DICOM Conformance Statement For IntelliSpace . - Philips

Transcription

DICOM Conformance StatementApplication Annex:IntelliSpace Collaboration Viewer on PhilipsIntelliSpace Portal V10.0 Koninklijke Philips N.V. 2018All rights are reserved.ICAP-T-030001.09bCorresponds to ICAP-W-030001.02

Issued by:Philips Medical Systems Nederland BV, a Philips Healthcare company,P.O. Box 10.0005680 DA BestThe NetherlandsEmail: dicom@philips.comInternet: www.philips.com/dicomDocument Number: Corresponds to ICAP-W-030001.02

DICOM Conformance Statement Annexure for IntelliSpace Collaboration Viewer on Philips IntelliSpace Portal V10.0Page 3 of 12Table of Contents1.INTRODUCTION . 41.1.REVISION HISTORY . 41.2.TERMINOLOGY . 42.DATA SPECIFICATIONS . 52.1.SUPPORTED IOD’S . 52.1.1. Acceptance Criteria . 52.1.2. Contents of Created IOD’s . 62.1.2.1.Secondary Capture Image Storage SOP Class . 6 2018 Koninklijke Philips N.V.ICAP-T-030001.09bCorresponds to ICAP-W-030001.0228-March-2018

DICOM Conformance Statement Annexure for IntelliSpace Collaboration Viewer on Philips IntelliSpace Portal V10.0Page 4 of 121. IntroductionThis DICOM Conformance Statement annex is applicable to the IntelliSpace Collaboration Viewer which is a zero install Web-Vieweron Philips IntelliSpace Portal V10.0 hosting platform, later referred to as IntelliSpace Collaboration Viewer. In general, the IntelliSpaceCollaboration Viewer is the user environment for reviewing CT, SC, MR, NM and PET images.1.1. Revision HistoryThe revision history below provides dates and differences among individual document versions.Table 1: Revision HistoryDocument VersionDate of al version.1.2. TerminologyDICOMIODUIDVRDigital Imaging and Communications in MedicineInformation Object DefinitionUnique IdentifierValue Representation 2018 Koninklijke Philips N.V.ICAP-T-030001.09bCorresponds to ICAP-W-030001.0228-March-2018

DICOM Conformance Statement Annexure for IntelliSpace Collaboration Viewer on Philips IntelliSpace Portal V10.0Page 5 of 122. Data Specifications2.1. Supported IOD’sThis section specifies each IOD accepted and / or created by IntelliSpace Collaboration Viewer.ACCEPTEDCREATEDThe applicable IOD is accepted for storage in the repository of the hosting platform and supported for import inIntelliSpace Collaboration Viewer for viewing and analysis.The IntelliSpace Collaboration Viewer supports generation of derived data by using the applicable IOD and is able tostore this data in the repository of the hosting platform.Table 2: Supported IOD’sIODSupportNameUIDACCEPTEDCREATEDCT Image Storage SOP Class1.2.840.10008.5.1.4.1.1.2YesNoEnhanced CT Image Storage1.2.840.10008.5.1.4.1.1.2.1YesNoSecondary Capture Image Storage SOP Class1.2.840.10008.5.1.4.1.1.7YesYesMR Image Storage SOP Class1.2.840.10008.5.1.4.1.1.4YesNoNuclear Medicine Image Storage SOP Class1.2.840.10008.5.1.4.1.1.20YesNoPositron Emission Tomography Image Storage SOP nce CriteriaThis section specifies the acceptance criteria applied by IntelliSpace Collaboration Viewer to which a dataset should adhere before itcan be imported into the application.All system models objects are accepted.Table 3: Accepted transfer syntaxes per IODIODNameTransfer e-ELEExplicit VR Little EndianImplicit VR Little EndianJPEG Lossless, NonHierarchical, FOP 0008.5.1.4.1.1.2.1CT-private-ELEExplicit VR Little EndianImplicit VR Little EndianJPEG Lossless, NonHierarchical, FOP .11.2.840.10008.1.21.2.840.10008.1.2.4.70Secondary Capture Image Storage SOP it VR Little EndianImplicit VR Little EndianJPEG Lossless, NonHierarchical, FOP .11.2.840.10008.1.21.2.840.10008.1.2.4.70MR Image Storage SOP it VR Little EndianImplicit VR Little .840.10008.1.21.2.840.10008.1.2.4.70CT Image Storage SOP ClassEnhanced CT Image Storage SOP Class 2018 Koninklijke Philips N.V.ICAP-T-030001.09bCorresponds to ICAP-W-030001.0228-March-2018

