuracy of the R&S ZVL3-75Power accuracy of the R&S ZVL13Power linearityPower resolutionHarmonicsSpurious9 kHz to 3 GHz3 GHz to 6 GHz (R&S ZVL6 only)9 kHz to 13.6 GHz (R&S ZVL13 only)R&S ZVL3, R&S ZVL6, andR&S ZVL3-75R&S ZVL13typ. 14 dBtyp. 14 dBtyp. 10 dB–50 dBm to 0 dBm,typ. –60 dBm to 10 dBm–35 dBm to –5 dBm,typ. –40 dBm to 0 dBmat –10 dBm source power above 10 MHz 2 dBin temperature range 18 C to 28 C 1 dB, typ. 0.3 dBat –10 dBm source power above 10 MHz 2.2 dBin temperature range 18 C to 28 C 1.2 dB, typ. 0.5 dBat –10 dBm source power above 10 MHz 3 dBin temperature range 18 C to 28 C 1.3 dB, typ. 0.5 dBreferenced to –10 dBm and above 10 MHz 2 dBin temperature range 18 C to 28 C 0.8 dB, typ. 0.3 dB0.01 dBat –10 dBm source powertyp. –35 dBcat –10 dBm source powertyp. –40 dBcTest port inputSpecifications apply to the two test ports, i.e. PORT 1 and PORT 2.Load matchAttenuationAttenuation stepsAttenuation uncertaintyMaximum nominal input level1 dB compression pointInherent spurious responseDamage DC voltageDamage CW RF powerDamage pulse voltageDamage pulse energy9 kHz to 3 GHz3 GHz to 6 GHz (R&S ZVL6 only)9 kHz to 13.6 GHz (R&S ZVL13 only )user-selectable9 kHz to 3 GHz3 GHz to 6 GHz(R&S ZVL6 and R&S ZVL13)6 GHz to 13.6 GHz (R&S ZVL13 only)with attenuation set to 0 dBwith attenuation set to 30 dBabove 200 MHz,with attenuation set to 0 dBwithout input signal andwith attenuation set to 0 dB10 µs pulse length10 µs pulse lengthtyp. 14 dBtyp. 14 dBtyp. 10 dB0 dB to 30 dB5 dB 0.3 dB 0.3 dB 0.7 dB–10 dBm 20 dBm0 dBm, nominal –90 dBm, nominal30 V 27 dBm150 V10 mWsRohde & Schwarz R&S ZVL Vector Network Analyzer7

Version 12.00, January 2019Additional front panel connectorsAUX OUTConnectorOutput impedanceOpen-circuit voltageadjustablePROBE POWERDC voltagesDC current3.5 mm mini jack (mono) 100 Ω0 V to 1.5 V 15 V, –12.6 V, and groundmax. 150 mARear panel connectorsLANlocal area network connectorRJ-45, 8 pinsVGAconnector for external monitorVGAEXT TRIGGER/GATE INConnectorInput signalEXT REFConnectorInput frequencyMaximum allowed deviationInput powerInput impedanceOutput frequencyOutput frequency accuracyOutput power8BNC, female, 50 ΩTTL-compatibleinput for external frequency reference signal and, with optional R&S FSL-B4 OCXOreference frequency, alternatively input or output for external frequency reference signalBNC, female, 50 Ω10 MHz1 kHz0 dBm to 10 dBm50 Ωrequires optional R&S FSL-B4 OCXO10 MHzreference frequency50 Hz0 dBm, nominalRohde & Schwarz R&S ZVL Vector Network Analyzer

