MINDRAY BS-200 Instrument Application - BrandSD

Transcription

MINDRAY BS-200Instrument ApplicationTest: TRIGLYCERIDESCatalog #: T7532TEST PARAMETERSTest Name:TRIGLYCERIDESR1 Volume:300No.:*R2 Volume:0Full Name:*Sample Volume:3Standard No.:R1 Blank:Reaction Type:End-pointMixed Reag. Blank:Primary Wavelength:510Linearity Range:Secondary Wavelength:670Linearity Limit:0Direction:AscendingSubstrate Limit:0Reaction Time:0 / 20Factor:0Incubation Time:0 Prozone s:CALIBRATION PARAMETERSRule:Two-point linearSensitivity:Replicates:2Calibrator 1:WaterCalibrator 2:Chem CalCalibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.12/6/2006Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: ALBUMINCatalog #: A7502TEST PARAMETERSTest Name:ALBUMINR1 Volume:300No.:*R2 Volume:0Full Name:*Sample Volume:3Standard No.:R1 Blank:Reaction Type:End-pointMixed Reag. Blank:Primary Wavelength:630Linearity Range:Secondary Wavelength:0Linearity Limit:0Direction:AscendingSubstrate Limit:0Reaction Time:0/4Factor:0Incubation Time:0 Prozone IBRATION PARAMETERSRule:Two-point linearSensitivity:Replicates:2Calibrator 1:WaterCalibrator 2:Chem CalCalibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.12/6/2006Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: ALKALINE PHOSPHATASECatalog #: A7516TEST PARAMETERSTest Name:ALKALINE PHOSPHATASER1 Volume:200No.:*R2 Volume:50Full Name:*Sample Volume:5Standard No.:R1 Blank:Reaction Type:KineticsMixed Reag. Blank:Primary Wavelength:405Linearity Range:Secondary Wavelength:670Linearity Limit:0.2Direction:AscendingSubstrate Limit:0Reaction Time:3 / 11Factor:2276Incubation Time:3 Prozone CALIBRATION PARAMETERSRule:Calibrator 1:Sensitivity:Calibrator 2:Replicates:2Calibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.12/6/2006Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: ALANINE TRANSAMINASECatalog #: A7526TEST PARAMETERSTest Name:ALANINE TRANSAMINASER1 Volume:200No.:*R2 Volume:50Full Name:*Sample Volume:10Standard No.:R1 Blank:Reaction Type:KineticsMixed Reag. Blank:Primary Wavelength:340Linearity Range:Secondary Wavelength:405Linearity Limit:0.2Direction:DescendingSubstrate Limit:0.15Reaction Time:3 / 11Factor:4700Incubation Time:3 Prozone CALIBRATION PARAMETERSRule:Calibrator 1:Sensitivity:Calibrator 2:Replicates:2Calibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.1/30/2008Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: ASPARTATE TRANSAMINASECatalog #: A7561TEST PARAMETERSTest Name:ASPARTATE TRANSAMINASER1 Volume:200No.:*R2 Volume:50Full Name:*Sample Volume:10Standard No.:R1 Blank:Reaction Type:KineticsMixed Reag. Blank:Primary Wavelength:340Linearity Range:Secondary Wavelength:405Linearity Limit:0.2Direction:DescendingSubstrate Limit:0.15Reaction Time:3 / 11Factor:4200Incubation Time:3 Prozone CALIBRATION PARAMETERSRule:Calibrator 1:Sensitivity:Calibrator 2:Replicates:2Calibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.12/6/2006Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: AMYLASECatalog #: A7564TEST PARAMETERSTest Name:AMYLASER1 Volume:200No.:*R2 Volume:0Full Name:*Sample Volume:5Standard No.:R1 Blank:Reaction Type:KineticsMixed Reag. Blank:Primary Wavelength:405Linearity Range:Secondary Wavelength:0Linearity Limit:0.2Direction:AscendingSubstrate