DICOM Conformance Statement Annexure for IntelliSpace Collaboration Viewer on Philips IntelliSpace Portal V10.0Page 6 of 12JPEG Lossless, NonHierarchical, FOP (Process14)Nuclear Medicine Image Storage SOP Class1.2.840.10008.5.1.4.1.1.20Positron Emission Tomography Image StorageSOP Class1.2.840.10008.5.1.4.1.1.1282.1.2.Explicit VR Little EndianImplicit VR Little te-ELE1.3.46.670589.33.1.4.1Explicit VR Little Endian1.2.840.10008.1.2.1Implicit VR Little Endian1.2.840.10008.1.2JPEG Lossless, NonHierarchical, FOP (Process14)1.2.840.10008.1.2.4.70Contents of Created IOD’sThis section specifies in detail the attribute contents of created data objects. Attributes are grouped together by its correspondingmodule as specified by DICOM standard. Philips private attributes are excluded for specification.Abbreviations used in the Module table for the column "Presence of Value" are:ALWAYSThe attribute is always present with a valueEMPTYThe attribute is always present without any value (attribute sent zero length)VNAPThe attribute is always present and its Value is Not Always Present (attribute sent zero length if no value is present)ANAPThe attribute is present under specified condition – if present then it will always have a valueThe abbreviations used in the Module table for the column "Source" are:AUTOThe attribute value is generated automaticallyCONFIGThe attribute value source is a configurable parameterCOPYThe attribute value source is another SOP instanceFIXEDThe attribute value is hard-coded in the applicationIMPLICITThe attribute value source is a user-implicit settingMPPSThe attribute value is the same as that use for Modality Performed Procedure StepMWLThe attribute value source is a Modality WorklistUSERThe attribute value source is explicit user input2.1.2.1.Secondary Capture Image Storage SOP ClassTable 4: IOD of Created Secondary Capture Image Storage SOP Class InstancesInformation EntityModulePresence Of ModulePatientPatient ModuleALWAYSStudyGeneral Study ModuleALWAYSStudyPatient Study ModuleCONDITIONALSeriesGeneral Series ModuleALWAYSEquipmentGeneral Equipment ModuleALWAYSEquipmentSC Equipment ModuleALWAYSImageGeneral Image ModuleALWAYSImageImage Pixel ModuleALWAYSImageSC Image ModuleALWAYSImageOverlay Plane ModuleCONDITIONALImageModality LUT ModuleCONDITIONALImageVOI LUT ModuleCONDITIONALImageSOP Common ModuleALWAYSImageExtended DICOM and Private attributesCONDITIONAL 2018 Koninklijke Philips N.V.ICAP-T-030001.09bCorresponds to ICAP-W-030001.0228-March-2018

DICOM Conformance Statement Annexure for IntelliSpace Collaboration Viewer on Philips IntelliSpace Portal V10.0Page 7 of 12Table 5: Patient ModuleSQPresenceof 010,0010PNVNAPCOPY-Patient ID0010,0020LOVNAPCOPY-Patient's Birth Date0010,0030DAVNAPCOPY-Patient's Birth Time0010,0032TMANAPCOPY-Patient's Sex0010,0040CSVNAPCOPY-Other Patient IDs0010,1000LOANAPCOPY-Other Patient Names0010,1001PNANAPCOPY-Ethnic Group0010,2160SHANAPCOPY-Patient Comments0010,4000LTANAPCOPY-Issuer of Patient ID0010,0021LOANAPCOPY-Attribute NameTagVRReferenced Patient Sequence0008,1120 Referenced SOP Class UID0008,1150 Referenced SOP Instance UIDPatient's NameValueSourceCommentTable 6: General Study ModuleDAPresenceof OPY-Referring Physician's Name0008,0090PNVNAPCOPY-Referring PhysicianIdentification Sequence0008,0096SQANAPCOPY-Study Description0008,1030LOANAPCOPY-Procedure Code Sequence0008,1032SQANAPCOPY-Physician(s) of Record0008,1048PNANAPCOPY-Physician(s) of RecordIdentification Sequence0008,1049SQANAPCOPY-Name of Physician(s) ReadingStudy0008,1060PNANAPCOPY-Physician(s) Reading StudyIdentification Sequence0008,1062SQANAPCOPY-Referenced Study Sequence0008,1110SQANAPCOPY-Study Instance UID0020,000DUIALWAYSAUTO-Study ID0020,0010SHVNAPAUTO-Attribute NameTagVRStudy Date0008,0020Study TimeAccession NumberValueSourceCommentTable 7: Patient Study ModuleLOPresenceof ValueANAPCOPY-0008,1084SQANAPCOPY-Patient's Age0010,1010ASANAPCOPY-Patient's Size0010,1020DSANAPCOPY-Patient's PCOPY-Additional Patient History0010,21B0LTANAPCOPY-Attribute NameTagVRAdmitting DiagnosesDescription0008,1080Admitting Diagnoses CodeSequence 2018 Koninklijke Philips N.V.ICAP-T-030001.09bValueCorresponds to ICAP-W-030001.02SourceComment28-March-2018