Version 12.00, January 2019Spectrum analysis optionThe specifications of the R&S ZVL-K1 spectrum analysis option apply to the RF INPUT connector (combined with PORT 2).FrequencyFrequency rangeR&S ZVL3 and R&S ZVL3-75R&S ZVL6R&S ZVL13AttenuationAttenuation steps9 kHz to 3 GHz9 kHz to 6 GHz9 kHz to 13.6 GHz(overrange 5 kHz to 15 GHz)aging per year 1 10–6temperature drift 0 C to 50 C 1 10–6with optional R&S FSL-B4 OCXO reference frequencyaging per year 1 10–7temperature drift 0 C to 50 C 1 10–7user-selectable0 dB to 30 dB5 dBR&S FSL-B22 RF preamplifier optionGainswitchableStatic reference frequency uncertaintyFrequency readoutMarker resolutionUncertaintyFrequency counter resolutionCounter uncertaintyFrequency span0 dB or 20 dB, nominalwith marker or frequency counterS/N 25 dBR&S ZVL3 and R&S ZVL3-75R&S ZVL6R&S ZVL13Frequency span uncertaintySpectral puritySingle sideband (SSB) phase noiseat 500 MHzat a carrier offset of 1 kHzat a carrier offset of 10 kHzat a carrier offset of 100 kHzat a carrier offset of 1 MHzSweep timeSweep times10 Hz span 3.2 kHz3.2 kHz span 1.5 GHz1.5 GHz span 3 GHz3 GHz span 6 GHz6 GHz span 13.6 GHzfor finite spanfor zero spanSweep filtersResolution bandwidths (RBW)Resolution bandwidth uncertaintyResolution filter shape factor 60 dB:3 dBtyp. –95 dBc (1 Hz) –96 dBc (1 Hz),typ. –100 dBc (1 Hz) –96 dBc (1 Hz),typ. –100 dBc (1 Hz) –113 dBc (1 Hz),typ. –116 dBc (1 Hz)2.5 ms to 5 s/Hz span2.5 ms to 16000 s5 ms to 16000 s10 ms to 16000 s25 ms to 16000 s1 µs to 5 µs in steps of 125 ns5 µs to 16000 s in steps of 5 % 3 %, nominal 0.1 %, nominalSweep times for zero spanSweep time uncertaintyspan/501 marker frequency referenceuncertainty 2 % span 10 % resolution bandwidth ½ last digit1 Hz frequency reference uncertainty ½ last digit0 Hz (zero span) and 10 Hz to 3 GHz0 Hz (zero span) and 10 Hz to 6 GHz0 Hz (zero span) and 10 Hz to 13.6 GHz(overrange 15 GHz)3%with optional R&S FSL-B7 narrowresolution filterszero spanRBW 3 MHz300 Hz to 10 MHz (–3 dB) in 1/3/10 steps10 Hz to 10 MHz (–3 dB) in 1/3/10 stepsadditionally 20 MHz (–3 dB) 3 %, nominal 5, nominal (Gaussian filters)Rohde & Schwarz R&S ZVL Vector Network Analyzer9

Version 12.00, January 2019EMI filters6 dB bandwidthswith optional R&S FSL-B7 narrowresolution filtersBandwidth uncertaintyShape factor 60 dB:3 dB 3 %, nominal 6, nominalChannel filtersBandwidthsVideo bandwidthsDemodulation bandwidthLevelDisplay range1 dB compression pointwith optional R&S FSL-B7 narrowresolution filtersone-pole lowpass filtersabove 200 MHz and at 0 dB attenuationIntermodulation of the R&S ZVL3, R&S ZVL6, and R&S ZVL3-75Third-order intermodulation (TOI)intermodulation-free dynamic range,above 10 MHz, level 2 –20 dBm,reference level –10 dBmSecond harmonic intercept (SHI)above 20 MHzIntermodulation of the R&S ZVL13Third-order intermodulation (TOI)Second harmonic intercept (SHI)Displayed average noise level of theR&S ZVL3, R&S ZVL6, andR&S ZVL3-75109 kHz, 120 kHz, and 1 MHz200 Hz, 9 kHz, 120 kHz, and 1 MHzintermodulation-free dynamic range,10 MHz to 6 GHz, level 2 –20 dBm,reference level –10 dBm6 GHz to 13.6 GHz20 MHz to 6.8 GHz300/500 4/15/16/18 (RRC)/20/ 21/24.3 (RRC)/25/30/50/100/150/192/200/300/500 kHz,1/1.228/1.28 (RRC)/1.5/2/3/ 3.84 (RRC)/4.096 (RRC)/ 5 MHz(RRC root raised cosine)100 Hz, additionally 200 Hz1 Hz to 10 MHz in 1/3/10 steps20 MHz, nominaldisplayed noise floor to 20 dBm0 dBm, nominal 50 dBc(TOI 5 dBm, typ. 12 dBm) 30 dBm, nominal 50 dBc(TOI 5 dBm, typ. 12 dBm)50 dBc (TOI 5 dBm), nominal 30 dBm, nominalat 0 dB attenuation, with resolution bandwidth (RBW) 1 kHz,and video bandwidth (VBW) 10 Hz, normalized to 1 Hzwith preamplifier OFF9 kHz to 1 MHz –100 dBm (1 Hz)1 MHz to 10 MHz –110 dBm (1 Hz)10 MHz to 50 MHz –130 dBm (1 Hz)50 MHz to 4 GHz –140 dBm (1 Hz)4 GHz to 6 GHz –135 dBm (1 Hz)with preamplifier ON (requires optional R&S FSL-B22 RF preamplifier)9 kHz to 1 MHz –115 dBm (1 Hz)1 MHz to 10 MHz –125 dBm (1 Hz)10 MHz to 50 MHz –150 dBm (1 Hz)50 MHz to 4 GHz –156 dBm (1 Hz)4 GHz to 6 GHz –151 dBm (1 Hz)500 MHztyp. –163 dBm (1 Hz)1 GHztyp. –163 dBm (1 Hz)3 GHztyp. –162 dBm (1 Hz)6 GHztyp. –156 dBm (1 Hz)Rohde & Schwarz R&S ZVL Vector Network Analyzer

Version 12.00, January 2019Displayed average noise level of theR&S ZVL13Immunity to interferenceImage frequency responseIntermediate frequency responseInherent spurious responseSpurious responseSpurious responseSpurious response, mixer level –10 dBmat 0 dB attenuation, with resolution bandwidth (RBW) 1 kHz,and video bandwidth (VBW) 10 Hz, normalized to 1 Hzwith preamplifier OFF9 kHz to 2 MHz –105 dBm (1 Hz)2 MHz to 10 GHz –125 dBm (1 Hz)10 GHz to 13.6 GHz –120 dBm (1 Hz)with preamplifier ON (requires optional R&S FSL-B22 RF preamplifier) –105 dBm (1 Hz)9 kHz to 2 MHz2 MHz to 20 MHz –125 dBm (1 Hz)20 MHz to 6 GHz –140 dBm (1 Hz)500 MHztyp. –151 dBm (1 Hz)1 GHztyp. –151 dBm (1 Hz)2 GHztyp. –149 dBm (1 Hz)3 GHztyp. –146 dBm (1 Hz)6 GHztyp. –146 dBm (1 Hz)f 2 48.375 MHzf 2 838.375 MHzf 2 7158.375 MHzat 48.375 MHz, 838.375 MHz, and7158.375 MHzabove 30 MHz, without input signal,at 0 dB attenuation and RBW 1 MHzat 7158.375 MHz – 2 finat fin – 3579.1875 MHzf 6 GHzf 6 GHzLevel displayLogarithmic level axisLinear axisNumber of tracesTrace detectorsNumber of measurement pointsdefault valueuser-selectableTrace functionsSetting range of reference levellogarithmic displayUnits of axislinear displaylogarithmic displaylinear display –60 dBc, typ. –80 dBc –60 dB, typ. –80 dBctyp. –60 dBc –60 dBc, typ. –80 dBc –90 dBmtyp. –60 dBctyp. –60 dBc –50 dBc, typ. –60 dBc –38 dBc, typ. –48 dBc10 dB to 100 dB0 % to 100 % with 10 divisions4max peak, min peak, auto peak, sample,RMS, quasi peak, and average501125 to 32001clear/write, max hold, average, min hold,or view–80 dBm to 20 dBmin steps of 2 dB, 5 dB, or 10 dB–80 dBm to 20 dBm or 0 % to 100 %dBm, dBmV, dBµV, dBµA, or dBpWV, mV, µV, A, mA, µA, W, mW, µW, nW,or pWRohde & Schwarz R&S ZVL Vector Network Analyzer11

Version 12.00, January 2019Level measurement uncertaintyAbsolute uncertaintyFrequency responseAttenuation uncertaintyUncertainty of reference level settingDisplay nonlinearityLogarithmic level displayBandwidth switching uncertainty95 % confidence level, 20 C to 30 C, S/N 16 dB,0 dB to –50 dB from reference levelR&S ZVL3, R&S ZVL6, and R&S ZVL1310 MHz to 3 GHz 0.5 dBR&S ZVL6 and R&S ZVL133 GHz to 6 GHz 0.8 dBR&S ZVL136 GHz to 13.6 GHz 1.2 dBR&S ZVL3-7510 MHz to 3 GHz 0.7 dBat internal calibration frequency of 65.833 MHzR&S ZVL3, R&S ZVL6 and 0.3 dBR&S ZVL13R&S ZVL3-75 0.5 dB 20 C to 30 C, at –10 dBm input level and 10 dB attenuationR&S ZVL3, R&S ZVL6, and R&S ZVL13up to 10 MHz 0.8 dB, nominal10 MHz to 3 GHz 0.5 dB, typ. 0.3 dBR&S ZVL6 and R&S ZVL133 GHz to 6 GHz 0.8 dB, typ. 0.3 dBR&S ZVL136 GHz to 13.6 GHz 1.2 dB, typ. 0.5 dBR&S ZVL3-75up to 10 MHz 1 dB, nominal10 MHz to 3 GHz 0.7 dB9 kHz to 3 GHz 0.3 dB3 GHz to 6 GHz (R&S ZVL6 and 0.3 dBR&S ZVL13)6 GHz to 13.6 GHz (R&S ZVL13 only) 0.7 dB 0.1 dB, nominalS/N 16 dB0 dB to –50 dBreference: RBW 10 kHzTrigger functionsTrigger sourceExternal trigger signalwith optional R&S FSL-B10 GPIBinterface12 0.1 dB, nominalfree run, video, external, IF powerTTLI/Q dataInterfaceMemory lengthSample rateSignal bandwidth 0.2 dBsample rate 65.8 MHzRohde & Schwarz R&S ZVL Vector Network AnalyzerLANLAN or IEC/IEEE bus (GPIB)max. 512 ksample I and Q10 kHz to 65.8 MHz20 MHz

Version 12.00, January 2019General dataRemote controlLAN interfaceIEC/IEEE bus (GPIB)with optional R&S FSL-B10 GPIBinterfaceDisplayTypeResolutionTemperaturecolor TFT640 480 pixeloperating temperature rangepermissible temperature rangestorage temperature range0 C to 50 C0 C to 55 C–40 C to 70 Cin line with IEC 60068-2-1 andIEC 60068-2-2 40 C at 85 % rel. humidity,in line with IEC 60068-2-30vibration, sinusoidal5 Hz to 150 Hz,in line with IEC 60068-2-610 Hz to 300 Hz,in line with IEC 60068-2-6440 g shock response spectrum,in line with IEC/EN 60068-2-27,MIL-STD-810Damp heatMechanical resistancevibration, randomshockEMC, RF emission10/100BaseTSCPI 1997.0In line with EN 61000-6-4; operation inresidential, commercial, and businessareas or in small-size companies is notcovered. The instrument may not beoperated in residential, commercial, andbusiness areas or in small-size companiesunless additional measures are taken toensure that EN 61000-6-3 is compliedwith.In line with CISPR 11/EN 55011 group 1class A (for a shielded test set-up). Theinstrument complies with the emissionrequirements stipulated by EN 55011class A. This means that the instrument issuitable for use in industrial environments.EMC, other emissions, and immunityin line with IEC/EN 61326,emission class B (residentialenvironment),immunity: industrial environment(excluding operating frequency)SafetyIEC 61010-1, EN 61010-1, UL 61010B-1,CSA C22.2 No. 1010-1Power supplyAC input voltageAC supply frequencyAC input currentDC power supplyDC input currentPower consumption100 V to 240 V (AC) with tolerance 10 %,safety class I in line with VDE 411for AC input voltages100 V to 120 V120 V to 240 Vrequires R&S FSL-B30 DC power supplyoptionrequires R&S FSL-B30 DC power supplyoptionR&S ZVL3, R&S ZVL6,and R&S ZVL3-75R&S ZVL1350 Hz to 400 Hz with tolerance 5 %50 Hz to 60 Hz with tolerance 5 %0.4 A to 1.2 A10 V to 28 V (DC)2.2 A to 8 A80 W, typ. 60 W,max. 80 W with all options110 W, typ. 85 W,max. 110 W with all optionsRohde & Schwarz R&S ZVL Vector Network Analyzer13

Version 12.00, January 2019Weight and dimensionsDimensions ( W H D )with handleShipping weight408.8 mm 158.1 mm 465.3 mm(16.1 in 6.2 in 18.3 in)342.3 mm 158.1 mm 367.0 mm(13.5 in 6.2 in 14.5 in)7 kg (16 lb)8.4 kg (18.5 lb)14 kg (31 lb)Recommended calibration interval12 monthswithout handleWeight14without optionswith battery pack and all other optionsRohde & Schwarz R&S ZVL Vector Network Analyzer

Version 12.00, January 2019Ordering informationDesignationVector Network Analyzer,3 GHz, 2 ports, 50 ΩVector Network Analyzer,6 GHz, 2 ports, 50 ΩVector Network Analyzer,13.6 GHz, 2 ports, 50 ΩVector Network Analyzer,3 GHz, 2 ports, 75 ΩOptionsOCXO Reference FrequencyAdditional Interfaces 1TV Trigger 1Narrow Resolution Filters 1Gated Sweep 1GPIB InterfaceRF Preamplifier (3GHz/6 GHz) 1DC Power SupplyNiMH Battery Pack 4.5 Ah 2Li-Ion Battery Pack 10 Ah with Battery Charger 2Power Sensor Support 1, 3Spectrum AnalysisDistance-to-Fault MeasurementTime DomainAM/FM/φM Measurement Demodulator 1Spectrogram Measurements 1Cable TV and TV Measurements 1Options for R&S ZVL3, R&S ZVL6 and R&S ZVL13 onlyBluetooth TX Measurements 1(1.1 and 2.0 EDR)Application Firmware for Noise Figure andGain Measurements 1, 43GPP FDD BTS Application Firmware 1WLAN IEEE 802.11a/b/g/j ApplicationFirmware 1Upgrade of R&S FSL-K91 toIEEE 802.11n Application Firmware 1WiMAX IEEE 802.16 OFDM/OFDMAApplication Firmware 1TypeR&S ZVL3Order No.1303.6509.03R&S ZVL61303.6509.06R&S ZVL131303.6509.13R&S ZVL3-751303.6509.75R&S FSL-B4R&S FSL-B5R&S FSL-B6R&S FSL-B7R&S FSL-B8R&S FSL-B10R&S FSL-B22R&S FSL-B30R&S FSL-B31R&S ETL-B235R&S FSL-K9R&S ZVL-K1R&S ZVL-K2R&S ZVL-K3R&S FSL-K7R&S FSL-K14R&S 301.9675.02R&S FSL-K81301.9398.02R&S FSL-K301301.9817.02R&S FSL-K72R&S FSL-K911302.0620.021302.0094.02R&S FSL-K91N1308.7903.02R&S FSL-K931302.0736.02The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Rohde & Schwarz is underlicense.“WiMAX Forum“ is a registered trademark of the WiMAX Forum. “WiMAX,“ the WiMAX Forum logo, “WiMAX Forum Certified,“ and the WiMAX ForumCertified logo are trademarks of the WiMAX Forum.1Requires the R&S ZVL-K1 spectrum analysis option.2Requires the R&S FSL-B30 DC power supply option.3Requires the R&S NRP-Z3/-Z4 or R&S FSL-B5 additional interfaces option.4Requires the R&S FSL-B5 additional interfaces option and a preamplifier.Rohde & Schwarz R&S ZVL Vector Network Analyzer15

Version 12.00, January 2019Recommended Calibration Kits for R&S ZVLDesignationCalibration Kit DC to 3 GHz, N(f), 50 Calibration Kit DC to 3 GHz, N(f), 75 Calibration Kit DC to 9 GHz, N(m), 50 Calibration Kit DC to 9 GHz, N(f), 50 Calibration Kit DC to 18 GHz, N, 50 Calibration Kit DC to 15 GHz, 3.5 mm, male, 50 Calibration Kit DC to 15 GHz, 3.5 mm, female, 50 TypeR&S ZCANR&S ZCANR&S ZV-Z170R&S ZV-Z170R&S ZV-Z270R&S ZV-Z135R&S ZV-Z135Order 35011.6536.021317.7677.021317.7677.03Power sensors supported by R&S ZVL with R&S FSL-K9DesignationAverage Power Sensor, 10 MHz to 8 GHz, 200 mWAverage Power Sensor, 10 MHz to 18 GHz, 200 mWAverage Power Sensor, 10 MHz to 18 GHz, 2 WAverage Power Sensor, 10 MHz to 18 GHz, 15 WAverage Power Sensor, 10 MHz to 18 GHz, 30 WPower Sensor Module with Power Splitter,DC to 18 GHz, 500 mWPower Sensor Module with Power Splitter,DC to 26.5 GHz, 500 mWAverage Power Sensor, 9 kHz to 6 GHz, 200 mWThermal Power Sensor, 0 Hz to 18 GHz, 100 mWThermal Power Sensor, 0 Hz to 40 GHz, 100 mWWideband Power Sensor, 50 MHz to 18 GHz, 100 mWTypeR&S NRP-Z11R&S NRP-Z21R&S NRP-Z22R&S NRP-Z23R&S NRP-Z24R&S NRP-Z27Order 21137.8502.021169.4102.02R&S NRP-Z371169.3206.02R&S NRP-Z91R&S NRP-Z51R&S NRP-Z55R&S 09.02Service optionsService optionsExtended warranty, one yearExtended warranty, two yearsExtended warranty, three yearsExtended warranty, four yearsExtended warranty with calibration coverage, one yearExtended warranty with calibration coverage, two yearsExtended warranty with calibration coverage, three yearsExtended warranty with calibration coverage, four yearsExtended warranty with accredited calibration coverage, one yearExtended warranty with accredited calibration coverage, two yearsExtended warranty with accredited calibration coverage, three yearsExtended warranty with accredited calibration coverage, four yearsR&S WE1R&S WE2R&S WE3R&S WE4R&S CW1R&S CW2R&S CW3R&S CW4R&S AW1R&S AW2R&S AW3R&S AW4Please contact your localRohde & Schwarz salesoffice.Extended warranty with a term of one to four years (WE1 to WE4)Repairs carried out during the contract term are free of charge 5. Necessary calibration and adjustments carried out during repairs arealso covered. Simply contact the forwarding agent we name; your product will be picked up free of charge and returned to you in topcondition a couple of days later.Extended warranty with calibration (CW1 to CW4)Enhance your extended warranty by a

Rohde & Schwarz R&S ZVL Vector Network Analyzer 5 Measurement accuracy This data is valid between 18 C and 28 C, provided the temperature has not varied by more than 1 K after calibration. Validity of the data is conditional on the use of a suitable calibration kit. This calibration kit is used to achieve the effective system data specified