Limit:0Reaction Time:3 / 11Factor:3178Incubation Time:0 Prozone CALIBRATION PARAMETERSRule:Calibrator 1:Sensitivity:Calibrator 2:Replicates:2Calibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.3/20/2008Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: UREA NITROGEN (LIQUID)Catalog #: B7552TEST PARAMETERSTest Name:UREA NITROGEN (LIQUID)R1 Volume:300No.:*R2 Volume:75Full Name:*Sample Volume:3Standard No.:R1 Blank:Reaction Type:Fixed-timeMixed Reag. Blank:Primary Wavelength:340Linearity Range:Secondary Wavelength:670Linearity Limit:0Direction:DescendingSubstrate Limit:0Reaction Time:2/7Factor:0Incubation Time:3 Prozone s:CALIBRATION PARAMETERSRule:Two-point linearSensitivity:Replicates:2Calibrator 1:WaterCalibrator 2:Chem CalCalibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.12/6/2006Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: CARBON DIOXIDE (405)Catalog #: C7502TEST PARAMETERSTest Name:CARBON DIOXIDE (405)R1 Volume:300No.:*R2 Volume:0Full Name:*Sample Volume:3Standard No.:R1 Blank:Reaction Type:Fixed-timeMixed Reag. Blank:Primary Wavelength:405Linearity Range:Secondary Wavelength:510Linearity Limit:0Direction:DescendingSubstrate Limit:0Reaction Time:2 / 15Factor:0Incubation Time:0 Prozone bs:CALIBRATION PARAMETERSRule:Two-point linearSensitivity:Replicates:2Calibrator 1:WaterCalibrator 2:Chem CalCalibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.1/22/2008Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: CHOLESTEROLCatalog #: C7510TEST PARAMETERSTest Name:CHOLESTEROLR1 Volume:300No.:*R2 Volume:0Full Name:*Sample Volume:3Standard No.:R1 Blank:Reaction Type:End-pointMixed Reag. Blank:Primary Wavelength:510Linearity Range:Secondary Wavelength:670Linearity Limit:0Direction:AscendingSubstrate Limit:0Reaction Time:0 / 20Factor:0Incubation Time:0 Prozone s:CALIBRATION PARAMETERSRule:Two-point linearSensitivity:Replicates:2Calibrator 1:WaterCalibrator 2:Chem CalCalibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.12/6/2006Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: CREATINE KINASECatalog #: C7522TEST PARAMETERSTest Name:CREATINE KINASER1 Volume:200No.:*R2 Volume:50Full Name:*Sample Volume:6Standard No.:R1 Blank:Reaction Type:KineticsMixed Reag. Blank:Primary Wavelength:340Linearity Range:Secondary Wavelength:405Linearity Limit:0.2Direction:AscendingSubstrate Limit:0Reaction Time:3 / 11Factor:5712Incubation Time:3 Prozone CALIBRATION PARAMETERSRule:Calibrator 1:Sensitivity:Calibrator 2:Replicates:2Calibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.12/6/2006Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: CALCIUM (ARSENAZO)Catalog #: C7529TEST PARAMETERSTest Name:CALCIUM (ARSENAZO)R1 Volume:250No.:*R2 Volume:0Full Name:*Sample Volume:3Reference No.:R1 Blank:Reaction Type:End-pointMixed Reag. Blank:Primary Wavelength:670Linearity Range:Secondary Wavelength:0Linearity Limit:0Direction:AscendingSubstrate Limit:0Reaction Time:0/8Factor:0Incubation Time:0 Prozone LIBRATION PARAMETERSRule:Two-point linearSensitivity:Replicates:2Calibrator 1:WaterCalibrator 2:Chem CalCalibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.12/6/2006Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: CREATININECatalog #: C7539TEST PARAMETERSTest