DICOM Conformance Statement Annexure for IntelliSpace Collaboration Viewer on Philips IntelliSpace Portal V10.0Table 8: General Series ModuleDAPresenceof 8,103ELOANAPCOPY-Performing Physician's Name0008,1050PNANAPCOPY-Operators' Name0008,1070PNANAPCOPY-Operator ed PerformedProcedure Step Sequence0008,1111SQANAPCOPY- Referenced SOP Class UID0008,1150UIALWAYSCOPY- Referenced SOP Instance UID0008,1155UIALWAYSCOPY-Body Part Examined0018,0015CSANAPCOPY-Protocol Name0018,1030LOANAPCOPY-Patient Position0018,5100CSANAPCOPY-Series Instance UID0020,000EUIALWAYSAUTO-Series PCOPY-Smallest Pixel Value in Series0028,0108US/SSANAPCOPY-Largest Pixel Value in Series0028,0109US/SSANAPCOPY-Request Attributes Sequence0040,0275SQANAPCOPY-Performed Procedure Step StartDate0040,0244DAANAPCOPY-Performed Procedure Step StartTime0040,0245TMANAPCOPY-Performed Procedure StepDescription0040,0254LOANAPCOPY-Performed Protocol CodeSequence0040,0260SQANAPCOPY-Comments on the PerformedProcedure Step0040,0280STANAPCOPY-Attribute NameTagVRSeries Date0008,0021Series Time0008,0031ModalitySeries DescriptionValueSourceCommentTable 9: General Equipment ModuleLOPresenceof ONFIG-0008,1010SHANAPCONFIG-Institutional Department Name0008,1040LOANAPCONFIG-Manufacturer's Model Name0008,1090LOANAPFIXEDPhilips PortalDevice Serial Number0018,1000LOANAPCONFIG-Software Version(s)0018,1020LOANAPFIXED10.xSpatial Resolution0018,1050DSANAPCOPY-Date of Last Calibration0018,1200DAANAPCOPY-Time of Last Calibration0018,1201TMANAPCOPY-Pixel Padding Value0028,0120US/SSANAPCOPY-Attribute NameTagVRManufacturer0008,0070Institution Name0008,0080Institution AddressStation Name 2018 Koninklijke Philips N.V.ICAP-T-030001.09bCorresponds to e 8 of 12

DICOM Conformance Statement Annexure for IntelliSpace Collaboration Viewer on Philips IntelliSpace Portal V10.0Page 9 of 12Table 10: SC Equipment ModuleAttribute NameCTPresenceof psIntelliSpace PortalANAPAUTOIntelliSpace ueModality0008,0060CSConversion Type0008,0064CSSecondary Capture Device ID0018,1010LOSecondary Capture DeviceManufacturer0018,1016LOPhilipsSecondary Capture DeviceManufacturer's Model Name0018,1018LOSecondary Capture DeviceSoftware Version(s)0018,1019LOVideo Image Format Acquired0018,1022SHDigital Image Format Acquired0018,1023LOSourceCommentTable 11: General Image ModuleDAPresenceof 0032TMANAPCOPY-Content Time0008,0033TMANAPCOPY-Referenced Image Sequence0008,1140SQANAPCOPY- Referenced SOP Class UID0008,1150UIALWAYSCOPY- Referenced SOP Instance UID0008,1155UIALWAYSCOPY-Derivation Description0008,2111STANAPCOPY-Source Image Sequence0008,2112SQANAPCOPY-Derivation Code Sequence0008,9215SQANAPCOPY-Instance Number0020,0013ISVNAPAUTO,COPY-Patient Orientation0020,0020CSANAPCOPY-Images in Acquisition0020,1002ISANAPCOPY-Quality Control Image0028,0300CSANAPCOPY-Burned In Annotation0028,0301CSANAPAUTO-Lossy Image Compression0028,2110CSANAPCOPY-Lossy Image CompressionRatio0028,2112DSANAPCOPY-Icon Image Sequence0088,0200SQANAPCOPY- Samples per Pixel0028,0002USALWAYSCOPY- Photometric Interpretation0028,0004CSALWAYSCOPY- Planar Configuration0028,0006USANAPCOPY- Rows0028,0010USALWAYSCOPY- Columns0028,0011USALWAYSCOPY- Pixel Aspect Ratio0028,0034ISANAPCOPY- Bits Allocated0028,0100USALWAYSCOPY- Bits Stored0028,0101USALWAYSCOPY- High Bit0028,0102USALWAYSCOPY- Pixel Representation0028,0103USALWAYSCOPY- Smallest Image Pixel Value0028,0106US/SSANAPCOPY-Attribute NameTagVRAcquisition Date0008,0022Content Date0008,0023Acquisition DatetimeAcquisition Time 2018 Koninklijke Philips N.V.ICAP-T-030001.09bValueCorresponds to ICAP-W-030001.02SourceComment28-March-2018

DICOM Conformance Statement Annexure for IntelliSpace Collaboration Viewer on Philips IntelliSpace Portal V10.0 Largest Image Pixel Value0028,0107US/SSANAPCOPY- Red Palette Color LookupTable Descriptor0028,1101US/SSANAPCOPY- Green Palette Color LookupTa

,& 3 7 e &ruuhvsrqgv wr ,& 3 : ,vvxhg e\ 3klolsv 0hglfdo 6\vwhpv 1hghuodqg %9 d 3klolsv hdowkfduh frpsdq\ 3 2 %r[ ' %hvw