Name:CREATININER1 Volume:200No.:*R2 Volume:40Full Name:*Sample Volume:12Standard No.:R1 Blank:Reaction Type:Fixed-timeMixed Reag. Blank:Primary Wavelength:510Linearity Range:Secondary Wavelength:578Linearity Limit:0Direction:AscendingSubstrate Limit:0Reaction Time:2/7Factor:0Incubation Time:3 Prozone LIBRATION PARAMETERSRule:Two-point linearSensitivity:Replicates:2Calibrator 1:WaterCalibrator 2:Chem CalCalibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.12/6/2006Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: CREATINE KINASE MBCatalog #: C7562TEST PARAMETERSTest Name:CREATINE KINASE MBR1 Volume:300No.:*R2 Volume:0Full Name:*Sample Volume:14Standard No.:R1 Blank:Reaction Type:KineticsMixed Reag. Blank:Primary Wavelength:340Linearity Range:Secondary Wavelength:405Linearity Limit:0.2Direction:AscendingSubstrate Limit:0Reaction Time:18 / 25Factor:6530Incubation Time:0 Prozone CALIBRATION PARAMETERSRule:Calibrator 1:Sensitivity:Calibrator 2:Replicates:2Calibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI CKMB Controls Cat.# C7562-12.5/1/2008Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: CRPCatalog #: C7568TEST PARAMETERSTest Name:CRPR1 Volume:180No.:*R2 Volume:180Full Name:*Sample Volume:7Standard No.:R1 Blank:Reaction Type:End-pointMixed Reag. Blank:Primary Wavelength:578Linearity Range:Secondary Wavelength:0Linearity Limit:0Direction:AscendingSubstrate Limit:0Reaction Time:0 / 18Factor:0Incubation Time:3 Prozone IBRATION PARAMETERSRule:SplineCalibrator 1:WaterCalibrator 2:CRP Cal 1Calibrator 3:CRP Cal 2Interval (day):Calibrator 4:CRP Cal 3Difference Limit:Calibrator 5:CRP Cal 4SD:Calibrator 6:CRP Cal 5Sensitivity:Replicates:1Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.4/08/2008Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: FRUCTOSAMINECatalog #: F7546TEST PARAMETERSTest Name:FRUCTOSAMINER1 Volume:250No.:*R2 Volume:0Full Name:*Sample Volume:14Standard No.:R1 Blank:Reaction Type:Fixed-timeMixed Reag. Blank:Primary Wavelength:546Linearity Range:Secondary Wavelength:670Linearity Limit:0Direction:AscendingSubstrate Limit:0Reaction Time:34 / 38Factor:0Incubation Time:0 Prozone ALIBRATION PARAMETERSRule:Two-point linearSensitivity:Replicates:2Calibrator 1:WaterCalibrator 2:Fruc. Std.Calibrator 3:Interval (day):Calibrator 4:Difference Limit:Calibrator 5:SD:Calibrator 6:Blank Response:Error Limit:Coefficient:0It is recommended that two levels of control material be assayed daily. Reorder PSI Chemistry Controls Cat.# C7590-50 & C7591-50.4/08/2008Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com

MINDRAY BS-200Instrument ApplicationTest: GLUCOSE HEXCatalog #: G7517TEST PARAMETERSTest Name:GLUCOSE HEX (Hitachi)R1 Volume:300No.:*R2 Volume:0Full Name:*Sample Volume:3Standard No.:R1 Blank:Reaction Type:End-pointMixed Reag. Blank:Primary Wavelength:340Linearity Range:Secondary Wavelength:405Linearity Limit:0Direction:AscendingSubstrate Lim

MINDRAY BS-200 Instrument Application Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com Test: UREA NITROGEN (LIQUID) Catalog #: B7552 TEST PARAMETERS Test Name: UREA NITROGEN (LIQUID) R1 Volume: 300 No.: * R2 Volume: 75 Full Name: * Sample Volume: 3 Standard No.: R1 Blank: Reaction Type: Fixed-time Mixed Reag. Blank: